Demultiplexing and FASTQ Generation
Sequencing data stored in BCL format are processed through the bcl2fastq conversion software. The bcl2fastq conversion software demultiplexes data and converts BCL files to standard FASTQ file formats for downstream analysis. For each sequencing run, the VeriSeq NIPT Assay Software creates a sample sheet (SampleSheet.csv). This file contains sample information provided to the software during the sample prep process (using the software API). These sample sheets contain a header with information about the run and descriptors for the samples processed in a particular flow cell.
The following table provides sample sheet data details.
Do not modify or edit this sample sheet file. It is system generated and modifications can cause adverse effects downstream, including incorrect results or analysis failure.
Column Name |
Description |
||||||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|---|
SampleID |
Sample identification. |
||||||||||||
SampleName |
Sample name. Default: same as SampleID. |
||||||||||||
Sample_Plate |
Plate identification for a given sample. Default: blank. |
||||||||||||
Sample_Well |
Well identification on the plate for a given sample. |
||||||||||||
I7_Index_ID |
Identification of the first index adapter. |
||||||||||||
index |
Nucleotide sequence of the first adapter. |
||||||||||||
I5_Index_ID |
Identification of the second adapter. |
||||||||||||
index2 |
Nucleotide sequence of the second adapter. |
||||||||||||
Sample_Project |
Project identification for a given sample. Default: blank. |
||||||||||||
SexChromosomes |
Analysis pertaining to sex chromosomes. One of the following:
|
||||||||||||
SampleType |
Sample type. One of the following:
|