Individual Tests
To help troubleshoot hardware problems encountered on the Workflow Manager, the following system tests may be required.
System Test |
Description |
---|---|
Barcode/Autoload |
Tests for proper configuration of the system deck, AutoLoader, and barcode scanning functionality. |
CPAC |
Tests functioning of the CPAC on-deck heater systems. Also checks for proper wiring of individual units to the control box. |
BVS Vacuum |
Tests functioning of the on-deck basic vacuum system (BVS) to confirm that the vacuum can engage and achieve operational pressures. |
Independent Channel |
Tests functioning of the independent pipette channels. Performs liquid retention test to detect dripping of pipette channels and consistency of delivery volumes. |
iSwap |
Tests iSwap robotic arm functioning and confirms gross deck teaching positions. |
96-Head |
Tests functioning of the CO-RE 96 pipette head. Performs liquid retention test to detect dripping of pipette channels and consistency of delivery volumes. |
Execute individual tests as follows.
1. | Select specific test to be executed. |
Full IOQ Execution runs all six tests sequentially.
2. | Follow the onscreen instructions, making note of observations for equipment function and any system errors encountered. |
3. | When finished, select Abort to exit method. |
4. | If asked to provide system trace logs generated during test, the logs are available at C:\Program Files (x86)\HAMILTON\LogFiles and begin with VeriSeqNIPT_Services. |