Illumina Preset Protocol - Illumina DNA Prep with Enrichment (S) Tagmentation
Protocol 1: Prep Sample (Illumina DNA Prep with Enrichment (S) Tagmentation)
Master Step 1: Sort Sample (Illumina DNA Prep with Enrichment (S) Tagmentation)
Master Step Type: No Outputs
Input = Blood, Saliva, gDNA, FFPE DNA
Output = Blood Lysis, Saliva Lysis, gDNA, FFPE DNA
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||||||||
|
Sample Table |
Table Columns |
Sample Type |
Text Dropdown |
Global field - derived sample Presets
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Sort & Set Next Step |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 2: Blood Lysis (Illumina DNA Prep with Enrichment (S) Tagmentation)
Step type: Standard
Input = Blood
Output = Blood Lysis
This protocol is expected to generate > 100 ng of DNA output at the end of the blood lysis step.
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Flex Lysis Reagent Kit |
Illumina |
20015884 |
|
Agencourt AMPure XP beads, 5 ml |
Beckman Coulter |
A63880 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
|
|
Placement Pattern |
Column |
|
|
Placement Screen |
Manual |
|
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Text |
Default - BLP |
||||||
|
Total Samples |
Numeric |
Calculated = Total Samples + 1 |
||||||||
|
|
|
BLB (μl) |
Numeric |
Calculated = Total Samples * 7 * 1.1 |
||||||
|
|
PK1 (μl) |
Numeric |
Calculated = Total Samples * 2 * 1.1 |
|||||||
|
Water (μl) |
Numeric |
Calculated = Total Samples * 31 * 1.1 |
||||||||
|
80% EtOH Prep Date |
Date |
|||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Table Columns |
NFE Sample |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Total Samples & Master Mix |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Remove & Copy NFE Sample Type |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Route NFE - Blood |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 3: Saliva Lysis (Illumina DNA Prep with Enrichment (S) Tagmentation)
Master Step Type: Standard
Input = Saliva
Output = Saliva Lysis
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Agencourt AMPure XP beads, 5 ml |
Beckman Coulter |
A63880 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
|
|
Placement Pattern |
Column |
|
|
Placement Screen |
Manual |
|
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
80% EtOH Prep Date |
Date |
|||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Table Columns |
NFE Sample |
Text |
|
||||||
|
Sample Volume (μl) |
Numeric |
|||||||||
|
80% EtOH (μl) |
Numeric |
|||||||||
|
Water for Resuspension (μl) |
Numeric |
Default - 30 |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy NFE Sample Type and Set UDFs |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Remove |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Route NFE - Saliva |
Trigger Location |
Step |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 4: Qubit gDNA QC (Illumina DNA Prep with Enrichment (S) Tagmentation)
Master Step Type: Standard QC
Input = gDNA (10–1000 ng)
Output = Measurement
|
Setting |
Value |
Specification |
|---|---|---|
|
Placement Pattern |
Tube |
|
|
Placement Screen |
Auto |
|
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Criteria 1 - Source Data Field |
Concentration |
|
|
Criteria 1 - Operator |
>= |
|||
|
Criteria 1 - Threshold Value |
||||
|
Criteria 2 - Source Data Field |
Concentration |
|||
|
Criteria 2 - Operator |
<= |
|||
|
Criteria 2 - Threshold Value |
||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
QC Log File |
Auto |
|||
|
QC Result File |
Auto |
|||
|
Upload File |
Manual |
|||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Derived Sample and Measurement custom field |
|
Conc. Units |
Text |
|||
|
NFE Sample |
Text |
Copied from previous step |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Assign QC Flags |
Command line
|
||
|
Set Next Step and Copy Concentration |
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 5: Dilute gDNA (Illumina DNA Prep with Enrichment (S) Tagmentation)
Master Step Type: Standard
Input = gDNA with concentration
Output = Diluted DNA (10-1000 ng)
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
|
|
Placement Pattern |
Column |
|
|
Placement Screen |
Manual |
|
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA Prep -Tagmentation (S) Beads |
Illumina |
20025519; 20025520 |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Desired gDNA Input (ng) |
Numeric |
Default - 10-1000 |
||||||
|
Dilution Buffer |
Text Dropdown |
|
||||||||
|
Notes |
Multiline Text |
Preset = For human gDNA samples and other large complex genomes, the recommended minimum gDNA input is 50 ng. |
||||||||
|
Sample Volume (μl) |
Numeric |
Default - 5 |
||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Copy from previous step |
||||||
|
Conc. Units (ng/μl) |
Text |
|||||||||
|
Total Volume (μl) |
Numeric |
Calculated = (Concentration * Sample Volume (μl)) / Desired gDNA Input (ng) |
||||||||
|
Dilution Volume (μl) |
Numeric |
Calculated = Total Volume (μl) - Sample Volume (μl) |
||||||||
|
Desired gDNA Input (ng) |
Numeric |
Copied from step field |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Concentration |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Calculate Dilution Volume & Total Volume |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Set Next Step - Remove & Copy NFE Sample Type and Desired gDNA Input (ng) |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
Command line
|
|||
|
Route NFE - gDNA |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 6: Qubit FFPE DNA QC (Illumina DNA Prep with Enrichment (S) Tagmentation)
Step Type: Standard QC
Input = FFPE DNA (50–1000 ng)
Output = Measurement
Standard QC
|
Setting |
Value |
|---|---|
|
Destination Containers |
Tube |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Criteria 1 - Source Data Field |
Concentration |
|
|
Criteria 1 - Operator |
>= |
|||
|
Criteria 1 - Threshold Value |
||||
|
Criteria 2 - Source Data Field |
Concentration |
|||
|
Criteria 2 - Operator |
<= |
|||
|
Criteria 2 - Threshold Value |
||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
QC Log File |
Auto |
|||
|
QC Result File |
Auto |
|||
|
Upload File |
Manual |
|||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Global fields - Derived Sample and Measurement |
|
Conc. Units |
Text |
Global fields - Derived Sample and Measurement |
||
|
NFE Sample type |
Text |
Copied from previous step |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Assign QC Flags |
Command line
|
||
|
Set Next Step and Copy Concentration |
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 7: Dilute FFPE DNA (Illumina DNA Prep with Enrichment (S) Tagmentation)
Input = FFPE DNA
Output = Diluted DNA (50-1000 ng)
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA Prep -Tagmentation (S) Beads |
Illumina |
20025519; 20025520 |
|
Setting |
Value |
|---|---|
|
Destination Containers |
96 well plate |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Desired gDNA Input (ng) |
50-1000 |
|||||||
|
Dilution Buffer |
Text Dropdown |
|
||||||||
|
Sample Volume (μl) |
Numeric |
Default - 5 |
||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Copied from previous step |
||||||
|
Conc. Units (ng/μl) |
Text |
|||||||||
|
Total Volume (μl) |
Numeric |
Calculated = (Concentration * Sample Volume (μl)) / (Desired gDNA Input (ng) * 5) |
||||||||
|
Dilution Volume (μl) |
Numeric |
Calculated = Total Volume (μl) - Sample Volume (μl) |
||||||||
|
Desired gDNA Input (ng) |
Numeric |
Copied from step field |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Concentration |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Calculate Dilution Volume & Total Volume |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Set Next Step - Advance & Copy NFE Sample Type and Desired gDNA Input |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Protocol 2: Library Prep (Illumina DNA Prep with Enrichment (S) Tagmentation)
Master Step 1: Tagment Genomic DNA (Illumina DNA Prep with Enrichment (S) Tagmentation)
Input = DNA, Blood Lysis, Saliva Lysis
Output = Tagmented DNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA Prep - Tagmentation (S) Beads |
Illumina |
20025519; 20025520 |
|
Illumina DNA/RNA Prep - Tagmentation PCR Reagents |
Illumina |
20025519; 20025520 |
|
Setting |
Value |
|---|---|
|
Destination Containers |
96 well plate |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Default - TAG |
|||||||||||||
|
Total Samples |
|
|||||||||||||||
|
eBLT (μl) |
Numeric |
Calculated = Total Samples * 11.5 * 1.1 |
||||||||||||||
|
TB1 (μl) |
Numeric |
Calculated = Total Samples * 11.5 * 1.1 |
||||||||||||||
|
Desired Input Amount (ng) |
Numeric |
Range = 50 - 1000 ng |
||||||||||||||
|
Total Volume (μl) |
Numeric |
Default - 30 |
||||||||||||||
|
Thermal Cycler Instructions |
Multiline Text |
Default
|
||||||||||||||
|
Notes |
Multiline Text |
Desired Input Amount should be between 50-1000 ng |
||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||||||||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Global field - derived sample Copied from previous step |
||||||||||||
|
Conc. Units (ng/μl) |
Text |
|||||||||||||||
|
Sample Volume (μl) |
Numeric |
Calculated = Concentration / Desired Input Amount (ng) |
||||||||||||||
|
Nuclease-free water (μl) |
Numeric |
Calculated = Total Volume (μl) - Sample Volume (μl) |
||||||||||||||
|
NFE Sample Type |
Text |
Copied from previous step |
||||||||||||||
|
Desired Input Amount (ng) |
Numeric |
Copied from step field |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Total Samples and Master Mix & Copy Conc. & NFE Sample Type |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Calculate Sample Volume & Water |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 2: Post Tagmentation Cleanup (Illumina DNA Prep with Enrichment (S) Tagmentation)
Step Type: No Outputs
Input = Tagmented DNA
Output = Cleaned Up Tagmented DNA
|
Setting |
Value |
Specification |
|---|---|---|
|
Placement Screen |
None |
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA/RNA Prep - Tagmentation Buffers |
Illumina |
20025519; 20025520 |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 3: Amplify Tagmented DNA (Illumina DNA Prep with Enrichment (S) Tagmentation)
Step Type: Add Labels
Input = Cleaned up Tagmented DNA
Output = Amplifed Tagmented DNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA Fast Hyb - Enrichment PCR + Buffers |
Illumina |
20025519; 20025520 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
|
|
Placement Screen |
None. Uses sample 96 well plate from previous step. |
|
Setting |
Value |
|---|---|
|
Label Groups |
IDT for Illumina DNA-RNA UD Indexes Set D Tagmentation-label-sheet (1).xlsx |
|
IDT for Illumina DNA-RNA UD Indexes Set C Tagmentation-label-sheet (1).xlsx |
|
|
IDT for Illumina DNA-RNA UD Indexes Set B Tagmentation-label-sheet (1).xlsx |
|
|
IDT for Illumina DNA-RNA UD Indexes Set A-D Tagmentation-label-sheet (1).xlsx |
|
|
Location of indexes = UDP Run Setup Resources |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Text |
Default - eBLT PCR |
|||||||||||||||||||||
|
|
|
Total Samples |
Numeric |
|
|||||||||||||||||||||
|
|
EPM (μl) |
Numeric |
Calculated = Total Samples * 23 *1.1 |
||||||||||||||||||||||
|
Nuclease-free water (μl) |
Numeric |
Calculated = Total Samples * 23 *1.1 |
|||||||||||||||||||||||
|
Thermal Cycler Instructions |
Multiline Text |
Default
X cycles of:
|
|||||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||||||||
|
Sample Table |
Table Columns |
PCR Cycles |
Numeric |
Preset with script |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Total Samples, Master Mix, Copy NFE Sample Type and Set PCR Cycles |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 4: Clean Up Libraries (Illumina DNA Prep with Enrichment (S) Tagmentation)
Step Type: Standard
Input = Amplifed Tagmented DNA
Output = Cleaned up libraries
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Agencourt AMPure XP purification beads |
Beckman Coulter |
A63880 |
|
Illumina DNA Prep -Tagmentation (S) Beads |
Illumina |
20025519; 20025520 |
|
Setting |
Value |
|---|---|
|
Destination Containers |
96 well plate |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step Data |
Master Step Fields |
80% EtOH Prep Date |
Date |
|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 5: Bioanalyzer QC Qualify Pre-Enriched Libraries (Illumina DNA Prep with Enrichment (S) Tagmentation)
Step Type: Standard QC
Input = Enriched DNA Fragments
Output = Result File
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Criteria 1 - Source Data Field |
Text |
Default: Peak 2 Size - bp |
||||||
|
Criteria 1 - Operator |
Text |
Default: >= |
||||||||
|
Criteria 1 - Threshold |
Numeric |
Default: 300 |
||||||||
|
Criteria 2 - Source Data Field |
Text |
Default: Peak 2 Size - bp |
||||||||
|
Criteria 2 - Operator |
Text |
Default: <= |
||||||||
|
Criteria 2 - Threshold |
Numeric |
Default: 400 |
||||||||
|
Use strict matching for Bioanalyzer results |
Toggle Switch |
|||||||||
|
Step File Placeholders |
|
File Name |
Bioanalyzer Input File |
Auto |
||||||
|
Bioanalyzer Input File Generation Log File |
Auto |
|||||||||
|
Bioanalyzer XML Result File |
|
|||||||||
|
Result File |
|
|||||||||
|
PDF Summary File |
|
|||||||||
|
Bioanalyzer XML Parsing Log File |
Auto |
|||||||||
|
QC Assignment Log File |
Auto |
|||||||||
|
QC Assignment Report |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Generate Bioanalyzer Input File |
Command line
|
||
|
Parse Bioanalyzer XML and assign QC flags |
Command line
|
|
Setting |
Value |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Start Next Step |
Optional next steps:
|
|||||||||
|
Assign Next Step |
|
|||||||||
|
Safe Stopping Point |
|
Master Step 6: Pool Pre-Enriched Libraries (Illumina DNA Prep with Enrichment (S) Tagmentation)
Step Type: Standard
Input = Diluted Pre-Enriched Libraries
Output = Pooled Pre-Enriched Libraries
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Agencourt AMPure XP purification beads |
Beckman Coulter |
A63880 |
|
Illumina DNA Prep -Tagmentation (S) Beads |
Illumina |
20025519; 20025520 |
|
Setting |
Value |
|---|---|
|
Destination Containers |
Tube |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Library Pool Plexity |
Text Dropdown |
|
||||||
|
Total Pooled Volume (μl) |
Numeric |
Read Only Default = 30 |
||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Table Columns |
Pre-Enriched Library Volume (μl) |
Numeric |
|
||||||
|
RSB Volume (μl) |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Pre-Enriched Library Volume & RSB Volume |
Command line
|
||
|
Set Next Step - Hybridize Probes |
Command line
|
|
Setting |
Value |
||||||
|---|---|---|---|---|---|---|---|
|
Start Next Step |
|
||||||
|
Assign Next Step |
Automatic |
||||||
|
Safe Stopping Point |
|
Master Step 7: Qubit (Illumina DNA Prep with Enrichment (S) Tagmentation)
Step Type: Standard QC
Input = Cleaned Up Libraries
Output = Measurement
|
Setting |
Value |
|---|---|
|
Destination Containers |
Tube |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Criteria 1 - Source Data Field |
Text |
Default: Concentration |
|
|
|
Criteria 1 - Operator |
Text |
Default: >= |
|
Criteria 1 - Threshold Value |
Numeric |
|||
|
Criteria 2 - Source Data Field |
Text |
Default: Concentration |
||
|
Criteria 2 - Operator |
Text |
Default: <= |
||
|
Criteria 2 - Threshold Value |
Numeric |
|||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
QC Log File |
Auto |
|||
|
QC Result File |
Auto |
|||
|
Upload File |
Manual |
|||
|
Sample Table |
Table Columns |
Concentration |
Numeric
|
Global field - derived sample |
|
Conc. Units |
Text |
Global field - derived sample |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Assign QC Flags |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Set Next Step and Copy Concentration |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 8: Pool Pre-Enriched Libraries - Mass (Illumina DNA Prep with Enrichment (S) Tagmentation)
Step Type: Standard
Input = Diluted Pre-Enriched Libraries
Output = Pooled Pre-Enriched Libraries
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Agencourt AMPure XP purification beads |
Beckman Coulter |
A63880 |
|
Illumina DNA Prep -Tagmentation (S) Beads |
Illumina |
20025519; 20025520 |
|
Setting |
Value |
|---|---|
|
Destination Containers |
Tube |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Library Pool |
Text Dropdown |
|
||||||
|
Pre-Enriched Library Sample (ng) |
Numeric |
|
||||||||
|
Total Pooled Volume (μl) |
Numeric |
|
||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
|
||||||
|
Conc. Unit |
Text |
|
||||||||
|
Pre-Enriched Library Volume (μl) |
Numeric |
Calculated = Pre-Enriched Library Sample (ng) / Concentration |
||||||||
|
RSB Volume (μl) |
Numeric |
Calculated = Total Pooled Volume (μl) - Pre-Enriched Library Volume (μl) |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Concentration |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Calculate Pre-Enriched Library Volume & RSB Volume |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 9: Hybridize Probes (Illumina DNA Prep with Enrichment (S) Tagmentation)
Input = Pooled Pre-Enriched Libraries
Output = Hybridized Probes
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA Fast Hyb - Enrichment Beads + Buffers |
Illumina |
20025519; 20025520 |
|
Illumina DNA Fast Hyb - Enrichment PCR + Buffers |
Illumina |
20025519; 20025520 |
|
Illumina Exome Panel (8 Enrichment Reactions) (CEX)[optional] |
Illumina |
20020183 |
|
TruSight One (6 Enrichment Reactions) (TOO)[optional] |
Illumina |
20029227 |
|
TruSight One Expanded (6 Enrichment Reactions) (TOE)[optional] |
Illumina |
20029226 |
|
TruSight Cancer (8 Enrichment Reactions)[optional] |
Illumina |
FC-121-0202 |
|
TruSight Hereditary Cancer (8 Enrichment Reactions)[optional] |
Illumina |
20029551 |
|
TruSight Cardio (8 Enrichment Reactions)[optional] |
Illumina |
20029229 |
|
TruSeq Neurodegeneration (8 Enrichment Reactions)[optional] |
Illumina |
20029550 |
|
Illumina Custom Enrichment Panel via DesignStudio (8 Enrichment Reactions)[optional] |
Illumina |
20025371 |
|
Setting |
Value |
|---|---|
|
Destination Containers |
96 well plate |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Text |
Default: NF-HYB program |
|
PCR Cycles |
Numeric |
|
||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 10: Capture Hybridize Probes (Illumina DNA Prep with Enrichment (S) Tagmentation)
Step Type: Standard
Input = Pooled Pre-Enriched Libraries
Output = Hybridized Probes
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA Fast Hyb - Enrichment PCR + Buffers |
Illumina |
20025519; 20025520 |
|
Illumina DNA Fast Hyb - Enrichment Beads + Buffers |
Illumina |
20025519; 20025520 |
|
Illumina Exome Panel (8 Enrichment Reactions) (CEX)[optional] |
Illumina |
20020183 |
|
TruSight One (6 Enrichment Reactions) (TOO)(optional)[optional] |
Illumina |
20029227 |
|
TruSight One Expanded (6 Enrichment Reactions) (TOE)[optional] |
Illumina |
20029226 |
|
TruSight Cancer (8 Enrichment Reactions)[optional] |
Illumina |
FC-121-0202 |
|
TruSight Hereditary Cancer (8 Enrichment Reactions)[optional] |
Illumina |
20029551 |
|
TruSight Cardio (8 Enrichment Reactions)[optional] |
Illumina |
20029229 |
|
TruSeq Neurodegeneration (8 Enrichment Reactions)[optional] |
Illumina |
20029550 |
|
Illumina Custom Enrichment Panel via DesignStudio (8 Enrichment Reactions)[optional] |
Illumina |
20025371 |
|
Setting |
Value |
|---|---|
|
Destination Containers |
96 well plate or tube |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Total Samples |
Numeric |
|
||||||
|
EE1 (μl) |
Numeric |
Calculated = Total Samples * 28.5 * 1.1 |
||||||||
|
HP3 (μl) |
Numeric |
Calculated = Total Samples * 1.5 *1.1 |
||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Total Samples and Calculate Master Mix |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 11: Amplify Enriched Library (Illumina DNA Prep with Enrichment (S) Tagmentation)
Step Type: Standard
Input = Pooled Pre-Enriched Libraries
Output = Hybridized Probes
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina DNA Fast Hyb - Enrichment PCR + Buffers |
Illumina |
20025519; 20025520 |
|
Setting |
Value |
|---|---|
|
Destination Containers |
96 well plate or tube |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Text |
Default = AMP |
|
PCR Cycles |
Numeric |
|||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Master Step 12: Clean Up Amplified Enriched Library (Illumina DNA Prep with Enrichment (S) Tagmentation)
Input = Amplified enriched libraries
Output = Cleaned up amplified enriched libraries
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Agencourt AMPure XP beads, 5 ml |
Beckman Coulter |
A63880 |
|
Illumina DNA Prep -Tagmentation (S) Beads |
Illumina |
20025519; 20025520 |
|
Setting |
Value |
|---|---|
|
Destination Containers |
96 well plate or tube |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step Data |
Master Step Fields |
80% EtOH Prep Date |
Date |
|
|
Step File Placeholders |
|
File Name |
Log |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
Safe Stopping Point