Illumina Genomics Architecture - NovaSeq Sequencing v1.0
The Illumina Genomics Architecture - NovaSeq Sequencing v1.0 includes the following functionality:
| • | Preset Illumina Genomics Architecture - NovaSeq Sequencing v1.0 protocols that support the setup of previously prepared libraries for sequencing on the NovaSeq platform. |
| • | Can be used with the Illumina Genomics Architecture - Library Prep Automated v1.0 workflow. |
| • | Tight file-based integration with Hamilton robots for performing various steps of liquid handling. |
| • | Automated calculation of sample and buffer volumes. |
| • | Automated calculation or display of reagents. |
| • | Automatic step transition when required. |
Protocol 1: Run Format (IGA v1.0)
Master Step 1: Define Run Format (IGA v1.0)
Step Type = No Outputs
Input = Indexed or Pooled/Captured 96-well plates
Output = Indexed or Pooled/Captured 96-well plates
|
Section |
Subsection |
Value |
|||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
|||||||||||||||
|
Expanded View Fields |
|
||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
|||||||||||||||
|
Well Sort Order |
Column |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Multiline Text Fields |
Quick Reference Checklist: Define Run Format |
Multiline Text |
Read Only Default User tasks for this page:
Field Order: Vertical |
|||||||||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||||||||||||||
|
Sample Table |
|
Table Heading |
Sample Table |
||||||||||||||||||||||||||||
|
Enable QC Flags |
No |
||||||||||||||||||||||||||||||
|
Sample Display Default |
Expand |
||||||||||||||||||||||||||||||
|
Well Sort Order |
Column |
||||||||||||||||||||||||||||||
|
|
Table Columns |
|
|
|
|||||||||||||||||||||||||||
|
Sample Name |
|||||||||||||||||||||||||||||||
|
Warning |
Text Dropdown |
Presets
|
|||||||||||||||||||||||||||||
|
Loading Workflow Type |
Text Dropdown |
Presets
|
|||||||||||||||||||||||||||||
|
Flowcell Type |
Text Dropdown |
Presets
|
|||||||||||||||||||||||||||||
|
Final Loading Concentration (pM) |
Numeric Dropdown |
Presets
|
|||||||||||||||||||||||||||||
|
Normalized Molarity (nM) |
Numeric |
|
|||||||||||||||||||||||||||||
|
Container Name |
|||||||||||||||||||||||||||||||
|
Well |
|||||||||||||||||||||||||||||||
|
LIMS ID (Container) |
|||||||||||||||||||||||||||||||
|
Project Name |
|||||||||||||||||||||||||||||||
|
Per Sample Volume (µl) |
Numeric |
|
|||||||||||||||||||||||||||||
|
Adjusted Per Sample Volume (µl) |
Numeric |
|
|||||||||||||||||||||||||||||
|
Minimum Molarity (nM) |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Define Run Format - Assign Workflow and Flowcell Type to Samples & Set Next Steps to Remove from Workflow (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Define Run Format - Routing Script (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Make Bulk Pool for NovaSeq Standard (IGA v1.0) |
|
|
Make Bulk Pool for NovaSeq Xp (IGA v1.0) |
|
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Protocol 2: NovaSeq Standard (IGA v1.0)
Master Step 1: Make Bulk Pool for NovaSeq Standard (IGA v1.0)
Step Type = Pooling
Input = Indexed or Pool/Captured 96-well plates
Output = Pooled libraries in a Bulk Pool Plate.
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Column |
|
Section |
Subsection |
Value |
|---|---|---|
|
Pool Settings |
Label Uniqueness |
On |
|
Defaults |
Sample Grouping |
Group by Containers |
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 Well Plate |
|
|
Defaults |
Well Sort Order |
Column |
|
|
Placement Pattern |
Column |
|
Skip Alternating Rows |
No |
|
|
Skip Alternating Columns |
No |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Number of Flowcells to Sequence |
Numeric |
Required Field = Yes Read Only = No Default = 1 Range = 1 to 10 Decimal Places Displayed = 0 |
||||||||||||||||||||||||
|
Minimum Per Sample Volume (µl) |
Numeric |
Required Field = Yes Read Only = No Default = 2 Decimal Places Displayed = 2 |
||||||||||||||||||||||||||
|
Liquid Handling Instrument ID |
Text |
Required Field = Yes Read Only = No |
||||||||||||||||||||||||||
|
|
Multiline Text Fields |
Quick Reference Checklist: Make Bulk Pool for NovaSeq Standard. i) pre Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User tasks for this page:
|
||||||||||||||||||||||||
|
Quick Reference Checklist: Make Bulk Pool for NovaSeq Standard. i) pre Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User tasks for this page:
|
||||||||||||||||||||||||||
|
Step File Placeholders |
|
Calculation |
File |
Auto |
||||||||||||||||||||||||
|
Calculation |
Log |
Auto |
||||||||||||||||||||||||||
|
Hamilton Make Bulk Pool Input |
File |
Auto |
||||||||||||||||||||||||||
|
Hamilton |
Log |
Auto |
||||||||||||||||||||||||||
|
Hamilton Output |
File |
Manual |
||||||||||||||||||||||||||
|
Sample Table |
|
Table Heading |
Sample Table |
|||||||||||||||||||||||||
|
Enable QC Flags |
No |
|||||||||||||||||||||||||||
|
Sample Display Default |
Expand |
|||||||||||||||||||||||||||
|
Well Sort Order |
Column |
|||||||||||||||||||||||||||
|
|
Table Columns |
|
|
|
||||||||||||||||||||||||
|
Sample Name |
||||||||||||||||||||||||||||
|
RSB Volume (µl) |
Numeric |
|
||||||||||||||||||||||||||
|
Volume of Pool to Denature (µl) |
Numeric |
|
||||||||||||||||||||||||||
|
NaOH Volme (µl) |
Numeric |
|
||||||||||||||||||||||||||
|
Tris-HCl Volume (µl) |
Numeric |
|
||||||||||||||||||||||||||
|
Flowcell Type |
Text Dropdown |
Presets
|
||||||||||||||||||||||||||
|
Loading Workflow Type |
Text Dropdown |
Required Field = Yes Read Only = No Custom Entries = No Presets:
|
||||||||||||||||||||||||||
|
Container Name |
||||||||||||||||||||||||||||
|
Well |
||||||||||||||||||||||||||||
|
LIMS ID (Container) |
||||||||||||||||||||||||||||
|
Project Name |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Validate Inputs Flowcell Type and Single Pool (IGA v1.0) |
Trigger Location |
Pooling |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
NovaSeq Make Bulk Pool - Output Plate BPP Barcode Check (Duplicates Allowed) (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Make Bulk Pool NovaSeq Standard - Calculate Volumes and Generate Hamilton Input File (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual button |
||
|
Command line
|
|||
|
Make Bulk Pool NovaSeq Standard - Validate Hamilton Output File Name if Present, Parse Instrument ID, Set Next Steps to Advance (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Dilute and Denature (IGA v1.0) |
|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Master Step 2: Dilute and Denature (IGA v1.0)
Step Type = STANDARD
Input = Bulk Pooled Libraries
Output = Bulk pooled libraries prepared in a Library Tube for sequencing.
|
Kit Name |
Supplier |
|---|---|
|
NaOH |
|
|
Resuspension Buffer (RSB) |
|
|
Tris-HCl |
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Library Tube |
|
|
Defaults |
Sample Grouping |
Group by Containers |
|
Well Sort Order |
Column |
|
|
|
Placement Pattern |
Column |
|
Skip Alternating Rows |
No |
|
|
Skip Alternating Columns |
No |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Experiment Name |
Text |
Required Field = Yes Read Only = No |
||||||||||||||||||||||||
|
BaseSpace Sequence Hub Configuration |
Text Dropdown |
Required Field = Yes Read Only = No Custom Entries = No Presets:
|
||||||||||||||||||||||||||
|
Workflow |
Text |
Required Field = No Read Only = Yes Default = GenerateFASTQ |
||||||||||||||||||||||||||
|
Workflow Type |
Text Dropdown |
Required Field = Yes Read Only = No Custom Entries = No
|
||||||||||||||||||||||||||
|
Index Read 1 |
Numeric Dropdown |
Required Field = Yes Read Only = No Custom Entries = Yes Range = 0 to 20 Decimal Places Displayed = 0 Presets:
|
||||||||||||||||||||||||||
|
Index Read 2 |
Numeric Dropdown |
Required Field = Yes Read Only = No Custom Entries = Yes Range = 0 to 20 Decimal Places Displayed = 0 Presets:
|
||||||||||||||||||||||||||
|
Paired End |
Text Dropdown |
Required Field = Yes Read Only = No Custom Entries = No Presets:
|
||||||||||||||||||||||||||
|
Read Cycles 1 |
Numeric Dropdown |
Required Field = Yes Read Only = No Custom Entries = Yes Range = 1 to 301 Decimal Places Displayed = 0 Presets:
|
||||||||||||||||||||||||||
|
Read Cycles 2 |
Numeric Dropdown |
Required Field = Yes Read Only = No Custom Entries = Yes Range = 0 to 301 Decimal Places Displayed = 0 Presets:
|
||||||||||||||||||||||||||
|
Use Custom Read 1 Primer |
Toggle Switch |
Required Field = Yes Read Only = No Default = No |
||||||||||||||||||||||||||
|
Use Custom Read 2 Primer |
Toggle Switch |
Required Field = Yes Read Only = No Default = No |
||||||||||||||||||||||||||
|
Use Custom Index Read 1 Primer |
Toggle Switch |
Required Field = No Read Only = No Default = No |
||||||||||||||||||||||||||
|
Output Folder |
Text |
Required Field = Yes Read Only = No |
||||||||||||||||||||||||||
|
Use Custom Recipe |
Toggle Switch |
Required Field = Yes Read Only = No Default = No |
||||||||||||||||||||||||||
|
Custom Recipe Path |
Text |
Required Field = No Read Only = No |
||||||||||||||||||||||||||
|
Adapter |
Text |
Required Field = Yes Read Only = No Default = CTGTCTCTTATACACATCT |
||||||||||||||||||||||||||
|
Adapter Read 2 |
Text |
Required Field = Yes Read Only = No Default = CTGTCTCTTATACACATCT |
||||||||||||||||||||||||||
|
BaseSpace Sequence Hub Analysis Workflow |
Text |
Required Field = Yes Read Only = No Default = TSS Connect v1.1 |
||||||||||||||||||||||||||
|
BaseSpace Sequence Hub Project Name |
Text |
Required Field = Yes Read Only = No |
||||||||||||||||||||||||||
|
Enable Secondary Analysis |
Toggle Switch |
Required Field = Yes Read Only = No Default = Yes |
||||||||||||||||||||||||||
|
Liquid Handling Instrument ID |
Text |
Required Field = No Read Only = Yes |
||||||||||||||||||||||||||
|
Required Yield Gbp |
Numeric |
Required Field = Yes Read Only = No Default = 90 |
||||||||||||||||||||||||||
|
SampleSheet Version |
Text Dropdown |
Required Field = Yes Read Only = No Custom Entries = No Presets:
|
||||||||||||||||||||||||||
|
|
Multiline Text Fields |
Quick Reference Checklist: Dilute and Denature |
Multiline Text |
Required Field = No Read Only = Yes Default User tasks for this page:
|
||||||||||||||||||||||||
|
Step File Placeholders |
|
Sample Sheet |
File |
Auto |
||||||||||||||||||||||||
|
Sample Sheet |
Log |
Auto |
||||||||||||||||||||||||||
|
Hamilton Load to Library tube |
File |
Auto |
||||||||||||||||||||||||||
|
Hamilton |
Log |
Auto |
||||||||||||||||||||||||||
|
Biosample Manifest |
File |
Auto |
||||||||||||||||||||||||||
|
Sample Table |
|
Table Heading |
Sample Table |
|||||||||||||||||||||||||
|
Enable QC Flags |
No |
|||||||||||||||||||||||||||
|
Sample Display Default |
Expand |
|||||||||||||||||||||||||||
|
Well Sort Order |
Column |
|||||||||||||||||||||||||||
|
|
Table Columns |
|
|
|
||||||||||||||||||||||||
|
Sample Name |
||||||||||||||||||||||||||||
|
Flowcell Type |
Text Dropdown |
Presets
|
||||||||||||||||||||||||||
|
Loading Workflow Type |
Text Dropdown |
Global Field = Derived Sample Required Field = Yes Read Only = No Custom Entries = No Presets:
|
||||||||||||||||||||||||||
|
Container Name |
||||||||||||||||||||||||||||
|
Well |
||||||||||||||||||||||||||||
|
LIMS ID (Container) |
||||||||||||||||||||||||||||
|
Project Name |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Validate Single Input (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Validate Library Tube Barcode (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Dilute and Denature NovaSeq Standard - Generate Hamilton Library Tube Input File, Validate Run Setup, and Generate SampleSheet (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Dilute and Denature NovaSeq Standard - Set Next Steps to Remove from Workflow (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Dilute and Denature NovaSeq Standard - Routing Script (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
AUTOMATED - NovaSeq Run (IGA v1.0) |
|
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Protocol 3: NovaSeq Xp (IGA v1.0)
Master Step 1: Make Bulk Pool for NovaSeq Xp (IGA v1.0)
Step Type = Pooling
Aliquot Generation = Fixed 1
Input = Indexed or Pool/Captured 96-well plates
Output = Indexed or Pool/Captured 96-well plates (ready to be pooled into a Bulk Pool Plate in the Next Step)
|
Section |
Subsection |
Value |
|||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
|||||||||||||||||||||
|
Expanded View Fields |
|
||||||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
|||||||||||||||||||||
|
Well Sort Order |
Row |
|
Section |
Subsection |
Value |
|---|---|---|
|
Pool Settings |
Label Uniqueness |
On |
|
Defaults |
Sample Grouping |
Group by Containers |
|
Well Sort Order |
Row |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 Well Plate |
|
|
Defaults |
Well Sort Order |
Row |
|
|
Placement Pattern |
Column |
|
Skip Alternating Rows |
No |
|
|
Skip Alternating Columns |
No |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Number of Lanes to Sequence |
Numeric |
Required Field = Yes Read Only = No Decimal Places Displayed = 0 |
||||||||||||||||||||||||
|
Minimum Per Sample Volume (µl) |
Numeric |
Required Field = Yes Read Only = No Default = 2 Decimal Places Displayed = 2 |
||||||||||||||||||||||||||
|
Multiline Text Fields |
Quick Reference Checklist: Make Bulk Pool for NovaSeq Xp. i) pre Liquid Handler |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
|||||||||||||||||||||||||
|
Step File Placeholders |
Calculation |
File |
Auto |
|||||||||||||||||||||||||
|
|
Log |
Auto |
||||||||||||||||||||||||||
|
Hamilton Make Bulk Pool Input |
File |
Auto |
||||||||||||||||||||||||||
|
Hamilton Input File Generation |
Log |
Auto |
||||||||||||||||||||||||||
|
Sample Table |
Table Heading |
Sample Table |
||||||||||||||||||||||||||
|
Enable QC Flags |
No |
|||||||||||||||||||||||||||
|
Sample Display Default |
Expand |
|||||||||||||||||||||||||||
|
|
Well Sort Order |
Row |
||||||||||||||||||||||||||
|
|
Table Columns |
|
|
|
||||||||||||||||||||||||
|
Sample Name |
||||||||||||||||||||||||||||
|
RSB Volume (µl) |
Numeric |
|
||||||||||||||||||||||||||
|
Flowcell Type |
Text Dropdown |
Presets:
|
||||||||||||||||||||||||||
|
Loading Workflow Type |
Text Dropdown |
Presets:
|
||||||||||||||||||||||||||
|
Container Name |
||||||||||||||||||||||||||||
|
Well |
||||||||||||||||||||||||||||
|
LIMS ID (Container) |
||||||||||||||||||||||||||||
|
Project Name |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Validate Inputs Flowcell Type and Single Pool (IGA v1.0) |
Trigger Location |
Pooling |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
NovaSeq Make Bulk Pool - Output Plate BPP Barcode Check (Duplicates Allowed) (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Make Bulk Pool NovaSeq Xp - Reset Loading Workflow Type and Flowcell Type (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Make Bulk Pool NovaSeq Xp - Calculate Volumes and Generate Hamilton Input File (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Set Flowcell Type and Next Step (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Dilute, Denature & ExAmp (IGA v1.0) |
|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Step Type = Standard
Input = Indexed or Pool/Captured 96-well plates
Output = Bulk pooled libraries ready to load to a flowcell.
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
DPX1 |
Illumina |
20019916 |
|
DPX2 |
Illumina |
20019917 |
|
DPX3 |
Illumina |
20019918 |
|
NaOH |
||
|
Resuspension Buffer (RSB) |
||
|
Tris-HCl |
|
Section |
Subsection |
Value |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||||||||
|
Expanded View Fields |
|
|||||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||||||||
|
Well Sort Order |
Row |
|
Section |
Subsection |
Value |
||||||
|---|---|---|---|---|---|---|---|---|
|
File Placeholders |
Hamilton Consolidated MakeBPP Bulk Pool Input File |
|
||||||
|
Upload Hamilton MakeBPP Output File |
|
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 Well Plate |
|
|
Defaults |
Sample Grouping |
Group by Containers |
|
Well Sort Order |
Row |
|
|
|
Placement Pattern |
Column |
|
Skip Alternating Rows |
No |
|
|
Skip Alternating Columns |
No |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
DPX1 Volume (µl) |
Numeric |
Required Field = No Read Only = Yes Decimal Places Displayed = 0 |
||||||||||||
|
DPX2 Volume (µl) |
Numeric |
Required Field = No Read Only = Yes Decimal Places Displayed = 0 |
||||||||||||||
|
DPX3 Volume (µl) |
Numeric |
Required Field = No Read Only = Yes Decimal Places Displayed = 0 |
||||||||||||||
|
Liquid Handling Instrument ID |
Text |
Required Field = No Read Only = Yes |
||||||||||||||
|
Multiline Text Fields |
Quick Reference Checklist: Dilute, Denature and ExAm |
Multiline Text |
Required Field = No Read Only = Yes Default User Tasks for the page:
|
|||||||||||||
|
Step File Placeholders |
Calculation |
File |
Auto |
|||||||||||||
|
|
Log |
Auto |
||||||||||||||
|
Incomplete WPP Hamilton Files (Do not use. Required for next step) |
File |
Auto |
||||||||||||||
|
Sample Table |
Table Heading |
Sample Table |
||||||||||||||
|
Enable QC Flags |
No |
|||||||||||||||
|
Sample Display Default |
Expand |
|||||||||||||||
|
|
Well Sort Order |
Row |
||||||||||||||
|
|
Table Columns |
|
|
|
||||||||||||
|
Sample Name |
||||||||||||||||
|
BP Aliquot Volume (µl) |
Numeric |
Global Field = Derived Sample Required Field = No Read Only = Yes Decimal Places Displayed = 0 |
||||||||||||||
|
Mastermix per Lane (µl) |
Numeric |
Global Field = Derived Sample Required Field = No Read Only = Yes Decimal Places Displayed = 0 |
||||||||||||||
|
NaOH Volume (µl) |
Numeric |
Global Field = Derived Sample Required Field = No Read Only = Yes Decimal Places Displayed = 2 |
||||||||||||||
|
Tris-HCl Volume (µl) |
Numeric |
Global Field = Derived Sample Required Field = No Read Only = Yes Decimal Places Displayed = 2 |
||||||||||||||
|
Flowcell Type |
Text Dropdown |
Global Field = Derived Sample Required Field = Yes Read Only = No Custom Entries = No Presets:
|
||||||||||||||
|
Loading Workflow Type |
Text Dropdown |
Global Field = Derived Sample Required Field = Yes Read Only = No Custom Entries = No Presets:
|
||||||||||||||
|
Container Name |
||||||||||||||||
|
Well |
||||||||||||||||
|
LIMS ID (Container) |
||||||||||||||||
|
Project Name |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Dil/Den/ExAmp NovaSeq Xp - Generate Hamilton Consolidated Bulk Pool MakeBPP Input File and Validate Input Flowcell Type (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Dil/Den/ExAmp NovaSeq Xp - Validate Hamilton BPP Output File Name & Parse Instrument ID (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Dil/Den/ExAmp NovaSeq Xp - Output Plate WPP Barcode Uniqueness Check (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Dil/Den/ExAmp NovaSeq Xp - Calculate Volumes and Generate Incomplete WPP Driver File for Next Step (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Dil/Den/ExAmp NovaSeq Xp - Set Next Steps to Advance (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Load to Flowcell (IGA v1.0) |
|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Master Step 3: Load to Flowcell (IGA v1.0)
Step Type = Standard
Input = Diluted and denatured Working Pool Plate
Output = Loaded flowcell ready for sequencing
|
Section |
Subsection |
Value |
|||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
|||||||||||||||
|
Expanded View Fields |
|
||||||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
|||||||||||||||
|
Well Sort Order |
Row |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
SP |
|
|
S1 |
||
|
S2 |
||
|
S4 |
||
|
Defaults |
Sample Grouping |
Group by Containers |
|
Well Sort Order |
Row |
|
|
|
Placement Pattern |
Column |
|
Skip Alternating Rows |
No |
|
|
Skip Alternating Columns |
No |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Experiment Name |
Text |
Required Field = Yes Read Only = No |
|||||||||||||||||||||
|
BaseSpace Sequence Hub Configuration |
Text Dropdown |
Required Field = Yes Read Only = No Custom Entries = No Presets:
|
|||||||||||||||||||||||
|
Workflow |
Text |
Required Field = No Read Only = Yes Default = GenerateFASTQ |
|||||||||||||||||||||||
|
Workflow Type |
Text Dropdown |
Required Field = Yes Read Only = No Custom Entries = No
|
|||||||||||||||||||||||
|
Index Read 1 |
Numeric Dropdown |
Required Field = Yes Read Only = No Custom Entries = Yes Range = 0 to 20 Decimal Places Displayed = 0 Presets:
|
|||||||||||||||||||||||
|
Index Read 2 |
Numeric Dropdown |
Required Field = Yes Read Only = No Custom Entries = Yes Range = 0 to 20 Decimal Places Displayed = 0 Presets:
|
|||||||||||||||||||||||
|
|
Paired End |
Text Dropdown |
Required Field = Yes Read Only = No Custom Entries = No Presets:
|
||||||||||||||||||||||
|
Read Cycles 1 |
Numeric Dropdown |
Required Field = Yes Read Only = No Custom Entries = Yes Range = 1 to 301 Decimal Places Displayed = 0 Presets:
|
|||||||||||||||||||||||
|
Read Cycles 2 |
Numeric Dropdown |
Required Field = Yes Read Only = No Custom Entries = Yes Range = 0 to 301 Decimal Places Displayed = 0 Presets:
|
|||||||||||||||||||||||
|
|
Use Custom Read 1 Primer |
Toggle Switch |
Required Field = Yes Read Only = No Default = No |
||||||||||||||||||||||
|
Use Custom Read 2 Primer |
Toggle Switch |
Required Field = Yes Read Only = No Default = No |
|||||||||||||||||||||||
|
Use Custom Index Read 1 Primer |
Toggle Switch |
Required Field = Yes Read Only = No Default = No |
|||||||||||||||||||||||
|
|
Output Folder |
Text |
Required Field = Yes Read Only = No |
||||||||||||||||||||||
|
|
Use Custom Recipe |
Toggle Switch |
Required Field = Yes Read Only = No Default = No |
||||||||||||||||||||||
|
Custom Recipe Path |
Text |
Required Field = No Read Only = No |
|||||||||||||||||||||||
|
Library Tube Barcode |
Text |
Required Field = Yes Read Only = No |
|||||||||||||||||||||||
|
Adapter |
Text |
Required Field = Yes Read Only = No Default = CTGTCTCTTATACACATCT |
|||||||||||||||||||||||
|
Adapter Read 2 |
Text |
Required Field = Yes Read Only = No Default = CTGTCTCTTATACACATCT |
|||||||||||||||||||||||
|
BaseSpace Sequence Hub Analysis Workflow |
Text |
Required Field = Yes Read Only = No Default = TSS Connect v1.1 |
|||||||||||||||||||||||
|
BaseSpace Sequence Hub Project Name |
Text |
Required Field = Yes Read Only = No |
|||||||||||||||||||||||
|
Enable Secondary Analysis |
Toggle Switch |
Required Field = Yes Read Only = No Default = Yes |
|||||||||||||||||||||||
|
Required Yield Gbp |
Numeric |
Required Field = No Read Only = No Default = 90 |
|||||||||||||||||||||||
|
SampleSheet Version |
Text Dropdown |
Required Field = Yes Read Only = No Custom Entries = No Presets:
|
|||||||||||||||||||||||
|
Multiline Text Fields |
Quick Reference Checklist: Load to Flowcell |
Multiline Text |
Required Field = No Read Only = Yes Default User tasks for this page:
|
||||||||||||||||||||||
|
Step File Placeholders |
|
Sample Sheet |
File |
Auto |
|||||||||||||||||||||
|
|
|
|
Log |
Auto |
|||||||||||||||||||||
|
Hamilton Input File (BPP to WPP to Flow Cell) |
File |
Auto |
|||||||||||||||||||||||
|
Biosample Manifest |
File |
Auto |
|||||||||||||||||||||||
|
Sample Table |
|
Table Heading |
Sample Table |
||||||||||||||||||||||
|
Enable QC Flags |
No |
||||||||||||||||||||||||
|
Sample Display Default |
Expand |
||||||||||||||||||||||||
|
Well Sort Order |
Row |
||||||||||||||||||||||||
|
|
Table Columns |
|
|
|
|||||||||||||||||||||
|
Sample Name |
|||||||||||||||||||||||||
|
Flowcell Type |
Text Dropdown |
Presets
|
|||||||||||||||||||||||
|
Loading Workflow Type |
Text Dropdown |
Required Field = Yes Read Only = No Custom Entries = No Presets:
|
|||||||||||||||||||||||
|
Container Name |
|||||||||||||||||||||||||
|
Well |
|||||||||||||||||||||||||
|
LIMS ID (Container) |
|||||||||||||||||||||||||
|
Project Name |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Validate Inputs and Selected Container (IGA v1.0) |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Validate Flowcell Barcode with New Mask Fixes (IGA v1.0) |
Trigger Location |
Placement |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Load to Flowcell NovaSeq Xp - Generate Hamilton WPP Input File (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Validate Run Setup and Generate SampleSheet (IGA v1.0) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
AUTOMATED - NovaSeq Run (IGA v1.0) |
|
|
Assign Next Step |
Manual |
|
Must move forward (30 minutes to start sequencing) |
|
Protocol 4: AUTOMATED - NovaSeq Run (IGA v1.0)
Master Step 1: AUTOMATED - NovaSeq Run (IGA v1.0)
Do not add samples to the Ice Bucket, start, or complete the AUTOMATED - NovaSeq Run (IGA v1.0) step. The integration does this automatically.
Step Type = Analysis
Input = Library Tube or Flowcell ready for sequencing
Output = None
|
Kit Name |
Supplier |
|---|---|
|
Buffer Cartridge |
Illumina |
|
Cluster Cartridge |
Illumina |
|
SBS Cartridge |
Illumina |
|
Section |
Subsection |
Value |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Sample Table |
Column Headers |
|
||||||||||||
|
Expanded View Fields |
|
|||||||||||||
|
Defaults |
Sample Grouping |
Group by Containers |
||||||||||||
|
Well Sort Order |
Row |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Run ID |
Text |
Required Field = No Read Only = Yes |
|
Current Cycle |
Numeric |
Required Field = Yes Read Only = No Decimal Places Displayed = 0 |
||
|
Flow Cell ID |
Text |
Required Field = No Read Only = Yes |
||
|
Flow Cell Part Number |
Text |
Required Field = No Read Only = Yes |
||
|
Instrument ID |
Text |
Required Field = No Read Only = Yes |
||
|
Firmware Version |
Text |
Required Field = No Read Only = Yes |
||
|
|
Run Status |
Text |
Required Field = No Read Only = Yes |
|
|
Current Read |
Numeric |
Required Field = No Read Only = Yes Decimal Places Displayed = 0 |
||
|
Flow Cell Side |
Text |
Required Field = No Read Only = Yes |
||
|
|
Flow Cell Lot Number |
Text |
Required Field = No Read Only = Yes |
|
|
Instrument Type |
Text |
Required Field = No Read Only = Yes |
||
|
RTA Version |
Text |
Required Field = No Read Only = Yes |
||
|
|
Output Folder |
Text |
Required Field = No Read Only = Yes |
|
|
|
Loading Workflow Type |
Text |
Required Field = No Read Only = Yes |
|
|
Flow Cell Mode |
Text |
Required Field = No Read Only = Yes |
||
|
Flow Cell Expiration Date |
Date |
Required Field = No Read Only = Yes |
||
|
Instrument Control Software Version |
Text |
Required Field = No Read Only = Yes |
||
|
Multiline Text Fields |
Sequencing Log |
Multiline Text |
Required Field = No Read Only = Yes |
|
|
Step File Placeholders |
|
None |
||
|
Sample Table |
|
Table Heading |
Sample Table |
|
|
Enable QC Flags |
No |
|||
|
Sample Display Default |
Collapse |
|||
|
Well Sort Order |
Row |
|||
|
|
Table Columns |
|
|
|
|
Sample Name |
||||
|
Container Name |
||||
|
Yield PF (Gb) R1 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 2 |
||
|
Yield PF (Gb) R2 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 2 |
||
|
Reads PF (M) R1 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 2 |
||
|
Reads PF (M) R2 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 2 |
||
|
Cluster Density (K/mm^2) R1 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 0 |
||
|
Cluster Density (K/mm^2) R2 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 2 |
||
|
%PF R1 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 2 |
||
|
%PF R2 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 2 |
||
|
% Bases >=Q30 R1 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 2 |
||
|
% Bases >=Q30 R2 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 0 |
||
|
Intensity Cycle 1 R1 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 0 |
||
|
Intensity Cycle 1 R2 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 0 |
||
|
% Phasing R1 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 3 |
||
|
% Phasing R2 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 3 |
||
|
% Prephasing R1 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 3 |
||
|
% Prephasing R2 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 3 |
||
|
% Aligned R1 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 2 |
||
|
% Aligned R2 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 2 |
||
|
% Error Rate R1 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 2 |
||
|
% Error Rate R2 |
Numeric |
Global field - Measurement & Container Required Field = No Read Only = Yes Decimal Places Displayed = 2 |
||
|
Well |
||||
|
LIMS ID (Container) |
||||
|
Project Name |
|
Setting |
Value |
|---|---|
|
None - End of Workflow |
|
|
Assign Next Step |
Manual |