NovaSeq X Series Integration v1.0 Configuration and User Interaction
The Illumina NovaSeq X Series Integration Package v1.0 supports the integration of Clarity LIMS to Illumina NovaSeq X series instruments. This integration supports Clarity LIMS v5.1 and later.
For instructions on validating and troubleshooting the Illumina NovaSeq X Series Integration, refer to NovaSeq X Series Integration v1.0 Validation and Troubleshooting.
The configuration provided in this integration has been established to support NovaSeq X Series lab processes. Any configuration changes to protocols or workflows (including renaming protocols, steps, and fields) could break the process.
The integration includes the following features:
| • | Preconfigured NovaSeq X Series Sequencing v1.0 workflow that maps to lab protocols and instrument runs. |
| • | The following steps in the NovaSeq X Series Sequencing v1.0 preconfigured protocol: |
| – | Assign Analysis Configuration Template (NovaSeq X Series Sequencing v1.0) |
| – | Make Bulk Pool (NovaSeq X Series Sequencing v1.0) |
| – | Dilute and Denature (NovaSeq X Series Sequencing v1.0) |
| – | Load to Library Tube Strip (NovaSeq X Series Sequencing v1.0) |
| – | AUTOMATED - Sequencing Run (NovaSeq X Series Sequencing v1.0) |
| • | Automatic validation of run setup information before sample sheet generation or planned run creation. |
| • | Automated generation of the sample sheet. This sample sheet is used to start the sequencing run on the NovaSeq X Series Control Software through Local run mode. |
| • | Automated creation of a planned run on Illumina Connected Analytics (ICA). The NovaSeq X Series Control Software retrieves the planned run for use with the Cloud run mode. |
| • | Automated tracking of the NovaSeq X Series sequencing run and parsing of run statistics into Clarity LIMS using the Stratus Integration Service (SIS). |
| • | [Optional] Preconfigured Library Prep Validation v2.3.0 workflow used for validation purposes only. The workflow contains a single-step protocol that models the library prep required to produce libraries that are ready for the NovaSeq X Series Sequencing v1.0 workflow. For more information, refer to NovaSeq X Series Integration v1.0 Validation and Troubleshooting. The Library Prep Validation v2.3.0 workflow is available in the Illumina Preset Protocols (IPP) v2.5 package. |
| • | [Optional] Clarity LIMS Product Analytics integration that can be used with ICA. The Clarity LIMS Product Analytics integration installation and configuration is required. For more information, refer to Clarity LIMS Product Analytics Service Integration v1.1 Installation and Clarity LIMS Product Analytics Integration v1.1 Configuration and User Interaction. |
The NovaSeq X Series Sequencing v1.0 workflow uses analysis configuration templates (ACTs) to configure secondary analysis for planned runs. Required ACTs must be created using BaseSpace Sequence Hub. The index adapter kit (the label group in Clarity LIMS) selected in the ACT is created in Clarity LIMS. The same label group must be used in the library preparation step. For more information, refer to NovaSeq X Series Integration v1.0 Validation and Troubleshooting.
Make sure that ACT names created in BaseSpace Sequence Hub do not have leading or trailing spaces. Otherwise, Clarity LIMS can have issues with recognizing the ACT names.
Samples must go through the library preparation and quantification process before they enter the NovaSeq X Series Sequencing v1.0 workflow. It is assumed that the following steps have completed before samples are assigned to the workflow:
| • | Samples have been accessioned into Clarity LIMS. |
| • | Samples have been run through QC and library prep. |
| • | Samples have the Molarity (nM) global field set to a value. This value is required for the Calculate Volumes automation in the Make Bulk Pool step. |
For more information on sample accessioning, refer to Sample Accessioning and Upload and Modify Samples in the Getting Started section of the Clarity LIMS Core Functionality documentation.
You can assign samples to workflows automatically, using a routing script, or manually—from the Projects & Samples dashboard. Refer to Assign and Process Samples in the Clarity LIMS Core Functionality documentation.
The NovaSeq X Series Integration Package v1.0 includes the following workflows:
| • | NovaSeq X Series Sequencing v1.0 |
| • | Library Prep Validation v2.3.0 (optional, but recommended for validation purposes) |
The following table describes the protocols and steps included in these workflows.
|
Protocol Name |
Protocol Purpose |
Protocol Steps |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Library Prep Validation v2.3.0 Workflow |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Library Prep Validation v2.3.0 |
Included for validation purposes only, this protocol models the library prep steps required to advance samples to the NovaSeq X Series Sequencing v1.0 workflow. The protocol contains a single step - Library Prep Validation v2.3.0. After this step, a routing script sends the samples to the first step of the NovaSeq X Series Sequencing v1.0 workflow (Assign Analysis Configuration Template (NovaSeq X Series Sequencing v1.0)). |
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
NovaSeq X Series Sequencing v1.0 Workflow |
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
NovaSeq X Series Sequencing v1.0 |
Allows for the assignment of per sample values for the following properties:
|
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||||
The Library Prep Validation v2.3.0 workflow allows for validation of the system after installation is complete. For details, refer to NovaSeq X Series Integration v1.0 Validation and Troubleshooting.
The following table lists the step automations included in the NovaSeq X Series Sequencing v1.0 workflow. Automation details are provided in the documentation for each step.
For details on the validation workflow, refer to NovaSeq X Series Integration v1.0 Validation and Troubleshooting.
|
Protocol |
Step |
Automation |
|||
|---|---|---|---|---|---|
|
NovaSeq X Series Sequencing v1.0 |
|
Register Step Completed¹ Retrieve ACT Information Validate Reagent Labels and Apply Selected ACT to Samples and Set Next Step Validate Sample Names and Retrieve Analysis Configuration Template List and Register Step Started² |
|||
|
Calculate Volumes Register Pools and Register Step Completed¹ Register Step Started¹ Set Next Step Validate Analysis Configurations |
||||
|
Calculate Volumes Register Step Completed¹ Set Next Step Validate Inputs Flowcell Type and Register Step Started² |
||||
|
Register Step Completed¹ Set Next Step Validate Analysis Configurations and Register Step Started² Validate Library Tube Strip Barcode Validate Run Setup and Create Planned Run |
||||
|
Register NovaSeq X RunCompleted and Register Step Completed¹ Register Step Started and Register NovaSeq X Run and Register NovaSeq X Run Association and Register NovaSeq X RunStarted¹ |
¹ These automations are required for Clarity LIMS Product Analytics support only.
² These automations are required for the NovaSeq X Series Sequencing v1.0 workflows and contain additional logic needed for Clarity LIMS Product Analytics support. If you would like to remove support for Clarity LIMS Product Analytics, contact Illumina Support.
All remaining automations are required for the NovaSeq X Series Sequencing v1.0 workflows.
The Illumina Universal Sample Identifier is a global field that is imported with the Library Prep Validation v2.3.0 and NovaSeq X Series Sequencing v1.0 workflows. Illumina Universal Sample Identifier is a text field that is reserved for Clarity LIMS Product Analytics support. The field is optional and not required for this integration.
As of Clarity LIMS v5, the term user-defined field (UDF) has been replaced with custom field in the user interface. However, the API resource is still called UDF.
There are two types of custom fields:
| • | Master step fields—Configured on master steps. Master step fields only apply to the following: |
| • | The master step on which the fields are configured. |
| • | The steps derived from those master steps. |
| • | Global fields—Configured on entities (eg, submitted sample, derived sample, measurement, etc.). Global fields apply to the entire Clarity LIMS system. |