QC Reason Messages
The qc_reason column in the NIPT Report displays a QC failure or warning when analysis results fall outside the expected range for an analytical QC metric.
QC Reason Message |
Description |
Recommended Action |
---|---|---|
FAILED iFACT |
individual Fetal Aneuploidy Confidence Test (iFACT)—QC metric that combines fetal fraction estimation with run metrics associated with coverage to determine whether the system has statistical confidence to make a call on a given sample. |
Reprocess sample. |
DATA OUTSIDE OF EXPECTED RANGE |
The average deviation from euploid coverage is not consistent with the trained data distribution. Possibly caused by contamination or incorrect sample processing. |
Reprocess sample. |
FRAGMENT SIZE DISTRIBUTION OUTSIDE OF EXPECTED RANGE |
The fragment size distribution is not consistent with the trained data distribution. Possibly caused by contamination or incorrect sample processing. |
Reprocess sample. |
FLOWCELL DATA OUTSIDE OF EXPECTED RANGE |
Flowcell data are not consistent with the trained data distribution. Possibly caused by error in flowcell setup. |
Reprocess sample. |
FAILED TO ESTIMATE FETAL FRACTION |
Unable to produce a valid fetal fraction estimate. |
Reprocess sample. |
SEQUENCING DATA OUTSIDE OF EXPECTED RANGE |
Input sequencing data are not consistent with the trained data distribution. Possibly caused by contamination or incorrect sample processing. |
Resequence flowcell. |
UNEXPECTED DATA |
The report generates a QC concern that does not correspond to any of the other QC reasons listed in this table. |
Contact Illumina Technical Support. |
NTC SAMPLE WITH HIGH COVERAGE |
High coverage detected for an NTC sample (no DNA material expected). Possibly caused by contamination or incorrect sample processing. |
Reprocess sample. |
CANCELLED |
Sample was canceled by a user. |
Not applicable. |
INVALIDATED |
Sample was invalidated by a user. |
Not applicable. |