1.
|
Combine the following volumes to prepare diluted LAM. For multiple samples, multiply each volume by the number of samples plus 10% overage. |
These volumes produce 5 µl per sample.
Store dilutions of the LAM for up to one week at -25°C to -15°C.
|
2.
|
Add 5 μl diluted LAM to each reaction. |
|
3.
|
Vortex to mix, and then centrifuge briefly. |
|
4.
|
Add 20 μl LPMA to each reaction. |
LPMA is viscous, pipette slowly to ensure that proper volumes are dispensed. The total volume per reaction is 75 μl.
|
5.
|
Pipette 10 times at 40 μl stroke, and then briefly centrifuge. |
|
6.
|
Place in the thermal cycler and run the Ligation program. |