Environmental Considerations
|
Element |
Specification |
||||||
|---|---|---|---|---|---|---|---|
|
Temperature |
|
||||||
|
Humidity |
Maintain a noncondensing relative humidity between 20–80% |
||||||
|
Altitude |
Locate components at an altitude below 2,000 m (6,500 ft). |
||||||
|
Air quality |
Operate components in an indoor environment with air particulate cleanliness levels per ISO 14644-1 Class 9 (ordinary room/laboratory air), or better. Keep components away from sources of dust. |
||||||
|
Ventilation |
Consult your facilities department for ventilation requirements sufficient for the level of heat output expected from components. |
||||||
|
Lighting |
Avoid exposing the Infinium Amplification System to direct sunlight and intense artificial light during operation. |
||||||
|
Vibration |
Limit the continuous vibration of the lab floor to ISO office level. During a run, do not exceed ISO operating room limits. Avoid intermittent shocks or disturbances near the instrument. |
|
Equipment |
Measured Power |
Thermal Output |
|---|---|---|
|
Infinium Amplification System |
750 Watts |
2,559 BTU/h |
|
Performance or Equipment |
Noise Output (dB) |
Distance |
|---|---|---|
|
Infinium Amplification System |
< 65 |
1 m (3.3 ft) |
A measurement of < 62 dB is within the level of a normal conversation at a distance of approximately 1 m (3.3 ft).