Specimen Collection, Transport, and Storage

Handle all specimens as if they are potentially infectious agents.

Whole blood specimens of 7–10 ml must be collected in Streck Cell-Free DNA BCT. Do not freeze.
Transportation of whole blood must comply with all applicable governing regulations for the transport of etiologic agents. Expedited shipping/transportation methods are recommended.
During transport, store at temperatures between 4°C and 30°C. After samples are received, store at 2°C to 8°C until ready to proceed. The time between blood collection and initial plasma isolation should not exceed 5 days.
If retesting is required, postprocessing samples may be recapped and stored at 4°C for an additional 5 days (up to a total of 10 days after blood collection).