Perform A-Tailing
This snippet causes dropdowns to open on page load. This text/paragraph does not appear in the output.
This step adds a single 'A' nucleotide to the 3' ends of blunt fragments. The A-tail prevents the fragments from ligating to each other during the adapter ligation reaction. A corresponding single 'T' nucleotide on the 3' end of the adapter provides a complementary overhang for ligating the adapter to the fragment.
Preparation
|
1.
|
Prepare ATL4—Vortex briefly on medium-high speed to mix, and then centrifuge briefly. |
|
2.
|
Save the following ATL program on the thermal cycler: |
|
•
|
Choose the preheat lid option and set to 100°C |
|
•
|
Set the reaction volume to 90 μl |
Procedure
|
1.
|
Add 35 μl ATL4 to each well of the LP plate, and pipette 10 times to mix. |
|
2.
|
Apply Microseal 'B', and then centrifuge briefly. |
|
3.
|
Place on the preprogrammed thermal cycler and run the ATL program. |
|
4.
|
When the program reaches the 4°C hold, immediately proceed to the next step. |