Illumina Genomics Architecture - Library Prep Automated v1.0
The Illumina Genomics Architecture - Library Prep Automated v1.0 includes the following functionality:
| • | Preset Illumina Genomics Architecture - Library Prep Automated v1.0 protocols that support the preparation of samples and libraries for sequencing. |
| • | Can be used with the Illumina Genomics Architecture - NovaSeq Sequencing v1.0 workflow. |
| • | Tight file-based integration with Hamilton robots for performing various steps of liquid handling. |
| • | Automated calculation of sample and buffer volumes. |
| • | Automated calculation or display of reagents. |
| • | Automatic step transition when required. |
Protocol 1: Sample Selection (IGA v1.0)
Master Step 1: Sample Selection
Step Type = No Outputs
Input = Blood, Saliva, or Genomic DNA (gDNA) samples (in any combination)
Output = None
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Row |
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Row |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Library Prep Method |
Text Dropdown |
Required Field = Yes Read Only = No Custom Entries = No Presets:
|
||||||||||||
|
|
Multiline Text Fields |
Quick reference checklist: Sample Selection |
Multiline Text |
Required Fields = No Read Only = Yes Default User tasks for this page:
|
||||||||||||
|
Step File Placeholders |
|
Set Library Prep Method Script |
Log |
Auto |
||||||||||||
|
Sample Table |
|
Table Heading |
Sample Table |
|||||||||||||
|
Enable QC Flags |
No |
|||||||||||||||
|
Sample Display Default |
Collapse |
|||||||||||||||
|
Well Sort Order |
Row |
|||||||||||||||
|
|
Table Columns |
|
|
|
||||||||||||
|
Sample Name |
||||||||||||||||
|
Library Prep Method |
Text |
Global Field = Submitted Sample Required field = No Read Only = No |
||||||||||||||
|
Units |
Text |
Global Field = Submitted Sample Required Field = No Read Only = No |
||||||||||||||
|
Sample Conc. |
Numeric |
|
||||||||||||||
|
Submitted Sample Type |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Validate Sample Attributes (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Sample Selection - Set Library prep method and Next steps (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Sample Selection - Route Plated gDNA (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Plate Samples (IGA v1.0) |
|
|
DNA Batching PCR-Free (IGA v1.0) |
|
|
DNA Batching with Enrichment (IGA v1.0) |
|
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Protocol 2: Sample Plating (IGA v1.0)
Master Step 1: Sample Plating
Step Type = Standard
Input = Blood, Saliva, or Genomic DNA (gDNA) samples (cannot be mixed sample types or mixed library preparation methods when step starts)
Output = Analytes plated on a 96-well plate
|
Section |
Subsection |
Value |
|||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
|||||||||||||||
|
Expanded View Fields |
|
||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Projects |
|||||||||||||||
|
Well Sort Order |
Column |
|
Section |
Subsection |
Value |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||
|
Expanded View Fields |
|
|||||||||||||
|
Defaults |
Sample Grouping |
Group by Projects |
||||||||||||
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 Well Plate |
|
|
Defaults |
Sample Grouping |
Group by Projects |
|
Well Sort Order |
Column |
|
|
|
Placement Pattern |
Column |
|
Skip Alternating Rows |
No |
|
|
Skip Alternating Columns |
No |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Step Sample Type |
Text |
Required Field = No Read Only = Yes |
|||||||||
|
Liquid Handling Instrument ID |
Text |
Required Field = No Read Only = Yes |
|||||||||||
|
|
Multiline Text Fields |
Quick reference checklist: Sample Plating. i) pre Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User tasks for this page:
|
|||||||||
|
Quick reference checklist: Sample Plating. ii) post Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User tasks for this page:
|
|||||||||||
|
Step File Placeholders |
|
Liquid Handler Input File |
File |
Auto |
|||||||||
|
Script Log File (Aggregated) |
Log |
Auto |
|||||||||||
|
Liquid Handler Output File |
File |
Manual |
|||||||||||
|
Sample Table |
|
Table Heading |
Sample Table |
||||||||||
|
Enable QC Flags |
No |
||||||||||||
|
Sample Display Default |
Collapse |
||||||||||||
|
Well Sort Order |
Row |
||||||||||||
|
|
Table Columns |
|
|
|
|||||||||
|
Sample Name |
|||||||||||||
|
Sample Type |
Text |
|
|||||||||||
|
Submitted Sample Type |
Text |
|
|||||||||||
|
Container Name |
|||||||||||||
|
Well |
|||||||||||||
|
Transfer Volume (ul) |
Numeric |
|
|||||||||||
|
Submitted Sample Labware Type |
Text |
|
|||||||||||
|
LIMS ID (Container) |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Sample Plating & Extraction - Validate Single Sample Type and Library Prep Method (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Placement Reminder (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Sample Plating - Sample Placement SAM Barcode Uniqueness Check (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Sample Plating - Set Transfer Volume & Create Liquid Handler Input File (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Sample Plating - Hamilton Output Plate Barcode Check, Store Instrument ID & Set Next Steps to Remove from Workflow before Custom Routing (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Sample Plating - Route Samples (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
DNA Extraction - Blood (IGA v1.0) |
|
|
DNA Extraction - Saliva (IGA v1.0) |
|
|
DNA Batching PCR-Free (IGA v1.0) |
|
|
DNA Batching with Enrichment (IGA v1.0) |
|
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Protocol 3: Blood Extraction (IGA v1.0)
Master Step 1: DNA Extraction - Blood
Step Type = Standard
Input = 96-well plates of Blood
Output = Analytes plated on a 96-well plate
|
Section |
Subsection |
Value |
|||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
|||||||||||||||
|
Expanded View Fields |
|
||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
|||||||||||||||
|
Well Sort Order |
Column |
|
Section |
Subsection |
Value |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
|||||||||
|
Expanded View Fields |
|
||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
|||||||||
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 Well Plate |
|
|
Defaults |
Sample Grouping |
Group by Containers |
|
Well Sort Order |
Column |
|
|
|
Placement Pattern |
Column |
|
Skip Alternating Rows |
No |
|
|
Skip Alternating Columns |
No |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
MLB |
Text |
Required Field = No Read Only = No |
|||||||||
|
ProK |
Text |
Required Field = No Read Only = No |
|||||||||||
|
ITB |
Text |
Required Field = No Read Only = No |
|||||||||||
|
RSB |
Text |
Required Field = No Read Only = No |
|||||||||||
|
Ethanol |
Text |
Required Field = No Read Only = No |
|||||||||||
|
Step Sample Type |
Text |
Required Field = No Read Only = Yes |
|||||||||||
|
Sample Count |
Numeric |
Required Field = No Read Only = Yes Default = 0 |
|||||||||||
|
Liquid Handling Instrument ID |
Text |
Required Field = No Read Only = Yes |
|||||||||||
|
Multiline Text Fields |
Quick reference checklist: DNA Extraction - Blood. i) pre Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
||||||||||
|
Quick reference checklist: DNA Extraction - Blood. ii) post Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
|||||||||||
|
Step File Placeholders |
Liquid Handling Robot (Hamilton) Input File |
File |
Auto |
||||||||||
|
Script Log File (Aggregated) |
Log |
Auto |
|||||||||||
|
Liquid Handling Robot Output File |
File |
Manual |
|||||||||||
|
Sample Table |
Table Heading |
Sample Table |
|||||||||||
|
Enable QC Flags |
No |
||||||||||||
|
Sample Display Default |
Collapse |
||||||||||||
|
|
Well Sort Order |
Row |
|||||||||||
|
|
Table Columns |
|
|
|
|||||||||
|
Sample Name |
|||||||||||||
|
Library Prep Method |
Text |
|
|||||||||||
|
Submitted Sample Type |
Text |
|
|||||||||||
|
Container Name |
|||||||||||||
|
Well |
|||||||||||||
|
LIMS ID (Container) |
|||||||||||||
|
Project Name |
|||||||||||||
|
Container |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Sample Plating & Extraction - Validate Single Sample Type and Library Prep Method (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Extraction - Sample Placement EDP Plate Barcode Uniqueness Check (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Extraction - Count Samples & Create LH Input File (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Extraction - Check LH Output File Barcode & Parse Reagent Info & Check for Missing Barcodes & Set Next Steps to Remove from Workflow (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Extraction - Routing (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
DNA Batching PCR-Free (IGA v1.0) |
|
|
DNA Batching with Enrichment (IGA v1.0) |
|
|
Assign Next Step |
Automatic |
|
Must Move Forward |
|
Protocol 4: Saliva Extraction (IGA v1.0)
Master Step 1: DNA Extraction - Saliva
Step Type = Standard
Input = 96-well plates of Saliva
Output = Analytes plated on a 96-well plate
|
Section |
Subsection |
Value |
|||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
|||||||||||||||
|
Expanded View Fields |
|
||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
|||||||||||||||
|
Well Sort Order |
Column |
|
Section |
Subsection |
Value |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
|||||||||
|
Expanded View Fields |
|
||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
|||||||||
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 Well Plate |
|
|
Defaults |
Sample Grouping |
Group by Containers |
|
Well Sort Order |
Column |
|
|
|
Placement Pattern |
Column |
|
Skip Alternating Rows |
No |
|
|
Skip Alternating Columns |
No |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
MLB |
Text |
Required Field = No Read Only = No |
|||||||||
|
ProK |
Text |
Required Field = No Read Only = No |
|||||||||||
|
ITB |
Text |
Required Field = No Read Only = No |
|||||||||||
|
RSB |
Text |
Required Field = No Read Only = No |
|||||||||||
|
Ethanol |
Text |
Required Field = No Read Only = No |
|||||||||||
|
Step Sample Type |
Text |
Required Field = No Read Only = Yes |
|||||||||||
|
Sample Count |
Numeric |
Required Field = No Read Only = Yes Default = 0 |
|||||||||||
|
Liquid Handling Instrument ID |
Text |
Required Field = No Read Only = No |
|||||||||||
|
Multiline Text Fields |
Quick reference checklist: DNA Extraction - Saliva. i) pre Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
||||||||||
|
Quick reference checklist: DNA Extraction - Saliva. ii) post Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
|||||||||||
|
Step File Placeholders |
Liquid Handling Robot (Hamilton) Input File |
File |
Manual |
||||||||||
|
Script Log File (Aggregated) |
Log |
Auto |
|||||||||||
|
Liquid Handling Robot Output File |
File |
Manual |
|||||||||||
|
Sample Table |
Table Heading |
Sample Table |
|||||||||||
|
Enable QC Flags |
No |
||||||||||||
|
Sample Display Default |
Collapse |
||||||||||||
|
|
Well Sort Order |
Row |
|||||||||||
|
|
Table Columns |
|
|
|
|||||||||
|
Sample Name |
|||||||||||||
|
Library Prep Method |
Text |
|
|||||||||||
|
Submitted Sample Type |
Text |
|
|||||||||||
|
Container Name |
|||||||||||||
|
Well |
|||||||||||||
|
LIMS ID (Container) |
|||||||||||||
|
Project Name |
|||||||||||||
|
Container |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Sample Plating & Extraction - Validate Single Sample Type and Library Prep Method (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Extraction - Sample Placement EDP Plate Barcode Uniqueness Check (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Extraction - Count Samples & Create LH Input File (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Extraction - Check LH Output File Barcode & Parse Reagent Info & Check for Missing Barcodes & Set Next Steps to Remove from Workflow (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Extraction - Routing (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
DNA Batching PCR-Free (IGA v1.0) |
|
|
DNA Batching with Enrichment (IGA v1.0) |
|
|
Assign Next Step |
Automatic |
|
Must Move Forward |
Protocol 5: Illumina DNA PCR-Free (IGA v1.0)
Master Step 1: DNA Batching NFPF
Step Type = Standard
Input = Blood, Saliva, or Genomic DNA (gDNA) samples on 96-well plates (can mix Blood and Saliva, but not Genomic DNA)
Output = Analytes plated on a 96-well plate
|
Section |
Subsection |
Value |
||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||||||||
|
Well Sort Order |
Row |
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Row |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 Well Plate |
|
|
Defaults |
Sample Grouping |
Group by Containers |
|
Well Sort Order |
Column |
|
|
|
Placement Pattern |
Column |
|
Skip Alternating Rows |
No |
|
|
Skip Alternating Columns |
No |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Final Volume (µl) |
Numeric |
Required Field = Yes Read Only = No Default = 35 |
||||||||||||
|
Input Amount (ng) |
Numeric |
Required Field = Yes Read Only = No Default = 350 |
||||||||||||||
|
Liquid Handling Instrument ID |
Text |
Required Field = No Read Only = Yes |
||||||||||||||
|
Multiline Text Fields |
Quick reference checklist: DNA Batching IDPF. i) pre Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
|||||||||||||
|
Quick reference checklist: DNA Batching IDPF. ii) post Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
||||||||||||||
|
Step File Placeholders |
Liquid Handler Batching Input File |
File |
Auto |
|||||||||||||
|
Script Log File (Aggregated) |
Log |
Auto |
||||||||||||||
|
Output File from Liquid Handler |
File |
Manual |
||||||||||||||
|
Sample Table |
Table Heading |
Sample Table |
||||||||||||||
|
Enable QC Flags |
No |
|||||||||||||||
|
Sample Display Default |
Collapse |
|||||||||||||||
|
|
Well Sort Order |
Row |
||||||||||||||
|
|
Table Columns |
|
|
|
||||||||||||
|
Sample Name |
||||||||||||||||
|
Sample Needed (µl) |
Numeric |
|
||||||||||||||
|
Final Amount (ng) |
Numeric |
|
||||||||||||||
|
Diluent Volume (µl) |
Numeric |
|
||||||||||||||
|
Warnings |
Text |
|
||||||||||||||
|
LIMS ID (Container) |
||||||||||||||||
|
Project Name |
||||||||||||||||
|
Well |
||||||||||||||||
|
Container |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
DNA Batching PF - Set Final Volume and Validate Sample Type Combination (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Placement Reminder (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
DNA Batching - Sample Placement DNA Plate Barcode Uniqueness Check (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
DNA Batching PF - Generate LH Input File (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
DNA Batching - Check LH Output File Barcode, Store Instrument ID, Set Next Steps to Advance (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Indexing Planning PCR-Free (IGA v1.0) |
|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
If Illumina extraction kit is used: Must Move Forward |
|
|
If external extraction kit is used: Safe Stopping Point |
Master Step 2: Indexing Planning
Step Type = No Outputs
Input = Blood, Saliva, or Genomic DNA (gDNA) 96-well batch plates
Output = No Outputs
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Row |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Index Plate Barcode |
Text |
Required Field = Yes Read Only = No |
|||||||||
|
Hamilton Indexing Offset Column |
Numeric |
Required Field = No Read Only = No Default = 1 Range = 1 to 12 |
|||||||||||
|
Multiline Text Fields |
Quick reference checklist:: Indexing Planning |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
||||||||||
|
Step File Placeholders |
Script Log File (Aggregated) |
Log |
Manual |
||||||||||
|
Sample Table |
Table Heading |
Sample Table |
|||||||||||
|
Enable QC Flags |
No |
||||||||||||
|
Sample Display Default |
Collapse |
||||||||||||
|
|
Well Sort Order |
Row |
|||||||||||
|
Table Columns |
Sample Name |
||||||||||||
|
Container Name |
|||||||||||||
|
Well |
|||||||||||||
|
LIMS ID (Container) |
|||||||||||||
|
Project Name |
|||||||||||||
|
INTERNALFIELD_IndexPlanning_Metadata |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Indexing Planning - Validate and Check for Duplicates of Index Plate Barcode, Validate Indexing Offset, and Set Next Steps (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Library Prep - Indexing PCR-Free (IGA v1.0) |
|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
If Illumina extraction kit is used: Must Move Forward |
|
|
If external extraction kit is used: Safe Stopping Point |
Master Step 3: Indexing NFPF
Step Type = Add Labels
Input = Blood, Saliva, or Genomic DNA (gDNA) 96-well batch plates
Output = Indexed 96-well plates
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Row |
|
Section |
Subsection |
Value |
|---|---|---|
|
Label Groups |
Nextera UDP A |
96 Labels |
|
Nextera UDP B |
96 Labels |
|
|
Nextera UDP C |
96 Labels |
|
|
Nextera UDP D |
96 Labels |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Sample Count |
Numeric |
Required Field = No Read Only = Yes Default = 0 Decimal Places Displayed = 0 |
||||||||||||
|
TB1 |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
ST2 |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
TWB |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
ELM-T |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
HP3 |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
IPB |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
RSB |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
BLT-PF |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
Library Plate Barcode |
Text |
Required Field = Yes Read Only = No |
||||||||||||||
|
Index Plate Barcode (read from input samples) |
Text |
Required Field = No Read Only = Yes |
||||||||||||||
|
Index Plate Offset (read from input samples) |
Text |
Required Field = No Read Only = Yes |
||||||||||||||
|
Liquid Handling Instrument ID |
Text |
Required Field = No Read Only = Yes |
||||||||||||||
|
Adapter |
Text |
Required Field = Yes Read Only = Yes Default = CTGTCTCTTATACACATCT |
||||||||||||||
|
Adapter Read 2 |
Text |
Required Field = Yes Read Only = Yes Default = CTGTCTCTTATACACATCT |
||||||||||||||
|
Multiline Text Fields |
Quick reference checklist: Library Prep IDPF. i) pre Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
|||||||||||||
|
Quick reference checklist: Library Prep IDPF. ii) post Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
||||||||||||||
|
Step File Placeholders |
Liquid Handler Input File |
File |
Auto |
|||||||||||||
|
Script Log File (Aggregated) |
Log |
Auto |
||||||||||||||
|
Hamilton Output File |
File |
Manual |
||||||||||||||
|
Sample Table |
Table Heading |
Sample Table |
||||||||||||||
|
Enable QC Flags |
No |
|||||||||||||||
|
Sample Display Default |
Collapse |
|||||||||||||||
|
|
Well Sort Order |
Row |
||||||||||||||
|
Table Columns |
Sample Name |
|||||||||||||||
|
Library Prep Method |
Text |
|
||||||||||||||
|
Normalized Molarity (nM) |
Numeric |
|
||||||||||||||
|
Container Name |
||||||||||||||||
|
Well |
||||||||||||||||
|
LIMS ID (Container) |
||||||||||||||||
|
Container |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Indexing - Validate Index Category and Index Plate Barcode (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Indexing - Custom Indexing with Hamilton Offsets and Dynamic Index Category Discovery (IGA v1.0) |
Trigger Location |
Add Labels |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Library Prep - Indexing PF - Validate/Copy LibPlate Barcode, Set Default Norm. Mol. if Applicable, and Generate LH Input File (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Library Prep - Indexing PF - Check for Valid, Unique LIB Plate Barcode & Match with Liquid Handler Output File & Parse Reagent Info & Set Next Steps to Remove from Workflow & Quant. Warning (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Library prep - Indexing PF - Route Samples to Sequencing Platform Selection with Automatic Workflow Detection (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Select Sequencing Platform (IGA v1.0) |
|
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
Protocol 6: Illumina DNA with Enrichment (IGA v1.0)
Master Step 1: DNA Batching NFE
Step Type = Standard
Input = Blood, Saliva, or Genomic DNA (gDNA) samples on 96-well plates (can mix types)
Output = Analytes plated on a 96-well plate
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Row |
|
Section |
Subsection |
Value |
|---|---|---|
|
Destination Containers |
96 Well Plate |
|
|
Defaults |
Sample Grouping |
Group by Containers |
|
Well Sort Order |
Row |
|
|
|
Placement Pattern |
Column |
|
Skip Alternating Rows |
No |
|
|
Skip Alternating Columns |
No |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Final Volume (µl) |
Numeric |
Required Field = Yes Read Only = No Default = 35 |
||||||||||||
|
Input Amount (ng) |
Numeric |
Required Field = Yes Read Only = No Default = 450 |
||||||||||||||
|
Liquid Handling Instrument ID |
Text |
Required Field = No Read Only = Yes |
||||||||||||||
|
Multiline Text Fields |
Quick reference checklist i): DNA Batching IDE, pre Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
|||||||||||||
|
Quick reference checklist ii): DNA Batching IDE, post Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
||||||||||||||
|
Step File Placeholders |
Liquid Handler Batching Input File |
File |
Auto |
|||||||||||||
|
Script Log File (Aggregated) |
Log |
Auto |
||||||||||||||
|
Output File from Liquid Handler |
File |
Manual |
||||||||||||||
|
Sample Table |
Table Heading |
Sample Table |
||||||||||||||
|
Enable QC Flags |
No |
|||||||||||||||
|
Sample Display Default |
Collapse |
|||||||||||||||
|
|
Well Sort Order |
Row |
||||||||||||||
|
Table Columns |
Sample Name |
|||||||||||||||
|
Sample Needed (µl) |
Numeric |
|
||||||||||||||
|
Final Amount (ng) |
Numeric |
|
||||||||||||||
|
Diluent Volume (µl) |
Numeric |
|
||||||||||||||
|
Warnings |
Text |
|
||||||||||||||
|
LIMS ID (Container) |
||||||||||||||||
|
Project Name |
||||||||||||||||
|
Well |
||||||||||||||||
|
Container Name |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
DNA Batching Enrichment - Set Final Volume (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Placement Reminder (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
DNA Batching - Sample Placement DNA Plate Barcode Uniqueness Check (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
DNA Batching Enrichment - Generate LH Input File (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
DNA Batching - Check LH Output File Barcode, Store Instrument ID, Set Next Steps to Advance (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Indexing Planning with Enrichment (IGA v1.0) |
|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
If Illumina extraction kit used: Must Move Forward. |
|
|
If external extraction kit used: Safe Stopping Point. |
Master Step 2: Indexing Planning
Step Type = No Outputs
Input = Blood, Saliva, or Genomic DNA (gDNA) 96-well batch plates
Output = No Outputs
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Row |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Index Plate Barcode |
Text |
Required Field = Yes Read Only = No |
|||||||||
|
Hamilton Indexing Offset Column |
Numeric |
Required Field = No Read Only = No Default = 1 Range = 1 to 12 |
|||||||||||
|
Multiline Text Fields |
Quick reference checklist:: Indexing Planning |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
||||||||||
|
Step File Placeholders |
Script Log File (Aggregated) |
Log |
Manual |
||||||||||
|
Sample Table |
Table Heading |
Sample Table |
|||||||||||
|
Enable QC Flags |
No |
||||||||||||
|
Sample Display Default |
Collapse |
||||||||||||
|
|
Well Sort Order |
Row |
|||||||||||
|
Table Columns |
Sample Name |
||||||||||||
|
INTERNALFIELD_IndexPlanning_Metadata |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Indexing Planning - Validate and Check for Duplicates of Index Plate Barcode, Validate Indexing Offset, and Set Next Steps (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Library Prep - Indexing with Enrichment (IGA v1.0) |
|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
If Illumina extraction kit used: Must Move Forward. |
|
|
If external extraction kit used: Safe Stopping Point. |
Master Step 3: Indexing NFE
Step Type = Add Labels
Input = Blood, Saliva, or Genomic DNA (gDNA) 96-well batch plates
Output = Indexed 96-well plates
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Row |
|
Section |
Subsection |
Value |
|---|---|---|
|
Label Groups |
Nextera UDP A |
96 Labels |
|
Nextera UDP B |
96 Labels |
|
|
Nextera UDP C |
96 Labels |
|
|
Nextera UDP D |
96 Labels |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Index Plate Barcode (read from input samples) |
Text |
Required Field = No Read Only = Yes |
|||||||||
|
Index Plate Offset (read from input samples) |
Text |
Required Field = No Read Only = Yes |
|||||||||||
|
Multiline Text Fields |
Quick reference checklist: Library Prep IDE |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
||||||||||
|
Step File Placeholders |
Script Log File (Aggregated) |
Log |
Auto |
||||||||||
|
Sample Table |
Table Heading |
Sample Table |
|||||||||||
|
Enable QC Flags |
No |
||||||||||||
|
Sample Display Default |
Collapse |
||||||||||||
|
|
Well Sort Order |
Row |
|||||||||||
|
Table Columns |
Sample Name |
||||||||||||
|
INTERNALFIELD_IndexPlanning_Metadata |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Indexing - Validate Index Category and Index Plate Barcode (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Indexing - Custom Indexing with Hamilton Offsets and Dynamic Index Category Discovery (IGA v1.0) |
Trigger Location |
Add Labels |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Library Prep - Indexing Enrichment - Copy Input Plate Barcode to Output Plate (IGA v1.0) |
Trigger Location |
Add Labels |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Library Prep - Indexing Enrichment - Copy Indexing Metadata to Outputs & Set Next Steps to Advance (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Library Prep - Pre-Capture Pooling with Enrichment (IGA v1.0) |
|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Must Move Forward |
Master Step 4: Library Prep NFE - Pool Libraries
Step Type = Pooling
Input = Blood, Saliva, or Genomic DNA (gDNA) 96-well batch plates
Output = One 96-well plate of pooled samples
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Row |
|
Section |
Subsection |
Value |
|---|---|---|
|
Pooling Settings |
Label Uniqueness |
On |
|
Defaults |
Sample Grouping |
Group by Containers |
|
Well Sort Order |
Row |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 Well Plate |
|
|
Defaults |
Well Sort Order |
Row |
|
|
Placement Pattern |
Column |
|
Skip Alternating Rows |
No |
|
|
Skip Alternating Columns |
No |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Sample Count |
Numeric |
Required Field = No Read Only = Yes Default = 0 |
||||||||||||
|
Liquid Handling Instrument ID |
Text |
Required Field = No Read Only = Yes |
||||||||||||||
|
Index Plate Barcode (read from input samples) |
Text |
Required Field = No Read Only = Yes |
||||||||||||||
|
Index Plate Offset (read from input samples) |
Text |
Required Field = No Read Only = Yes |
||||||||||||||
|
Total Pool Volume (µl) |
Numeric |
Required Field = No Read Only = No Default = 30 |
||||||||||||||
|
eBLT |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
TB1 |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
ST2 |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
TWB |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
EPM |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
SPB |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
RSB |
Text |
Required Field = No Read Only = No |
||||||||||||||
|
Adapter |
Text |
Required Field = Yes Read Only = Yes Default = CTGTCTCTTATACACATCT |
||||||||||||||
|
Adapter Read 2 |
Text |
Required Field = Yes Read Only = No Default = CTGTCTCTTATACACATCT |
||||||||||||||
|
Multiline Text Fields |
Quick reference checklist: Library Prep IDE - pool libraries. i) pre Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
|||||||||||||
|
Quick reference checklist: Library Prep IDE - pool libraries. ii) post Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
||||||||||||||
|
Step File Placeholders |
Liquid Handler Input File |
File |
Auto |
|||||||||||||
|
|
Script Log File (Aggregated) |
Log |
Auto |
|||||||||||||
|
Liquid Handler Output File |
File |
Manual |
||||||||||||||
|
Sample Table |
|
Table Heading |
Sample Table |
|||||||||||||
|
Enable QC Flags |
No |
|||||||||||||||
|
Sample Display Default |
Collapse |
|||||||||||||||
|
|
Well Sort Order |
Row |
||||||||||||||
|
Table Columns |
Sample Name |
|||||||||||||||
|
Sample Needed (µl) |
Numeric |
|
||||||||||||||
|
Diluent Volume (µl) |
Numeric |
|
||||||||||||||
|
Number of Samples in Pool |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Library prep - Pooling Enrichment - Validate Pool Sizes (IGA v1.0) |
Trigger Location |
Pooling |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Library Prep - Pooling Enrichment - Sample Placement PCPP Plate Barcode Uniqueness Check (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Library Prep - Pooling Enrichment - Create LH Input File (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Library Prep - Pooling Enrichment - Validate LH Output File Name, Parse Reagents & Set Next Steps to Advance (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Library Prep - Capture and Wash with Enrichment (IGA v1.0) |
|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Must Move Forward |
Master Step 5: Capture and Wash NFE
Step Type = Standard
Input = Blood, Saliva, or Genomic DNA (gDNA) 96-well batch plates
Output = Indexed 96-well plates
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Row |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 Well Plate |
|
|
Defaults |
Sample Grouping |
Group by Containers |
|
|
Well Sort Order |
Row |
|
|
Placement Pattern |
Same Shape |
|
Skip Alternating Rows |
No |
|
|
Skip Alternating Columns |
No |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Pool Count |
Numeric |
Required Field = No Read Only = Yes Default = 0 |
||||||
|
Liquid Handling Instrument ID |
Text |
Required Field = No Read Only = Yes |
||||||||
|
EHB2 |
Text |
Required Field = No Read Only = No |
||||||||
|
Probes |
Text |
Required Field = No Read Only = No |
||||||||
|
EEW |
Numeric |
Required Field = No Read Only = No |
||||||||
|
EE1 |
Text |
Required Field = No Read Only = No |
||||||||
|
HP3 |
Text |
Required Field = No Read Only = No |
||||||||
|
ET2 |
Text |
Required Field = No Read Only = No |
||||||||
|
EPM |
Text |
Required Field = No Read Only = No |
||||||||
|
PPC |
Text |
Required Field = No Read Only = No |
||||||||
|
Multiline Text Fields |
Quick reference checklist: Capture and Wash IDE. i) pre Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
|||||||
|
Quick reference checklist: Capture and Wash IDE. i) post Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
||||||||
|
Step File Placeholders |
Liquid Handler Input File |
File |
Auto |
|||||||
|
|
Script Log File (Aggregated) |
Log |
Auto |
|||||||
|
Liquid Handler Output File |
File |
Manual |
||||||||
|
Sample Table |
|
Table Heading |
Sample Table |
|||||||
|
Enable QC Flags |
No |
|||||||||
|
Sample Display Default |
Collapse |
|||||||||
|
|
Well Sort Order |
Row |
||||||||
|
Table Columns |
Sample Name |
|||||||||
|
Container Name |
||||||||||
|
Well |
||||||||||
|
LIMS ID (Container) |
||||||||||
|
Project Name |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Capture and Wash Enrichment - Sample Placement LIB Plate Barcode Uniqueness Check & Update Pool Count (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Capture and Wash Enrichment - Create LH Input File (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Capture and Wash Enrichment - Validate LH Output File Name, Parse Reagents & Set Next Steps to Advance (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Post-Capture Pool QC with Enrichment (IGA v1.0) |
|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
Master Step 6: Generic Pool QC NFE
Step Type = Standard QC
Input = Blood, Saliva, or Genomic DNA (gDNA) 96-well captured plates
Output = No Outputs
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Row |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Multiline Text Fields |
Quick reference checklist: Pool QC IDE |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
||||||||||||
|
Step File Placeholders |
QC Instrument Results (Generic Placeholder) |
File |
Manual |
|||||||||||||
|
|
Script Log File (Aggregated) |
Log |
Auto |
|||||||||||||
|
Sample Table |
|
Table Heading |
Sample Table |
|||||||||||||
|
Enable QC Flags |
Yes |
|||||||||||||||
|
Sample Display Default |
Expand |
|||||||||||||||
|
|
Well Sort Order |
Row |
||||||||||||||
|
File Column Options |
File Column Display |
Hide |
||||||||||||||
|
File Attachment Method |
Auto |
|||||||||||||||
|
Table Columns |
Sample Name |
|||||||||||||||
|
Container Name |
||||||||||||||||
|
Well |
||||||||||||||||
|
Normalized Molarity (nM) |
Numeric |
|
||||||||||||||
|
Normalized Molarity (nM) |
Numeric |
|
||||||||||||||
|
LIMS ID (Container) |
||||||||||||||||
|
Project Name |
||||||||||||||||
|
INTERNALFIELD_QCValue_forRouting |
Text |
|
||||||||||||||
|
INTERNALFIELD_QCValue_forRouting |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Pool QC - Generic - Enrichment - Copy Normalized Molarity from Output to Input, Copy QC Value to Inputs & Set QC Passed Sample Next Steps to Remove from Workflow (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Pool QC - Generic - Enrichment - Route QC Passed Samples to Sequencing Platform Selection (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Select Sequencing Platform (IGA v1.0) |
|
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
Protocol 7: Sequencing Platform Selection (IGA v1.0)
Master Step 1: Sequencing Platform Selection
Step Type = No Outputs
Input = Indexed or Pool/Captured 96-well plates
Output = None
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Row |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Sequencing Platform |
Text |
Required Field = Yes Read Only = Yes Default = NovaSeq_IGA |
|||||||||
|
|
Multiline Text Fields |
Quick reference checklist: Sequencing Platform Selection |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
|||||||||
|
Step File Placeholders |
Set Sequencing Platform Script Log |
Log |
Auto |
||||||||||
|
Sample Table |
|
Table Heading |
Sample Table |
||||||||||
|
Enable QC Flags |
No |
||||||||||||
|
Sample Display Default |
Collapse |
||||||||||||
|
|
Well Sort Order |
Row |
|||||||||||
|
Table Columns |
Sample Name |
||||||||||||
|
Container Name |
|||||||||||||
|
Well |
|||||||||||||
|
LIMS ID (Container) |
|||||||||||||
|
Project Name |
|||||||||||||
|
Sequencing Platform |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Sequencing Platform Selection - Copy Instrument Type to Analyte & Set Next Steps to Remove from Workflow (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Sequencing Platform Selection - Route to Sequencing Workflow by Type (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Define Run Format (IGA v1.0) |
|
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |