Uninterruptible Power Supply
Illumina recommends using an uninterruptible power supply (UPS) to help protect critical runs and provide valuable emergency response time if a sudden power failure occurs. The Infinium Amplification System instrument and its control computer are installed with, and connected to, a UPS.
:
The following table provides UPS recommendations. The output voltage for the recommended models varies based on your region.
|
Specification |
APC Smart UPS SRT 1500 VA 120 V* Part # SRT1500XLA |
|---|---|
|
Maximum output capacity |
1350 W/1500 VA |
|
Input voltage (nominal) |
120 VAC |
|
Input frequency |
50/60 Hz |
|
Input connection |
NEMA 5-15P |
|
Dimensions (H × W × D) |
8.4 cm (3.3 in) × 43.2 cm (17 in) × 50.8 cm (20 in) |
|
Weight |
21.9 kg (48.5 Lb) |
|
Typical combined (UPS+Battery) run time (50% load) |
60 minutes |
|
Typical combined (UPS+Battery) run time (100% load) |
20 minutes |
* Use with APC Smart-UPS SRT 48 V 1 kVA 1.5 kVA Battery Pack (part # SRT48BP).