To ensure that your Illumina instrument warranty remains valid, this integration must be performed and maintained by the BaseSpace Clarity LIMS Support team as needed.
Overview of the BaseSpace Clarity LIMS ↔ Illumina NextSeq Instrument Interface
|
Image Action #
|
Description
|
|
0
|
NextSeq Run (NextSeq) 1.0 initiated in the LIMS.
|
|
1
|
|
1.
|
End Run event fires on Illumina instrument. |
|
2.
|
Instrument fires configured LIMS script. |
|
3.
|
Script writes End Run event to NAS network root folder in \gls_events . |
|
|
2
|
|
1.
|
Timer-based daemon on the LIMS server wakes up and looks for events. |
|
2.
|
Daemon discovers End Run event and processes it. |
|
3.
|
Daemon uses the supplied process run directory to locate the run folder on the NAS. |
|
4.
|
Daemon parses key information from the run folder including the Reagent Cartridge ID. |
|
5.
|
Daemon uses LIMS REST API to do the following: |
|
–
|
Populate step UDFs: Finish Date, Run Type, Status, Flowcell ID, Experiment Name, Read 1 Cycles, Index 1 Read Cycles, Index 2 Read Cycles, Read 2 Cycles, Output Folder, Run ID, Reagent Cartridge ID, PR2 Bottle ID, Chemistry, Workflow |
|
–
|
Examine list of files to collect, and attach any that are available: |
|
–
|
RunInfo.xml, RunParameters.xml, Link to Run Folder.html |
|
–
|
Run statistics are parsed from several binary InterOp files and stored (visible in the NextSeq run). |
|
–
|