User Accounts
Illumina instrument control computers running
Permissions |
Administrator |
User |
---|---|---|
Set up, start, and monitor sequencing runs. |
X |
X |
Download and update software. |
X |
|
See status for active run started by other user. |
X |
|
Terminate unresponsive UCS process. |
X |
|
The instrument control computer is designed to operate Illumina sequencing systems. Do not consider it a general-purpose computer. For quality and security reasons, do not use the control computer for web browsing, checking email, reviewing documents, or other unnecessary activity. These activities can result in degraded performance or loss of data.
Illumina instrument control computers running Windows 7 support one account type: Administrator. Illumina recommends upgrading the current software version to be able to use the User account type.
Only use the Administrator account type when it is required.
You are required to set a password upon initial setup of your system. The following table identifies the required password policies for the control computer.
Policy |
Security Setting |
---|---|
Enforce password history |
Five passwords remembered |
Maximum password age |
180 days |
Minimum password age |
0 days |
Minimum password length |
10 characters |
Password must meet complexity requirements |
Disabled |
Store passwords using reversible encryption |
Disabled |
Illumina does not store or maintain customer login credentials, and unknown passwords cannot be reset. An unknown password requires that an Illumina representative restore the factory default, which removes all data from the system and extends the necessary support time.
1. | Press Ctrl+Alt+Delete, and then select Change a password. |
2. | In the Old password field, type the default password. |
3. | In the New password field, type the new password. |
For security, the password must be at least 10 characters long, comply with local policies, and contain numbers, letters, and symbols.
4. | In the Confirm password field, reenter the new password. |
5. | Press Enter to confirm the reset and return to the desktop. |
6. | In the search box, type netplwiz, and then press Enter. |
7. | In the User Accounts dialog box, select the Users must enter a user name and password to use this computer checkbox. |
8. | Select Apply, and then select OK. |
Illumina instrument control computers running
• | Root (super administrator) |
• | Ilmnadmin (administrator) |
• | Ilmnuser (user) |
The Illumina Field Service Engineer initiates a password change for all three accounts after completing instrument installation. Update each password every 180 days, when prompted.
Policy |
Setting |
---|---|
Enforce password history |
Five passwords remembered |
Lockout threshold |
10 invalid log in attempts |
Minimum password length |
10 characters |
Minimum character variety |
Disabled |
Maximum repeating characters |
Disabled |
Password must meet complexity requirements |
Basic checks enabled (for example, checks for qwerty or password) |
Store passwords using reversible encryption |
Disabled |
Password must meet dictionary check requirements |
Disabled |
1. | Log in to ilmnadmin. |
2. | Select the power button, and then expand the ilmnadmin drop-down list. |
3. | Select Account Settings. |
4. | Select Unlock, and then enter the ilmnadmin password. |
5. | Select Add User. |
6. | Select the Standard account type, and then enter a new user name. |
7. | Select Set password now, and then enter a password. |
8. | Select Add. |
The new user is added to the Users list.
9. | Grant the user access to the Control Software (CS) as follows. |
a. | Open terminal. |
b. | Enter the following: |
$ sudo usermod -a -G ilmnusers <new user name>
c. | If prompted, enter the ilmnadmin password. |
10. | To confirm that the user permissions were set successfully, do as follows. |
a. | Log in to the new user account. |
b. | Navigate to the Control Software (CS). |
c. | From the control software menu, select Settings. |
d. | Under Default Output Folder, make sure you can select and save the output folder path. |
If you can select and save the output folder path without any errors, the permissions are successfully set.
This section describes how to reset the ilmnuser, ilmnadmin, and root passwords. Password recovery is not available. Resetting your password does not bypass account lockout after 10 incorrect password attempts. You must wait 10 minutes before you can reset your password or try to log in.
Reset ilmnuser Password
You can reset the ilmnuser password if you know the ilmnadmin or root password.
1. | Log in to ilmnadmin. |
2. | Open terminal. |
3. | Enter sudo passwd ilmnuser. |
4. | Enter the ilmnadmin password at the prompt. |
5. | Enter a new ilmnuser password at the prompt. |
6. | Retype the new ilmnuser password at the prompt to confirm the new password. |
Reset ilmnadmin Password
You can reset the ilmnadmin password if you know the root password.
1. | Log in to root. |
2. | Open terminal. |
3. | Enter passwd ilmnadmin to change the ilmadmin password, or enter passwd ilmnuser to change the ilmnuser password. |
4. | Enter the new password at the prompt. |
5. | Retype the new password at the prompt to confirm the new password. |
Reset Root Password
To reset the root password, do as follows.
1. | Log in to ilmnadmin. |
2. | Open terminal. |
3. | Enter sudo passwd root. |
4. | Enter the new password at the prompt. |
5. | Retype the new password at the prompt to confirm the new password. |