Metrics Output
The MetricsOutput.tsv file contains the following quality control metrics for all samples:
This TSV file also includes expanded DNA library QC metrics per sample, based on total reads, collapsed reads, chimeric reads, and on-target reads. Analysis using RNA samples also produces RNA library QC metrics and expanded RNA library QC metrics per sample based on total reads and coverage.
The MetricsOutput.tsv file is a final combined metrics report with sample status, key analysis metrics, and metadata. Sample metrics within the report include suggested lower specification limits (LSL) and upper specification limits (USL) for each sample in the run.
For troubleshooting information, refer to Troubleshooting.
Run metrics from the analysis module indicate the quality of the sequencing run.
Review the following metrics to assess run data quality:
|
Metric |
Description |
Recommended Threshold |
|---|---|---|
|
PCT_PF_READS |
Percentage of reads on the sequencing flow cell that pass the filter. |
≥ 55.0 |
|
PCT_Q30_R1 |
Percentage of bases with a quality score ≥ 30 from Read 1. |
≥ 80.0 |
|
PCT_Q30_R2 |
Percentage of bases with a quality score ≥ 30 from Read 2. |
≥ 80.0 |
The values in the Run Metrics section are listed as NA in the following situations:
| • | If the analysis was started from FASTQ files. |
| • | If the analysis was started from BCL files and the InterOp files are missing or corrupt. |
Review the following metrics to assess sample data quality:
|
Metric |
Description |
Recommended Threshold |
Variant Class |
|---|---|---|---|
|
CONTAMINATION_SCORE and CONTAMINATION_P_VALUE |
The contamination score based on VAF distribution of SNPs.
The contamination p-value used to assess highly rearranged genomes. A p‑value ≤ 0.049 suggests that the sample has large-scale rearrangements that could lead to high-contamination scores without actual sample contamination. The contamination p-value is only needed when the contamination score is above USL. |
Contamination Score ≤
OR
Contamination Score > |
All |
|
MEDIAN_EXON_COVERAGE |
Median exon fragment coverage across all exon bases. |
≥ 1300 |
Small variant, TMB, fusion, MSI |
|
PCT_EXON_1000X |
Percent exon bases with 1000X fragment coverage. |
≥ 80.0 |
Small variant, TMB |
|
GENE_SCALED_MAD |
The median of absolute deviations normalized by gene fold change. |
≤ 0.059* |
CNV |
|
MEDIAN_BIN_COUNT_CNV_TARGET |
The median raw bin count per CNV target. |
≥ 6.0 |
CNV |
* The recommended threshold of 0.059 for GENE_SCALED_MAD only applies to real cell‑free DNA.