Illumina Preset Protocol - Respiratory Virus Panel
The Respiratory Virus Panel workflow includes the following functionality:
| • | Preconfigured protocols that support the preparation of upper and lower respiratory specimens for real-time RT-PCR qualitative detection of nucleic acid from the 2019-nCoV. |
| • | Automated calculation of sample and buffer volumes. |
| • | Automated calculation or display of reagents at every step in the protocol. |
| • | Automatic step transition when required. |
| • | Automatic placement of samples when necessary. |
| • | Automated assignment of QC Pass/Fail, based on user-selected threshold values. |
| • | Routing to iSeq, MiniSeq, MiSeq, NextSeq 550, NextSeq 2000. |
Protocol 1: Sort Sample Type (Respiratory Virus Panel)
Master Step 1: Sort Sample Type (Respiratory Virus Panel)
Step Type: No Output
Input = Sample
Output = Sample
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Multiline Text Fields |
Comment |
||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Table Columns |
Respiratory Virus Extraction Protocol |
Required Presets
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Remove |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Routing Script - Extraction |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Protocol 2: QIAamp Viral RNA Mini Kit (Respiratory Virus Panel)
Master Step 1: Sample Concentration (Respiratory Virus Panel)
Step Type: Standard
Input = Plasma, serum, urine, cerebrospinal fluid, bone marrow and other body fluids
Output = Sample at 140 µl
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Sample Volume (mL) |
Numeric |
Default
|
||||||
|
Multiline Text Fields |
Directions |
Read Only Default
|
||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Table Columns |
Purification Protocol |
Required Presets
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Purification |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Master Step 2: Purification of Viral RNA - Spin Protocol (Respiratory Virus Panel)
Step Type: Standard
Input = Concentrated sample
Output = Purified viral RNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Buffer AW1 |
61904 |
|
|
Buffer AW2 |
61904 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on master step and step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Incubation Start Time |
Text |
||||||||||
|
Incubation End Time |
Text |
||||||||||||
|
96-100% EtOH Prep Date |
Date |
||||||||||||
|
Total Samples |
Numeric |
Default = 0 |
|||||||||||
|
Buffer AVL Volume (ml) |
Numeric |
Calculated |
|||||||||||
|
Carrier RNA AVE Volume (µl) |
Numeric |
Calculated |
|||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||
|
Sample Table |
Table Columns |
cDNA Synthesis Protocol |
Required Presets
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Total Samples and Master Mix |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Set Next Step - Remove |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing Script - cDNA synthesis |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Master Step 3: Purification of Viral RNA - Vacuum Protocol (Respiratory Virus Panel)
Step Type: Standard
Input = Concentrated sample
Output = Purified viral RNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
QIAvac 24 Plus vacuum manifold |
19413 |
|
|
Buffer AW1 |
61904 |
|
|
Buffer AW2 |
61904 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on master step and step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Incubation Start Time |
Text |
|||||||||||||
|
Incubation End Time |
Text |
|||||||||||||||
|
96-100% EtOH Prep Date |
Date |
|||||||||||||||
|
Total Samples |
Numeric |
Default = 0 |
||||||||||||||
|
Buffer AVL Volume (ml) |
Numeric |
Calculated |
||||||||||||||
|
Carrier RNA AVE Volume (µl) |
Numeric |
Calculated |
||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||||||||
|
Sample Table |
Table Columns |
cDNA Synthesis Protocol |
Required Presets
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Total Samples and Master Mix |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Set Next Step - Remove |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing Script - cDNA synthesis |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Protocol 3: AllPrep PowerViral DNA/RNA Kit (Respiratory Virus Panel)
Master Step 1: Sort Sample (Respiratory Virus Panel)
Step Type: Standard
Input = Plasma, serum, urine, cerebrospinal fluid, bone marrow and other body fluids
Output = Plasma, serum, urine, cerebrospinal fluid, bone marrow and other body fluids
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Table Columns |
Is Bead Beating Required? |
Required Presets
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Purification |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Master Step 2: Prep Samples (Respiratory Virus Panel)
Step Type: Standard
Input = Concentrated sample
Output = Purified viral RNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
AllPrep PowerViral DNA/RNA Kit |
28000-50 |
|
|
β-mercaptoethanol |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on master step and step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
100% EtOH Open Date |
Date |
||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||
|
Sample Table |
Table Columns |
cDNA Synthesis Protocol |
Required Presets
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Remove |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing Script - cDNA synthesis |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Master Step: Prep Samples - Bead Beating (Respiratory Virus Panel)
Step Type: Standard
Input = Plasma, serum, urine, cerebrospinal fluid, bone marrow and other body fluids
Output = Extracted Sample
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
AllPrep PowerViral DNA/RNA Kit |
28000-50 |
|
|
β-mercaptoethanol |
||
|
Phenol:chloroform:isoamyl alcohol |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on master step and step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
100% EtOH Open Date |
Date |
|||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||||||||
|
Sample Table |
Table Columns |
cDNA Synthesis Protocol |
Required Presets
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Remove |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing Script - cDNA synthesis |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Protocol 4: Maxima H Minus Double-Stranded cDNA Synthesis Kit (Respiratory Virus Panel)
Master Step 1: First Strand cDNA Synthesis (Respiratory Virus Panel)
Step Type: Standard
Input = Total RNA or poly(A)+ mRNA
Output = First Stranded Sample
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Maxima H Minus Double-Stranded cDNA Synthesis Kit |
K2561 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on master step and step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Multiline Text Fields |
Directions |
Preset Please record the volumes for the following fields:
|
||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||
|
Sample Table |
Table Columns |
Volume of Oligo(dT)18 Used (µl) |
Numeric |
2 decimal |
|||||||||
|
Volume of Template RNA Used (µl) |
Numeric |
2 decimal |
|||||||||||
|
Nuclease-free Water Used (µl) |
Numeric |
2 |
|||||||||||
|
4x First Strand Reaction Mix (µl) |
Numeric |
||||||||||||
|
First Strand Enzyme Mix (µl) |
Numeric |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Master Mix UDFs |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Automatic |
|
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Master Step 2: Second Strand cDNA Synthesis (Respiratory Virus Panel)
Step Type: No Output
Input = First Stranded Sample
Output = Second Stranded Sample
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Maxima H Minus Double-Stranded cDNA Synthesis Kit |
K2561 |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Sample Table |
Table Columns |
First Strand cDNA Synthesis (µl) |
Numeric |
2 decimal |
|
5x Second Strand Reaction Mix (µl) |
Numeric |
2 decimal |
||
|
Nuclease-free Water (µl) |
Numeric |
2 |
||
|
Second Strand Enzyme Mix (µl) |
Numeric |
2 decimal |
||
|
Total Volume (µl) |
Numeric |
2 decimal |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Master Mix UDFs |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Master Step 3: Removal of RNA (Respiratory Virus Panel)
Step Type: No Output
Input = Second Stranded Sample
Output = ds cDNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Maxima H Minus Double-Stranded cDNA Synthesis Kit |
K2561 |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
RNase I (µl) to add to each tube |
Numeric |
Default
|
|||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Master Step 4: Purification of ds cDNA (Respiratory Virus Panel)
Step Type: Standard
Input = ds cDNA
Output = purified ds cDNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
GeneJET PCR Purification Kit |
K0701, K0702 |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Sample Table |
Table Columns |
Sample Type |
Text |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Sample Type - ds CDNA output |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Set Next Step - Remove |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing Script - Nextera Flex for Enrichment |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Protocol 5: NEBNext Ultra II First Strand Synthesis Module (Respiratory Virus Panel)
Master Step 1: mRNA Isolation, Fragmentation and Priming (Respiratory Virus Panel)
Step Type: Standard
Input = mRNA
Output = Isolated mRNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
NEBNext Poly(A) mRNA Magnetic Isolation Module |
E7490S, E7490L |
|
|
NEBNext First Strand Synthesis Reaction Buffer |
E7525S, E7525L |
|
|
NEBNext Random Primers |
S1330S |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on master step and step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
MM1: NEBNext First Strand Synthesis Reaction Buffer (µl) |
Numeric |
2 decimal places |
|||
|
|
|
MM1: NEBNext Random Primers (µl) |
Numeric |
2 decimal places |
|||
|
|
MM1: Nuclease-free Water (µl) |
Numeric |
2 decimal places |
||||
|
Beads MM: Oligo dT Beads d(T)25 (µl) |
Numeric |
2 decimal places |
|||||
|
Beads MM: RNA Binding Buffer (2X) (µl) |
Numeric |
2 decimal places |
|||||
|
Total Samples |
Numeric |
Default
|
|||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Total Samples and Master Mix |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Master Step 2: First Strand cDNA Synthesis (Respiratory Virus Panel)
Step Type: Standard
Input = Total RNA
Output = First Stranded Sample
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
NEBNext Ultra II Non-Directional RNA Second Strand Synthesis Module |
E6111S, E6111L |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on master step and step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Murine RNase Inhibitor (µl) |
Numeric |
2 decimal places |
||||||||||||
|
|
|
ProtoScript II Reverse Transcriptase (µl) |
Numeric |
2 decimal places |
||||||||||||
|
|
Nuclease-free Water (µl) |
Numeric |
2 decimal places |
|||||||||||||
|
Total Samples |
Numeric |
Preset
|
||||||||||||||
|
Multiline Text Fields |
Thermal Cycler Program |
Multiline Text |
Default
|
|||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Total Samples and Master Mix |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Protocol 6: NEBNext Ultra II II Non-Directional RNA Second Strand Synthesis Module (Respiratory Virus Panel)
Master Step 1: Second Strand cDNA Synthesis (Respiratory Virus Panel)
Step Type: No Output
Input = First Stranded Sample
Output = ds cDNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
NEBNext Ultra II Non-Directional RNA Second Strand Synthesis Module |
E6111S, E6111L |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
First-Strand Synthesis Product Volume per Well (µl) |
Numeric |
Default
|
|||
|
|
|
NEBNext Second Strand Reaction Buffer (µl) |
Numeric |
2 decimal places |
|||
|
|
NEBNext Second Strand Synthesis Enzyme Mix (µl) |
Numeric |
2 decimal places |
||||
|
Nuclease-free Water (µl) |
Numeric |
2 decimal places |
|||||
|
Total Samples |
Numeric |
Default
|
|||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Total Samples and Master Mix |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Master Step 2: Purification of Double-stranded cDNA (Respiratory Virus Panel)
Step Type: No Output
Input = ds cDNA
Output = Purified ds cDNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
NEBNext Ultra II Non-Directional RNA Second Strand Synthesis Module |
E6111S, E6111L |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step Data |
Master Step Fields |
80% EtOH Date |
Date |
|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Sample Type - ds cDNA |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Set Next Step - Remove |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing Script - Nextera Flex for Enrichment INPUTS |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Protocol 7: SuperScript Double-Stranded cDNA Synthesis Kit (Respiratory Virus Panel)
Master Step 1: First-Strand Synthesis (Respiratory Virus Panel)
Step Type: Standard
Input = Total RNA or mRNA to first-strand cDNA
Output = First-strand cDNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
SuperScript Double-Stranded cDNA Synthesis Kit |
11917-010, 11917-020 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on master step and step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Multiline Text Fields |
Directions |
Default
|
|||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Table Columns |
Prep Sample Type |
Text Dropdown |
Presets
|
||||||
|
mRNA/Total RNA (ug) |
Numeric |
Required
|
||||||||
|
|
Primer (100 pmol/µl) |
Numeric |
2 decimal places |
|||||||
|
RNA in DEPC-treated water µl) |
Numeric |
2 decimal places |
||||||||
|
5X First-Strand Reaction Buffer (µl) |
Numeric |
0 decimal |
||||||||
|
0.1M DTT (µl) |
Numeric |
|||||||||
|
10 mM dNTP mix (µl) |
Numeric |
|||||||||
|
SuperScript II RT (µl) |
Numeric |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Master Mix |
Trigger Location |
Record Details |
|
|
Trigger Style |
Button |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|
Master Step 2: Second-Strand Synthesis (Respiratory Virus Panel)
Step Type: Standard
Input = First strand cDNA
Output = ds cDNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
SuperScript Double-Stranded cDNA Synthesis Kit |
11917-010, 11917-020 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on master step and step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
DEPC-treated water per tube (µl) |
Numeric |
Default
|
||||||
|
5X Second-Strand Reaction Buffer per tube (µl) |
Numeric |
Default
|
||||||||
|
10 mM dNTP mix per tube (µl) |
Numeric |
Default
|
||||||||
|
E. coli DNA Ligase (10 U/μL) per tube (µl) |
Numeric |
Default
|
||||||||
|
E. coli DNA Polymerase I (10 U/μL) per tube (µl) |
Numeric |
Default
|
||||||||
|
E. coli RNase H (2 U/μL) per tube (µl) |
Numeric |
Default
|
||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Table Columns |
Sample Type |
Default
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Sample Type - ds cDNA output |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Set Next Step - Remove |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|||
|
Routing Script - Nextera Flex for Enrichment OUTPUTS |
Trigger Location |
Step |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Safe Stopping Point |
|