Illumina Preset Protocol - Illumina DNA PCR-Free Library Prep Manual
The Illumina DNA PCR-Free Library Prep Manual workflow includes the following functionality.
| • | Preset Illumina DNA PCR-Free Library Prep Manual protocols that support the preparation of up to 96 dual-indexed paired-end single-stranded libraries from DNA using the Illumina DNA PCR-Free Library Prep workflow. |
| • | Automated calculation of sample and buffer volumes. |
| • | Automated calculation or display of reagents at every step in the protocol. |
| • | Automatic step transition when required. |
| • | Automatic placement of samples when necessary. |
| • | Automated assignment of QC Pass/Fail, based on user-selected threshold values. |
Protocol 1: Sample Prep (Illumina DNA PCR-Free Library Prep Manual)
Protocol Type: Sample Prep
Master Step 1: Sort Sample (Illumina DNA PCR-Free Library Prep Manual)
Step Type: No Outputs
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
||||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||||||||
|
Sample Table |
Table Columns |
Sample Type for DNA PCR-Free Lib Prep |
Text Dropdown |
Presets
|
|||||||||||||||||||||
|
Protocol Type |
Text Dropdown |
Presets
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - DNA PCR-Free Lib Prep |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 2: Qubit - Hybex (Illumina DNA PCR-Free Library Prep Manual)
Step Type: Standard QC
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Criteria 1 - Source Data Field |
Text |
Default
|
||||||
|
Criteria 1 - Operator |
Text |
Default
|
||||||||
|
Criteria 1 - Threshold Value |
Numeric |
Default
|
||||||||
|
Criteria 2 - Source Data Field |
Text |
Default
|
||||||||
|
Criteria 2 - Operator |
Text |
Default
|
||||||||
|
Criteria 2 - Threshold Value |
Numeric |
Default
|
||||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Default Read Only
|
|||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
QC Log File |
Auto |
|||||||||
|
QC Result File |
Manual |
|||||||||
|
Upload File |
Manual |
|||||||||
|
Sample Table |
Well sort order |
Column |
||||||||
|
Table Columns |
||||||||||
|
Concentration |
Numeric |
Global field - derived sample & measurement |
||||||||
|
Conc. Units |
Text |
Global field - derived sample & measurement |
||||||||
|
Sample Type |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Sample Type for DNA PCR-Free Lib Prep |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Assign QC Flags |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Set Next Step-REMOVE and Copy Concentration & Sample Type |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing - gDNA Hybex |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 3: Qubit - Thermal Cycler (Illumina DNA PCR-Free Library Prep Manual)
Step Type: Standard QC
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Criteria 1 - Source Data Field |
Text |
Default
|
||||||
|
Criteria 1 - Operator |
Text |
Default
|
||||||||
|
Criteria 1 - Threshold Value |
Numeric |
Default
|
||||||||
|
Criteria 2 - Source Data Field |
Text |
Default
|
||||||||
|
Criteria 2 - Operator |
Text |
Default
|
||||||||
|
Criteria 2 - Threshold Value |
Numeric |
Default
|
||||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
|||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
QC Log File |
Auto |
|||||||||
|
QC Result File |
Manual |
|||||||||
|
Upload File |
Manual |
|||||||||
|
Sample Table |
Well sort order |
Column |
||||||||
|
Table Columns |
Concentration |
Numeric |
Global field - derived sample & measurement |
|||||||
|
Conc. Units |
Text |
Global field - derived sample & measurement |
||||||||
|
Sample Type |
Text |
|
||||||||
|
Prep Input Type |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Sample Type for DNA PCR-Free Lib Prep |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Assign QC Flags |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Set Next Step - REMOVE, Set Prep Input Type, Copy Concentration & Sample Type |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing - gDNA Thermal Cycler |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 4: Whole Blood Lysis (Illumina DNA PCR-Free Library Prep Manual)
Step Type: Standard
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina Lysis Reagent Kit |
Illumina |
20042221 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
80% EtOH Prep Date |
Date |
||||||||||||||||||||||||||||||||||
|
Total Samples |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||
|
MLB (µl) |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||
|
PK1 (µl) |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||
|
Nuclease-free water (µl) |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Default Read Only
|
||||||||||||||||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||||||||||||||||||||
|
Sample Table |
|
Well Sort Order |
Column |
|
|||||||||||||||||||||||||||||||||
|
|
Table Columns |
Sample Type |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Whole Blood Lysis Master Mix & Copy Sample Type |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Remove |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing - Whole Blood Lysis |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 5: Dried Blood Spot Lysis (Illumina DNA PCR-Free Library Prep Manual)
Step Type: Standard
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina Lysis Reagent Kit |
Illumina |
20042221 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
Configured on master step |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
80% EtOH Prep Date |
Date |
||||||||||||||||||||||||||||||||||||||||
|
Total Samples |
Numeric |
|
|||||||||||||||||||||||||||||||||||||||||
|
10x Lysis Buffer (µl) |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||||||||
|
PK1 (µl) |
Numeric |
Calculated = Total Samples * 2 * 1.1 |
|||||||||||||||||||||||||||||||||||||||||
|
Nuclease-free water (µl) |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
||||||||||||||||||||||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||||||||||||||||||||||||||
|
Sample Table |
Well Sort Order |
Column |
|||||||||||||||||||||||||||||||||||||||||
|
|
Table Columns |
Sample Type |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Dried Blood Lysis Master Mix & Copy Sample Type |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Remove |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing - Dried Blood Spot Lysis |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 6: Saliva Lysis (Illumina DNA PCR-Free Library Prep Manual)
Step Type: Standard
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina Lysis Reagent Kit |
Illumina |
20042221 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
80% EtOH Prep Date |
Date |
||||||||||||||||||||||||||||||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
||||||||||||||||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||||||||||||||||||||
|
Sample Table |
Well Sort Order |
Column |
|||||||||||||||||||||||||||||||||||
|
|
Table Columns |
Sample Type |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Sample Type for DNA PCR-Free Lib Prep |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
|
|
|
Command line
|
|||
|
Set Next Step - Remove |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing - Saliva Lysis |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Protocol 2: Library Prep - Hybex (Illumina DNA PCR-Free Library Prep Manual)
Protocol Type: Library Prep
Master Step 1: Tagment Genomic DNA (Illumina DNA PCR-Free Library Prep Manual)
Step Type: Standard
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA PCR-Free Library Prep |
Illumina |
24 - 20041794; 96 - 20041795 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
80% EtOH Prep Date |
Date |
User input |
|||||||||||||||
|
Desired DNA Input (ng) |
Numeric |
|
|||||||||||||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||
|
Sample Table |
Well Sort Order |
Column |
|||||||||||||||||
|
|
Table Columns |
Concentration |
Numeric |
|
|||||||||||||||
|
Conc. Units |
Numeric |
|
|||||||||||||||||
|
DNA (µl) |
Numeric |
Read Only
|
|||||||||||||||||
|
RSB (µl) = 25 - DNA (µl) |
|||||||||||||||||||
|
Desired DNA Input (ng) |
Numeric |
Read Only
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Concentration |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
|
|
|
Command line
|
|||
|
Calculate DNA Volume & Copy Desired DNA Input |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 2: Post Tagmentation Cleanup (Illumina DNA PCR-Free Library Prep Manual)
Step Type: No Outputs
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA PCR-Free Library Prep |
Illumina |
24 - 20041794; 96 - 20041795 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
|||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||||||||
|
Sample Table |
Well Sort Order |
Column |
|||||||||||||||||||||||
|
|
Table Columns |
Desired DNA Input (ng) |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Desired DNA Input |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
|
|
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 3: Ligate Indexes (Illumina DNA PCR-Free Library Prep Manual)
Step Type: Add Labels
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA PCR-Free Library Prep |
Illumina |
24 - 20041794; 96 - 20041795 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Setting |
Value |
|---|---|
|
Label Groups |
IDT for Illumina Nextera DNA Unique Dual Indexes Set A for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
IDT for Illumina Nextera DNA Unique Dual Indexes Set B for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
|
IDT for Illumina Nextera DNA Unique Dual Indexes Set C for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
|
IDT for Illumina Nextera DNA Unique Dual Indexes Set D for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
|
IDT for Illumina Nextera DNA Unique Dual Indexes Set A-D for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Total Samples |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||
|
HP3 (µl) |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||
|
Nuclease-free water (µl) |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||
|
Multiline Text Fields |
Instruction Notes for Hybex Protocol |
Multiline Text |
Read Only Default
|
||||||||||||||||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||||||||||||||||||||
|
Sample Table |
Well Sort Order |
Column |
|||||||||||||||||||||||||||||||||||
|
|
Table Columns |
Desired DNA Input (ng) |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Dilution for HP3 & Copy Desired DNA Input |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
|
|
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 4: Clean Up Libraries (Illumina DNA PCR-Free Library Prep Manual)
Step Type: Standard
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA PCR-Free Library Prep |
Illumina |
24 - 20041794; 96 - 20041795 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
80% EtOH Prep Date |
Date |
User select |
|||||||||||||||||||||||||||||||||||||||||||||||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||||||||||||||||||||||||||||||||||||||
|
Sample Table |
Well Sort Order |
Column |
|||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
Table Columns |
Desired DNA Input (ng) |
Numeric |
|
|||||||||||||||||||||||||||||||||||||||||||||||||||
|
Protocol Type |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Desired DNA Input |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
|
|
|
Command line
|
|||
|
Set Next Step-Remove & Protocol Type |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing - Quantify & Pool Libraries |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Protocol 3: Library Prep - Thermal Cycler Standard Input (Illumina DNA PCR-Free Library Prep Manual)
Protocol Type: Library Prep
Master Step 1: Tagment Genomic DNA - Standard Input
Step Type: Standard
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA PCR-Free Library Prep |
Illumina |
24 - 20041794; 96 - 20041795 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
80% EtOH Prep Date |
Date |
User input |
|||||||||||||||
|
Desired DNA Input (ng) |
Numeric |
Required
|
|||||||||||||||||
|
Thermal Cycler Program |
Text |
Default
|
|||||||||||||||||
|
Multiline Text Fields |
Thermal Cycler Program Notes |
Multiline Text |
Default Read Only
|
||||||||||||||||
|
Instruction Notes |
Multiline Text |
Default Read Only
|
|||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||
|
Sample Table |
Well Sort Order |
Column |
|||||||||||||||||
|
|
Table Columns |
Concentration |
Numeric |
|
|||||||||||||||
|
Conc. Units |
Numeric |
|
|||||||||||||||||
|
Sample Type |
Text |
|
|||||||||||||||||
|
DNA (µl) |
Numeric |
|
|||||||||||||||||
|
RSB (µl) |
Numeric |
|
|||||||||||||||||
|
BLT-PF (µl) |
Numeric |
|
|||||||||||||||||
|
Desired DNA Input (ng) |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Sample Type & Concentration |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
|
|
|
Command line
|
|||
|
Calculate Dilution & BLT-PF Volumes and Copy Desired DNA Input |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Set Next Step-Post TAG |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 2: Tagment Genomic DNA - Low Input
Step Type: Standard
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA PCR-Free Library Prep |
Illumina |
24 - 20041794; 96 - 20041795 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
80% EtOH Prep Date |
Date |
User input |
|||||||||||||||
|
Desired DNA Input (ng) |
Numeric |
Required
|
|||||||||||||||||
|
Total Samples |
Numeric |
Read Only
|
|||||||||||||||||
|
BLT-PF (µl) |
Numeric |
Read Only
|
|||||||||||||||||
|
TB1 (µl) |
Numeric |
Read Only
|
|||||||||||||||||
|
Thermal Cycler Program |
Text |
Default = TAG |
|||||||||||||||||
|
Multiline Text Fields |
Thermal Cycler Program Notes |
Multiline Text |
Read Only Default
|
||||||||||||||||
|
Instruction Notes |
Multiline Text |
Read Only Default
|
|||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||
|
Sample Table |
|
Well Sort Order |
Column |
||||||||||||||||
|
|
Table Columns |
Concentration |
Numeric |
|
|||||||||||||||
|
Conc. Units |
Numeric |
|
|||||||||||||||||
|
Sample Type |
Numeric |
|
|||||||||||||||||
|
DNA (µl) |
Numeric |
|
|||||||||||||||||
|
RSB (µl) |
Numeric |
|
|||||||||||||||||
|
Desired DNA Input (ng) |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Tag Master Mix and Copy Concentration & Sample Type |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
|
|
|
Command line
|
|||
|
Calculate Dilution and Copy Desired DNA Input |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 3: Post Tagmentation Cleanup
Step Type: No Outputs
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA PCR-Free Library Prep |
Illumina |
24 - 20041794; 96 - 20041795 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
|||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||
|
Sample Table |
Well Sort Order |
Column |
|||||||||||||||||
|
|
Table Columns |
Desired DNA Input (ng) |
Numeric |
|
|||||||||||||||
|
Sample Type |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Desired DNA Input & Sample Type |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
|
|
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 4: Ligate Indexes - Thermal Cycler (Illumina DNA PCR-Free Library Prep Manual)
Step Type: Add Labels
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA PCR-Free Library Prep |
Illumina |
24 - 20041794; 96 - 20041795 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Setting |
Value |
|---|---|
|
Label Groups |
IDT for Illumina Nextera DNA Unique Dual Indexes Set A for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
IDT for Illumina Nextera DNA Unique Dual Indexes Set B for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
|
IDT for Illumina Nextera DNA Unique Dual Indexes Set C for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
|
IDT for Illumina Nextera DNA Unique Dual Indexes Set D for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
|
IDT for Illumina Nextera DNA Unique Dual Indexes Set A-D for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Total Samples |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||||||
|
HP3 (µl) |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||||||||
|
Nuclease-free water (µl) |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||||||||
|
Thermal Cycler Program |
Text |
Default = ELM |
|||||||||||||||||||||||||||||||||||||||||
|
Multiline Text Fields |
Thermal Cycler Program Notes |
Multiline Text |
Default ELM Program
|
||||||||||||||||||||||||||||||||||||||||
|
Instruction Notes |
Multiline Text |
Default Read Only
|
|||||||||||||||||||||||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||||||||||||||||||||||||||
|
Sample Table |
Well Sort Order |
Column |
|||||||||||||||||||||||||||||||||||||||||
|
|
Table Columns |
Desired DNA Input (ng) |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Dilution for HP3 & Copy Desired DNA Input |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
|
|
|
Command line
|
|||
|
Set Next Step - Clean Up Libraries |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 5: Clean Up Libraries - Thermal Cycler
Step Type: Standard
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA PCR-Free Library Prep |
Illumina |
24 - 20041794; 96 - 20041795 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||||||||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
80% EtOH Prep Date |
Date |
User select |
||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Multiline Text Fields |
Instruction Notes for Standard Input |
Multiline Text |
Read Only Default
|
|||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Sample Table |
Well Sort Order |
Column |
||||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
|
Table Columns |
Desired DNA Input (ng) |
Numeric |
|
||||||||||||||||||||||||||||||||||||||||||||||||||||||
|
Prep Input Type |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Desired DNA Input |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
|
|
|
Command line
|
|||
|
Set Next Step-Remove & Prep Input Type |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing - Quantify & Pool Libraries |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 6: Clean Up Libraries - Thermal Cycler
Step Type: Standard
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA PCR-Free Library Prep |
Illumina |
24 - 20041794; 96 - 20041795 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
80% EtOH Prep Date |
Date |
User select |
|||||||||||||||||||||||||||||||||||||||
|
Multiline Text Fields |
Instruction Notes for Low Input |
Multiline Text |
Read Only Default
|
||||||||||||||||||||||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||||||||||||||||||||||||||
|
Sample Table |
Well Sort Order |
Column |
|||||||||||||||||||||||||||||||||||||||||
|
|
Table Columns |
Desired DNA Input (ng) |
Numeric |
|
|||||||||||||||||||||||||||||||||||||||
|
Prep Input Type |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Desired DNA Input |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
|
|
|
Command line
|
|||
|
Set Next Step-Remove & Prep Input Type |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing - Quantify & Pool Libraries |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Protocol 4: Quantify and Pool Libraries - Standard Input (Illumina DNA PCR-Free Library Prep Manual)
Protocol Type: Library Prep
Master Step 1: Pool Samples (Illumina DNA PCR-Free Library Prep Manual)
Step Type: Pooling
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA PCR-Free Library Prep |
Illumina |
24 - 20041794; 96 - 20041795 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Default Read Only
|
||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 2: Qubit (Illumina DNA PCR-Free Library Prep Manual)
Step Type: Standard QC
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Criteria 1 - Source Data Field |
Text |
Default
|
|||
|
Criteria 1 - Operator |
Text |
Default
|
|||||
|
Criteria 1 - Threshold Value |
Numeric |
|
|||||
|
Criteria 2 - Source Data Field |
Text |
Default
|
|||||
|
Criteria 2 - Operator |
Text |
Default
|
|||||
|
Criteria 2 - Threshold Value |
Numeric |
|
|||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||
|
QC Log File |
Auto |
||||||
|
QC Result File |
Manual |
||||||
|
Upload File |
Manual |
||||||
|
Sample Table |
Well Sort Order |
Column |
|||||
|
|
Table Columns |
Concentration |
Numeric |
Global field - derived sample |
|||
|
Conc. Units |
Text |
Global field - derived sample |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Assign QC Flags |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Set Next Step-Remove and Copy Concentration |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Protocol 5: Quantify and Pool Libraries - Low Input (Illumina DNA PCR-Free Library Prep Manual)
Protocol Type: Library Prep
Master Step 1: KAPA Sample Prep - Illumina/Universal
Step Type: Standard
|
Kit Name |
Supplier |
|---|---|
|
10 mM Tris-HCl, pH 8.0 – 8.5 (25°C) |
|
|
0.05% Tween® 20 |
|
Control Name |
Supplier |
Cat. # |
Specification |
|---|---|---|---|
|
KAPA Library Quantification Dilution Control |
Roche |
KK4906-07960417001 |
Concentration = 200 pM |
|
NTC |
Concentration = 0 |
||
|
KAPA DNA Standard 1 |
Roche |
KK4828-07960166001 |
Concentration = 20 pM |
|
KAPA DNA Standard 2 |
Roche |
KK4828-07960166001 |
Concentration = 2 pM |
|
KAPA DNA Standard 3 |
Roche |
KK4828-07960166001 |
Concentration = 0.2pM |
|
KAPA DNA Standard 4 |
Roche |
KK4828-07960166001 |
Concentration = 0.02 pM |
|
KAPA DNA Standard 5 |
Roche |
KK4828-07960166001 |
concentration = 0.002 pM |
|
KAPA DNA Standard 6 |
Roche |
KK4828-07960166001 |
concentration = 0.0002 pM |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|
|
Destination Containers |
|
Configured on step |
||||||
|
Well Sort Order |
Column |
Configured on master step |
||||||
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
DNA Dilution Buffer Prep Date |
Date |
|||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
|||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Well Sort Order |
Column |
||||||||
|
|
Sample Display Default |
Expand |
||||||||
|
|
Table Columns |
Desired DNA Input (ng) |
Numeric |
Calculated = copy from previous step |
||||||
|
Expected Concentration (pM) |
Numeric |
|
||||||||
|
Dilution Factor |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Desired DNA Input for KAPA |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Calculate Expected Concentration & Dilution Factor |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Set Next Step - Advance |
|||
|
Trigger Location |
Record Details |
||
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 2: KAPA qPCR Prep & Analysis - Illumina/Universal
Step Type: Standard QC
Measurement Generation = Variable
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
KAPA qPCR library quantification kit - Illumina/Universal |
Roche |
KK4828-07960166001 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
qPCR Master Mix with Primer Prep Date |
Numeric |
||||||||||||||||||||||||||||||||||||||||
|
Reaction Volume (µl) |
Numeric Dropdown |
|
|||||||||||||||||||||||||||||||||||||||||
|
Adding Rox? |
Text Dropdown |
|
|||||||||||||||||||||||||||||||||||||||||
|
Total Samples |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||||||||
|
KAPA SYBR FAST qPCR Master Mix + Primer Premix (µl) |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||||||||
|
ROX (µl) |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||||||||
|
PCR-grade water (µl) |
Numeric |
Read Only
|
|||||||||||||||||||||||||||||||||||||||||
|
Criteria 1 - Source Data Field |
Text |
Default
|
|||||||||||||||||||||||||||||||||||||||||
|
Criteria 1 - Operator |
Text |
Default
|
|||||||||||||||||||||||||||||||||||||||||
|
Criteria 1 - Threshold Value |
Numeric |
Default
|
|||||||||||||||||||||||||||||||||||||||||
|
Criteria 2 - Source Data Field |
Text |
Default
|
|||||||||||||||||||||||||||||||||||||||||
|
Criteria 2 - Operator |
Text |
Default
|
|||||||||||||||||||||||||||||||||||||||||
|
Criteria 2 - Threshold Value |
Numeric |
Default
|
|||||||||||||||||||||||||||||||||||||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Default
|
||||||||||||||||||||||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||||||||||||||||||||||||||
|
QC Log File |
Auto |
||||||||||||||||||||||||||||||||||||||||||
|
QC Result File |
Manual |
||||||||||||||||||||||||||||||||||||||||||
|
Upload File |
Manual |
||||||||||||||||||||||||||||||||||||||||||
|
Sample Table |
Well Sort Order |
Column |
|||||||||||||||||||||||||||||||||||||||||
|
Sample Display Default |
Expand |
||||||||||||||||||||||||||||||||||||||||||
|
|
Table Columns |
qPCR Master Mix (µl) |
Numeric |
|
|||||||||||||||||||||||||||||||||||||||
|
Sample (µl) |
Numeric |
|
|||||||||||||||||||||||||||||||||||||||||
|
Cq |
Numeric |
|
|||||||||||||||||||||||||||||||||||||||||
|
Average Cq |
Numeric |
|
|||||||||||||||||||||||||||||||||||||||||
|
Dilution Factor |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate qPCR Master Mix & Sample Volume |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Calculate Average Cq |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Assign QC Flags |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Set Next Step - Advance & Copy Dilution Factor and Average Cq |
|||
|
Trigger Location |
Record Details |
||
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 3: KAPA Library Quantification - Illumina/Universal
Step Type: Standard QC
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Library Length (bp) |
Numeric |
Default
|
||||||
|
Standard Curve Slope |
Numeric |
User input |
||||||||
|
Standard Curve Y-Intercept |
User input |
|||||||||
|
Library Length Ratio |
Numeric |
Read Only
|
||||||||
|
Criteria 1 - Source Data Field |
Text |
Default
|
||||||||
|
Criteria 1 - Operator |
Text |
Default
|
||||||||
|
Criteria 1 - Threshold Value |
Numeric |
|
||||||||
|
Criteria 2 - Source Data Field |
Text |
Default
|
||||||||
|
Criteria 2 - Operator |
Text |
Default
|
||||||||
|
Criteria 2 - Threshold Value |
Numeric |
|
||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
QC Log File |
Auto |
|||||||||
|
QC Result File |
Manual |
|||||||||
|
Upload File |
Manual |
|||||||||
|
Sample Table |
Well Sort Order |
Column |
||||||||
|
|
Table Columns |
Average Cq |
Numeric |
|
||||||
|
Diluted Avg. Conc. (pM) |
Numeric |
|
||||||||
|
Concentration (nM) |
Numeric |
|
||||||||
|
Dilution Factor |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Library Length Ratio & Copy Dilution Factor and Average Cq |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Calculate Concentration |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Assign QC Flags |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Set Next Step and Copy Concentration |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 4: Dilute Libraries - Low Input (Illumina DNA PCR-Free Library Prep Manual)
Step Type: Standard
|
Kit Name |
Supplier |
|---|---|
|
RSB |
Illumina |
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|
|
Destination Containers |
|
Configured on step |
||||||
|
Well Sort Order |
Column |
Configured on master step |
||||||
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Final Concentration (nM) |
Numeric |
Default = 2 |
||||||
|
Final Volume (µl) |
Numeric |
|||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Well Sort Order |
Column |
||||||||
|
|
Table Columns |
Concentration (nM) |
Numeric |
|
||||||
|
Sample (µl) |
Numeric |
|
||||||||
|
RSB (µl) |
Numeric |
|
||||||||
|
Final Concentration (nM) |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Concentration (nM) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Calculate Dilution Volumes |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 5: Pool Libraries - Low Input (Illumina DNA PCR-Free Library Prep Manual)
Step Type: Pooling
|
Setting |
Value |
Specification |
|---|---|---|
|
Well Sort Order |
Column |
|
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Sample Volume (µl) |
Numeric |
|||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
|||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Well Sort Order |
Column |
||||||||
|
|
Table Columns |
Concentration (nM) |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Final Concentration (nM) |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
End of Protocol