Illumina Preset Protocol - Illumina COVIDSeq Workflow
The Illumina COVIDSeq Workflow workflow includes the following functionality:
| • | Preset Illumina COVIDSeq Workflow protocols that track samples through the quantitative detection of SARS-CoV-2 RNA from nasopharyngeal (NP), oropharyngeal (OP), and nasal swab as well as mid-turbinate specimens collected from individuals who meet COVID-19 clinical or epidemiological criteria. |
| • | Automated calculation of sample and buffer volumes. |
| • | Automated calculation or display of reagents at every step in the protocol. |
| • | Automatic step transition when required. |
| • | Automatic placement of samples when necessary. |
| • | Automated assignment of QC Pass/Fail, based on user-selected threshold values. |
| • | Sample preparation for use on the NovaSeq 6000 Sequencing System. |
| • | Parsing of analysis results into Clarity LIMS. |
Protocol 1: RNA Extraction (Illumina COVIDSeq Workflow)
Master Step 1: Prep Sample - COVIDSeq
Step Type: Standard
Input = Purified Total RNA
Output = Denatured purified total RNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina COVIDSeq Positive Control HT |
Illumina |
|
|
Illumina COVIDSeqTest Box 4 |
Illumina |
20043436 |
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
1.7 mL Tube |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Placement Screen |
Auto |
Configured on step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
|||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||
|
Sample Table |
|
Well Sort Order |
Column |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 2: Spin Purification of RNA - COVIDSeq
Step Type: Standard
Input = Swab sample
Output = Purified viral RNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
QIAmp Viral RNA Mini kit |
Qiagen |
(50) - 52904; (250) - 52906 |
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Placement Screen |
Auto |
Configured on step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
% EtOH |
Numeric |
|
||||||
|
% EtOH Prep Date |
Date |
|||||||||
|
No. of Preps |
Numeric Dropdown |
|
||||||||
|
AW1 Concentrate (mL) |
Numeric |
Logic = If No. of Preps = 50 then 19; if No. of Preps = 250 then 98 |
||||||||
|
% EtOH for AW1 (mL) |
Numeric |
Logic = If No. of Preps = 50 then 25; if No. of Preps = 250 then 130 |
||||||||
|
AW2 Concentrate (mL) |
Numeric |
Logic = If No. of Preps = 50 then 13; if No. of Preps = 250 then 66 |
||||||||
|
% EtOH for AW2 (mL) |
Numeric |
Logic = If No. of Preps = 50 then 30; if No. of Preps = 250 then 160 |
||||||||
|
Total Samples |
Numeric |
|
||||||||
|
Buffer AVL (mL) |
Numeric |
Calculated = 0.56 * Total Samples |
||||||||
|
Carrier RNA-Buffer AVE (uL) |
Numeric |
Calculated = Buffer AVL (mL) * 10 |
||||||||
|
Elution Incubation Start Time |
Text |
|||||||||
|
Elution Incubation End Time |
Text |
|||||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
|||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
|
Well Sort Order |
Column |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Total Samples |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Calculate AW1, AW2 and AVL Reagent Volumes |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual button |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 3: Anneal Sample - COVIDSeq
Step Type: Standard
Input = Purified viral RNA
Output = Annealed samples
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina COVIDSeq Test Box 5 – 3072 Samples |
Illumina |
20043648 |
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Thermal Cycler Program |
Text |
Default
|
||||||||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
|||||||||||||
|
COVIDSeq Anneal Program Notes |
Multiline Text |
Read Only Default
|
||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||||||||
|
Sample Table |
|
Well Sort Order |
Column |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Protocol 2: cDNA Synthesis (Illumina COVIDSeq Workflow)
Master Step 1: Synthesize First Strand cDNA - COVIDSeq
Step Type: Standard
Input = Annealed RNA
Output = cDNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina COVIDSeq Test Box 6 – 3072 Samples |
Illumina |
20043647 |
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step and step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Total Samples |
Numeric |
Default = 0 Calculated = Total Samples + 1 |
||||||||||||||||||
|
|
|
FSA (uL) |
Numeric |
Calculated = 9 * Total Samples |
||||||||||||||||||
|
|
RVT (uL) |
Numeric |
Calculated = 1 * Total Samples |
|||||||||||||||||||
|
Thermal Cycler Program |
Text |
Default = COVIDSeq FSS |
||||||||||||||||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
|||||||||||||||||||
|
COVIDSeq FSS Program Notes |
Multiline Text |
Read Only Default
|
||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||||||||||||||
|
Sample Table |
|
Well Sort Order |
Column |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate First Strand cDNA Master Mix |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Protocol 3: Target Amplification (Illumina COVIDSeq Workflow)
Master Step 1: Amplify cDNA - COVIDSeq
Step Type: Standard
Input = cDNA
Output = Amplified Targeted cDNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina COVIDSeq Test Box 5 – 3072 Samples |
Illumina |
20043648 |
|
Illumina COVIDSEQ Test Box 7 – 3072 Samples |
Illumina |
20043439 |
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step and step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Total Samples |
Numeric |
Default = 0 Calculated = Total Samples + 1 |
||||||||||||||||||
|
|
|
IPM HT for COVIDSeq PCR 1 Master Mix (uL) |
Numeric |
Calculated = 15 * Total Samples |
||||||||||||||||||
|
|
CPP1 HT for COVIDSeq PCR 1 Master Mix (uL) |
Numeric |
Calculated = 4.3 * Total Samples |
|||||||||||||||||||
|
Nuclease-free water for COVIDSeq PCR 1 Master Mix (uL) |
Numeric |
Calculated = 4.7 * Total Samples |
||||||||||||||||||||
|
IPM HT for COVIDSeq PCR 2 Master Mix (uL) |
Numeric |
Calculated = 15 * Total Samples |
||||||||||||||||||||
|
CPP2 HT for COVIDSeq PCR 2 Master Mix (uL) |
Numeric |
Calculated = 4.3 * Total Samples |
||||||||||||||||||||
|
Nuclease-free water for COVIDSeq PCR 2 Master Mix (uL) |
Numeric |
Calculated = 4.7 * Total |
||||||||||||||||||||
|
Thermal Cycler Program |
Text |
Default = COVIDSeq PCR |
||||||||||||||||||||
|
Multiline Text Fields |
COVIDSeq PCR Program Notes |
Multiline Text |
Read Only Default
35 cycles of:
|
|||||||||||||||||||
|
Instruction Notes |
Multiline Text |
Read Only Default
|
||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||||||||||||||
|
Sample Table |
|
Well Sort Order |
Column |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate COVIDSeq PCR 1 and PCR2 Master Mixes |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Automatic |
|
Safe Stopping Point |
|
Protocol 4: Library Prep (Illumina COVIDSeq Workflow)
Master Step 1: Tagment PCR Amplicons - COVIDSeq
Step Type: Standard
Input = Amplified Targeted cDNA
Output = Tagmented amplicons
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina COVIDSeqTest Box 3 – 3072 Samples |
Illumina |
20043646 |
|
Illumina COVIDSEQ Test Box 7 – 3072 Samples |
Illumina |
20043439 |
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step and step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Total Samples |
Numeric |
Default = 0 Calculated = Total Samples + 1 |
||||||||||||
|
|
|
TB1 (uL) |
Numeric |
Calculated = 12 * Total Samples |
||||||||||||
|
|
eBLT (uL) |
Numeric |
Calculated = 4 * Total Samples |
|||||||||||||
|
Nuclease-free water (uL) |
Numeric |
Calculated = 20 * Total Samples |
||||||||||||||
|
Thermal Cycler Program |
Text |
Default = COVIDSeq TAG |
||||||||||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
|||||||||||||
|
COVIDSeq TAG Program Notes |
Multiline Text |
Read Only Default
|
||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||||||||
|
Sample Table |
Well Sort Order |
Column |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Tagmentation Master Mix |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 2: Post Tagmentation Clean Up - COVIDSeq
Step Type: No Input
Input = Tagmented amplicons
Output = Cleaned up tagmented amplicons
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina COVIDSeq Test Box 2 – 3072 Samples |
Illumina |
20043434 |
|
Illumina COVIDSeqTest Box 4 – 3072 Samples |
Illumina |
20043436 |
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Section |
Subsection |
Setting |
Value |
Specification |
|---|---|---|---|---|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|
Sample Table |
Well Sort Order |
Column |
||
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 3: Amplify Tagmented Amplicons - COVIDSeq
Step Type: Add Labels
Input = Cleaned up tagmented amplicons
Output = Amplified tagmented amplicons
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
llumina COVIDSeq Test Box 5 – 3072 Samples |
Illumina |
20043648 |
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
96 well plate |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step and step |
|
Setting |
Value |
|---|---|
|
Label Groups |
IDT for Illumina Nextera DNA Unique Dual Indexes Set A for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
IDT for Illumina Nextera DNA Unique Dual Indexes Set B for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
|
IDT for Illumina Nextera DNA Unique Dual Indexes Set C for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
|
IDT for Illumina Nextera DNA Unique Dual Indexes Set D for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
|
IDT for Illumina Nextera DNA Unique Dual Indexes Set A-D for NovaSeq, MiSeq, HiSeq 2500, HiSeq 2000 |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||||||||||||||||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Total Samples |
Numeric |
|
|||||||||||||||||||||||||||
|
|
|
EPM HT (uL) |
Numeric |
Calculated = 24 * Total Samples |
|||||||||||||||||||||||||||
|
|
Nuclease-free water (uL) |
Numeric |
Calculated = 24 * Total Samples |
||||||||||||||||||||||||||||
|
Thermal Cycler Program |
Text |
Default
|
|||||||||||||||||||||||||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
||||||||||||||||||||||||||||
|
COVIDSeq TAG PCR Program Notes |
Multiline Text |
Read Only Default
7 cycles of:
|
|||||||||||||||||||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||||||||||||||||||||||||||
|
Sample Table |
Well Sort Order |
Column |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate PCR Master Mix |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 4: Pool and Clean Up Libraries - COVIDSeq
Step Type: Pooling
Input = Amplified tagmented amplicons
Output = Pooled and cleaned up libraries
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina COVIDSeqTest Box 3 – 3072 Samples |
Illumina |
20043646 |
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|
|
Destination Containers |
|
Configured on step |
||||||
|
Well Sort Order |
Column |
Configured on master step |
||||||
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
80% EtOH Date |
Date |
|||||||
|
|
|
Total Pooled ITB Tubes |
Numeric |
Calculated = logic (total pooled ITB tubes) |
||||||
|
|
ITB (uL) |
Numeric |
Calculated = 0.9 * (55 * Total Pooled ITB Tubes) |
|||||||
|
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Read Only Default
|
|||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Well Sort Order |
Column |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate ITB Volume |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Protocol 5: Quantification (Illumina COVIDSeq Workflow)
Master Step 1: Quantify Libraries - COVIDSeq Qubit
Step Type: Standard QC
Input = Pooled and cleaned up libraries
Output = Measurement
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Criteria 1 - Source Data Field |
Text |
Default
|
|||
|
Criteria 1 - Operator |
Text |
Default
|
|||||
|
Criteria 1 - Threshold Value |
Numeric |
||||||
|
Criteria 2 - Source Data Field |
Text |
Default
|
|||||
|
Criteria 2 - Operator |
Text |
Default
|
|||||
|
Criteria 2 - Threshold Value |
Numeric |
||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||
|
QC Log File |
Auto |
||||||
|
QC Result File |
Manual |
||||||
|
Upload File |
Manual |
||||||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Global field - derived sample & measurement |
|||
|
Conc. Units |
Text |
Global field - derived sample & measurement |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Assign QC Flags |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Set Next Step and Copy Concentration |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
||||||
|---|---|---|---|---|---|---|---|
|
Start Next Step |
Manual |
||||||
|
Assign Next Step |
Manual |
||||||
|
Must move forward |
Optional
|
Master Step 2: Dilute Libraries - COVIDSeq
Step Type: Standard
Input = DNA with concentration
Output = Diluted DNA
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina COVIDSeqTest Box 3 – 3072 Samples |
Illumina |
20043646 |
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Final Volume (uL) |
Numeric |
Default
|
||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
|
||||||
|
Conc. Units |
Text |
|
||||||||
|
SampleVolume (uL) |
Numeric |
|
||||||||
|
RSB (uL) |
Numeric |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Copy Concentration |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon entry |
||
|
Command line
|
|||
|
Calculate Sample & RSB Volumes |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 3: Quantify Libraries - COVIDSeq Qubit
Step Type: Standard QC
Input = Diluted libraries
Output = Measurement
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Criteria 1 - Source Data Field |
Text |
Default
|
|||
|
Criteria 1 - Operator |
Text |
Default
|
|||||
|
Criteria 1 - Threshold Value |
Numeric |
||||||
|
Criteria 2 - Source Data Field |
Text |
Default
|
|||||
|
Criteria 2 - Operator |
Text |
Default
|
|||||
|
Criteria 2 - Threshold Value |
Numeric |
||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||
|
QC Log File |
Auto |
||||||
|
QC Result File |
Manual |
||||||
|
Upload File |
Manual |
||||||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Global field - derived sample |
|||
|
Conc. Units |
Text |
Global field - derived sample |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Assign QC Flags |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Set Next Step and Copy Concentration |
Trigger Location |
Record Details |
|
|
Trigger Style |
Automatic upon exit |
||
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Manual |
|
Assign Next Step |
Manual |
|
Must move forward |
|
Master Step 3: Normalize Libraries - COVIDSeq
Step Type: Standard
Input = Quantified diluted library
Output = Normalized library
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina COVIDSeqTest Box 3 – 3072 Samples |
Illumina |
20043646 |
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
Final Molarity (nM) |
Numeric |
Default
|
|||
|
Final Volume (uL) |
Numeric |
Default
|
|||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||
|
Sample Table |
Table Columns |
Concentration |
Numeric |
Copied from previous step |
|||
|
Conc. Units |
Text |
Copied from previous step |
|||||
|
Starting Molarity (nM) |
Numeric |
Calculated = Concentration / 660 * 400 |
|||||
|
Library Pool (uL) |
Numeric |
Calculated = (Final Molarity (nM) * Final Volume (uL))/Starting Molarity (nM) |
|||||
|
RSB (uL) |
Numeric |
Calculated = Final Volume (uL) - Library Pool (uL) |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate Molarity, Library Pool & RSB |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 4: Pool and Dilute Libraries - COVIDSeq
Step Type: Pooling
Input = Normalized library
Output = Pooled and diluted libraries
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
Illumina COVIDSeqTest Box 3 – 3072 Samples |
Illumina |
20043646 |
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Tube |
Configured on step |
|
Well Sort Order |
Column |
Configured on master step |
|
Placement Pattern |
Column |
Configured on master step |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||||||||
|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|---|
|
Step Data |
Multiline Text Fields |
Instruction Notes |
Multiline Text |
Default
|
||||||||||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||||||||
|
Sample Table |
Well Sort Order |
Column |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Protocol 6: NovaSeq 6000 Xp Setup (Illumina COVIDSeq Workflow)
Master Step 1: NovaSeq Denature & ExAmp - COVIDSeq
Step Type: Standard
|
Kit Name |
Supplier |
Cat. # |
|---|---|---|
|
DPX1 |
Illumina |
20019916 |
|
DPX2 |
Illumina |
20019917 |
|
DPX3 |
Illumina |
20019918 |
|
Setting |
Value |
Specification |
|---|---|---|
|
Destination Containers |
Configured on step |
|
Section |
Subsection |
Setting |
Value |
Specification |
|||
|---|---|---|---|---|---|---|---|
|
Step Data |
Master Step Fields |
No. of Lanes |
Numeric |
Default
|
|||
|
DPX1 Volume (ul) |
Numeric |
Calculated = 315 * No. of Lanes |
|||||
|
DPX2 Volume (ul) |
Numeric |
Calculated = 45 * No. of Lanes |
|||||
|
DPX3 Volume (ul) |
Numeric |
Calculated = 165 * No. of Lanes |
|||||
|
0.2 N NaOH Prep Date |
Date |
||||||
|
0.2 N NaOH per pool (uL) |
Numeric |
Default
|
|||||
|
400 mM Tris-HCI Prep Date |
Date |
||||||
|
400 mM Tris-HCI per pool (uL) |
Numeric |
Default
|
|||||
|
Flowcell Type |
Numeric |
Default Read Only
|
|||||
|
Step File Placeholders |
|
File Name |
Log |
Auto |
|||
|
Sample Table |
Table Columns |
Flowcell Type |
Numeric |
Calculated = step Flowcell Type |
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
Calculate DPX1, DPX2, DPX3 |
Trigger Location |
Record Details |
|
|
Trigger Style |
Manual Button |
||
|
Command line
|
|||
|
Set Next Step - Advance |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Automatic upon exit |
|
|
|
|
|
Command line
|
|
Setting |
Value |
|---|---|
|
Start Next Step |
Automatic |
|
Assign Next Step |
Automatic |
|
Must move forward |
|
Master Step 2: Load to Flowcell (Illumina COVIDSeq Workflow)
For configuration details, see the Clarity LIMS NovaSeq Integration documentation.
Protocol 7: Automated - NovaSeq Run (Illumina COVIDSeq Workflow)
Master Step 1: Automated - NovaSeq Run (Illumina COVIDSeq Workflow)
For configuration details, see the Clarity LIMS NovaSeq Integration documentation.
Protocol 8: Analysis Results (Illumina COVIDSeq Workflow)
Master Step 1: COVIDSeq analysis results
Step Type: Demultiplexing
|
Setting |
Value |
|---|---|
|
Well Sort Order |
Column |
|
Section |
Subsection |
Setting |
Value |
Specification |
||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
Step File Placeholders |
|
File Name |
Log |
Auto |
||||||
|
COVIDSeq TSV_Run Report File |
Manual |
|||||||||
|
COVIDSeq_RunReport PDF |
Manual |
|||||||||
|
Parsing Log |
Auto |
|||||||||
|
Sample Table |
Well Sort Order |
Column |
||||||||
|
Sample Table |
Table Columns |
Quality Control |
Text |
|
||||||
|
Internal Control |
Text |
|
||||||||
|
Result |
Text |
|
||||||||
|
Consensus sequence |
Text |
|
||||||||
|
Lane |
Numeric |
|
||||||||
|
Index Set |
Text |
|
||||||||
|
Index ID |
Text |
|
||||||||
|
Control Type |
Text |
|
||||||||
|
Human Control |
Text |
|
||||||||
|
SARS-CoV-2 for Control Sample |
Text |
|
||||||||
|
Lane for Control Sample |
Numeric |
|
||||||||
|
Index Set for Control Sample |
Text |
|
||||||||
|
Index ID for Control Sample |
Text |
|
|
Automation Name |
Setting |
Value |
Specification |
|---|---|---|---|
|
CovidSEQ: Parse TSV Run Report |
Trigger Location |
Record Details |
|
|
|
Trigger Style |
Manual Button |
|
|
|
|
|
Command line
|
End of Workflow