NovaSeq X Series Settings
For descriptions of each field, including the specific requirements and if the expected value is an integer or string, refer to Parameters.
Standalone Sections
Parameter |
Required |
Instrument-Specific Requirements |
---|---|---|
FileFormatVersion |
Yes |
|
RunName |
Yes |
|
RunDescription |
No |
|
InstrumentType |
No |
|
InstrumentPlatform |
No |
Enter NovaSeqXSeries for NovaSeq X Series. |
Custom_* |
No |
Parameter |
Required |
Instrument-Specific Requirements |
|||||||||
---|---|---|---|---|---|---|---|---|---|---|---|
Read1Cycles |
Yes |
||||||||||
Read2Cycles |
No¹ |
||||||||||
Index1Cycles |
No¹ |
Value must be in the 6–12 range for the following run types:
The total number of read cycles and index cycles cannot exceed the number of cycles specified by the reagent kit. The index cycle limit applies to cycles used as an index, not UMI cycles or trimmed reads. |
|||||||||
Index2Cycles |
No¹ |
Value must be in the 6–12 range for the following run types:
The total number of read cycles and index cycles cannot exceed the number of cycles specified by the reagent kit. The index cycle limit applies to cycles used as an index, not UMI cycles or trimmed reads. |
¹ Depending on the run or library prep kit, this parameter might be required.
Parameter |
Required |
Instrument-Specific Requirements |
---|---|---|
CustomRead1Primer |
No¹ |
|
CustomRead2Primer |
No¹ |
|
CustomIndex1Primer |
No¹ |
|
CustomIndex2Primer |
No¹ |
¹ Depending on the run or library prep kit, this parameter might be required.
Application Sections
For NovaSeq X Series, there is a limit of four application sections for onboard analysis or eight application sections for cloud analysis.
Parameter |
Required |
Section |
Instrument-Specific Requirements |
|||||||||
---|---|---|---|---|---|---|---|---|---|---|---|---|
SoftwareVersion |
Yes |
Settings |
|
|||||||||
FastqCompressionFormat |
Yes |
Settings |
Supported values are gzip or dragen. To use DRAGEN Illumina Complete Long Reads WGS (ICLR WGS), the recommended value is gzip. |
|||||||||
AdapterRead1 |
No |
Data |
Possible values are a concatenation from the set [A,C,T,G]. NovaSeq X Series also allows na. A value of na (case insensitive) must be used if both the following statements are accurate:
|
|||||||||
AdapterRead2 |
No |
Data |
Possible values are a concatenation from the set [A,C,T,G]. NovaSeq X Series also allows na. A value of na (case insensitive) must be used if both the following statements are accurate:
|
|||||||||
BarcodeMismatchesIndex1 |
No |
Data |
NovaSeq X Series allows the value na in addition to the values 0, 1, or 2. Enter a value of na if:
|
|||||||||
BarcodeMismatchesIndex2 |
No |
Data |
NovaSeq X Series allows the value na in addition to the values 0, 1, or 2. Enter a value of na if:
|
|||||||||
OverrideCycles |
No |
Data |
OverrideCycles example: Y151;I8N2;N2I8;Y151 For DRAGEN ICLR WGS, use the following guidelines:
|
|||||||||
Sample_ID |
Yes |
Data |
|
|||||||||
Lane |
Yes |
Data |
For NovaSeq X, values must be in the range of lanes specified in RunInfo.xml:
|
|||||||||
Index |
Yes |
Data |
For NovaSeq X, a value of na must be used if no indexes are specified in OverrideCycles for a sample. When using DRAGEN ICLR WGS, leave blank. |
|||||||||
Index2 |
No |
Data |
Always write index sequences in the forward orientation. BCL Convert handles index orientation. When using DRAGEN ICLR WGS, leave blank. |
Parameter |
Required |
Section |
Instrument-Specific Requirements |
---|---|---|---|
SoftwareVersion |
Yes |
Settings |
|
AppVersion |
Yes |
Settings |
|
KeepFastQ |
Yes |
Settings |
|
MapAlignOutFormat |
Yes |
Settings |
If MapAlignOutFormat is set to none, VariantCallingMode cannot be none for any sample. |
ReferenceGenomeDir |
Yes |
Data |
|
Bedfile |
Conditionally required |
Data |
Only required if VariantCallingMode is not None. Must include the prefix DragenEnrichment/ before the BED file name. The value must be na if BedFile is placed in the Data section and VariantCallingMode is set to None. |
GermlineOrSomatic |
Yes |
Data |
|
AuxNoiseBaselineFile |
No |
Data |
The value must be na if AuxNoiseBaselineFile is placed in the Data section and no AuxNoiseBaselineFile is provided for the sample. |
AuxCnvPanelOfNormalsFile |
Conditionally required |
Data |
Optional if VariantCallingMode is AllVariantCallers. The value must be na if AuxCnvPanelOfNormalsFile is placed in the Data section and no AuxCnvPanelOfNormalsFile is provided for the sample. |
VariantCallingMode |
Yes |
Data |
|
Sample_ID |
Yes |
Data |
Parameter |
Required |
Section |
Instrument-Specific Requirements |
---|---|---|---|
SoftwareVersion |
Yes |
Settings |
|
AppVersion |
Yes |
Settings |
|
KeepFastQ |
Yes |
Settings |
|
MapAlignOutFormat |
Yes |
Settings |
If MapAlignOutFormat is set to none, VariantCallingMode cannot be none for any sample. |
ReferenceGenomeDir |
Yes |
Data |
|
VariantCallingMode |
Yes |
Data |
|
QcCoverage1BedFile |
No |
Data |
Only supported for v4.1.23 and later. |
QcCoverage2BedFile |
No |
Data |
Only supported for v4.1.23 and later. |
QcCoverage3BedFile |
No |
Data |
Only supported for v4.1.23 and later. |
QcCrossContaminationVcfFile |
No |
Data |
Only supported for v4.1.23 and later. |
Sample_ID |
Yes |
Data |
Parameter |
Required |
Section |
Instrument-Specific Requirements |
---|---|---|---|
SoftwareVersion |
Yes |
Settings |
|
AppVersion |
Yes |
Settings |
|
MapAlignOutFormat |
Yes |
Settings |
For the DRAGEN RNA application, selecting none is not allowed if RnaPipelineMode is set to MapAlign for any sample. |
KeepFastQ |
Yes |
Settings |
|
DifferentialExpressionEnable |
No |
Settings |
|
ReferenceGenomeDir |
Yes |
Data |
|
RnaGeneAnnotationFile |
No |
Data |
|
RnaPipelineMode |
Yes |
Data |
|
DownSampleNumReads |
No |
Data |
|
Sample_ID |
Yes |
Data |
|
Comparison1 |
No |
Data |
|
Comparison2 |
No |
Data |
|
Comparison3 |
No |
Data |
|
Comparison4 |
No |
Data |
|
Comparison5 |
No |
Data |
Parameter |
Required |
Section |
Instrument-Specific Requirements |
---|---|---|---|
SoftwareVersion |
Yes |
Settings |
|
AppVersion |
Yes |
Settings |
|
KeepFastQ |
Yes |
Settings |
|
MapAlignOutFormat |
Yes |
Settings |
|
ReferenceGenomeDir |
Yes |
Data |
|
AuxNoiseBaselineFile |
No |
Data |
|
AuxSvNoiseBaselineFile |
No |
Data |
|
AuxCnvPopBAlleleVcfFile |
No |
Data |
|
AuxGermlineTaggingFile |
No |
Data |
|
VariantCallingMode |
Yes |
Data |
|
Sample_ID |
Yes |
Data |
Parameter |
Required |
Section |
Instrument-Specific Requirements |
---|---|---|---|
SoftwareVersion |
Yes |
Settings |
|
AppVersion |
Yes |
Settings |
|
KeepFastQ |
Yes |
Settings |
|
MapAlignOutFormat |
Yes |
Settings |
|
UsesTaps |
Yes |
Settings |
|
ReferenceGenomeDir |
Yes |
Data |
|
MethylationProtocol |
No |
Data |
|
Sample_ID |
No |
Data |