System liquid leakage
|
|
•
|
Switch off instrument immediately. Perform decontamination and/or maintenance and contact Illumina Technical Support. |
|
Communication error, power or communication is interrupted, or no communication
|
|
•
|
Power on the instrument. |
|
•
|
Check instrument cable and plug. |
|
•
|
Switch off the instrument and PC. Wait until the status lamp is dark and then switch on the instrument and PC. |
|
•
|
Make sure that the water circulator is in Standby mode (on) and has not been turned off completely. |
|
•
|
Power cycle the water circulator using the large main switch on the back of the circulator. |
|
Initialization error
|
|
•
|
Make sure that the arms can move freely so that other objects do not obstruct their movement range. |
|
•
|
Make sure that the front safety door is securely closed on both sides. |
|
•
|
Make sure that the water circulator is in Standby mode (on) and has not been turned off completely. |
|
Front safety panel does not unlock properly
|
|
•
|
Switch off the instrument. If front safety panel does not unlock, contact Illumina Technical Support. |
|
Front safety panel does not lock properly
|
|
•
|
Make sure that the front safety door is securely closed on both sides. |
|
•
|
Power cycle the instrument and test the front safety panel again during a run to see if it locks correctly. |
|
•
|
Switch off the instrument and contact Illumina Technical Support. |
|
Positioning error for PosID3, LiHa, or RoMa (for example, crash, failure to move)
|
|
•
|
Check container, rack, and carrier positions. |
|
•
|
If problem persists, contact Illumina Technical Support. |
|
Barcode not read
|
|
•
|
Check if barcode type is permissible. |
|
•
|
Check container position on carrier. |
|
•
|
Check label quality and test a new barcode label. |
|
•
|
Check settings in application software. |
|
•
|
Check and clean both PosID3 or Keyence scanner windows and make sure their lasers are firing properly. |
|
•
|
If barcode still does not read, manually enter the barcode in ILASS. |
|
Carrier or tube presence not detected
|
|
•
|
Clean the No Tube sensor on the PosID3. |
|
•
|
Make sure the PosID3 reflector is in place and carefully clean its reflective surface. |
|
•
|
If the problem persists, contact Illumina Technical Support. |
|
Unusual noise during PosID3, RoMa, or LiHa movement
|
|
•
|
Contact Illumina Technical Support. |
|
Overflow of wash station
|
|
•
|
Check the wash station outlet tubing for any kinks. |
|
•
|
Make sure the waste tube is not below liquid surface in the waste container. |
|
•
|
Clean the wash station and check for any obstructions. |
|
•
|
Use a waste container with fixed wash tubing inlet. |
|
Damaged or bent tip.
|
|
•
|
Contact Illumina Technical Support. |
|
ILASS error
|
|
•
|
If problem persists, contact Illumina Technical Support. |
|
Insufficient access rights to run Freedom EVO2
|
|
•
|
If permission settings have been altered, check and correct those settings. |
|
•
|
Contact Illumina Technical Support. |
|
EVOWare warnings regarding folders/files in your virtual store
|
|
•
|
Check the following path for a folder labeled Tecan and delete it: C:\Users\<user>\AppData\Local\VirtualStore\ProgramData\. |
|
•
|
If the error persists, contact Illumina Technical Support. |
|