Control Materials (Optional)

The control materials described in this section are optional for use with the Illumina Viral Surveillance Panel v2.

Enterobacteria Phage T7 and Phage MS2 Internal Control Sample

Add Enterobacteria Phage T7 and/or Bacteriophage MS2 as an internal spike-in control to each test sample before starting DNA extraction to serve as a process control for nucleic acid extraction. The recommended concentration for these controls is to spike T7 phage at 1.21E+07 copies/ml or spike MS2 phage at 5.38E+08 copies/ml of the sample volume before DNA extraction. The concentration can be adjusted based on your workflow.
The recommended T7 and MS2 phage materials must be purchased directly from the vendor. Contact Microbiologics to request a quote.

Running the optional external controls consumes reagents and reduces the total number of unknown test samples that can be processed with this kit.