Run on Illumina Run Manager
Illumina personnel must install Illumina Run Manager prior to installing DRAGEN TruSight Oncology 500 ctDNA Analysis Application on NovaSeq 6000Dx.
Installation Requirements
The following requirements must be met to install and run DRAGEN TruSight Oncology 500 ctDNA Analysis Application on NovaSeq 6000Dx.
| • | A NovaSeq 6000Dx sequencing instrument with Illumina Run Manager v1.6.2 or later installed and paired with DRAGEN Server v4. |
| • | Administrator privileges on Illumina Run Manager. |
Installation Instructions
Perform the following steps to download and install the DRAGEN TruSight Oncology 500 ctDNA Analysis Application on NovaSeq 6000Dx installation package.
| 1. | Contact Illumina Customer Care for the link to the installation package. |
The link expires after 7 days.
| 2. | Download the installation package with the link provided in the email from Illumina Customer Care. The installation package contains the following files: |
| • | DRAGEN IRES file—drageninstaller_<DRAGEN_VERSION>.el8.x86_64_prod.ires |
| • | Illumina Run Manager—TSO500L_v<APP_VERSION>.iapp |
| 3. | Install the DRAGEN version using the IRES file. |
| a. | Log into Illumina Run Manager as an administrator. |
| b. | From the Settings menu, select DRAGEN. |
| c. | Select Add DRAGEN Installer. |
| d. | Select the DRAGEN IRES file to begin installation. |
| e. | Check the install versions list to verify that the appropriate DRAGEN version is listed. |
| 4. | Install the Illumina Run Manager-compatible application. |
| a. | Log into Illumina Run Manager as an administrator. |
| b. | From the left-hand menu, select Applications. |
| c. | Select the RUO tab, select Install Application, and then select the IAPP file to install the application. |
| d. | Check the RUO Applications list to verify that the appropriate application version is listed. |
You can configure application settings such as required users from this list.
External Storage Configuration
Perform the following steps to specify a location to store analysis results.
| 1. | Log into Illumina Run Manager as an administrator. |
| 2. | From the Settings menu, select External Storage for Analysis Results. |
| 3. | Configure the following parameters: |
| • | Server Location—The path of the preferred output file location. A change to the server location may prompt you to enter the domain, user name, and password to the new server location. |
In addition to storing the analysis results from any sequencing runs, this location contains the input data created when setting up runs. Make sure that the server location has sufficient storage space for your run.
| • | Encryption—Select Require encryption during file transfer. |
| 4. | Select Save to preserve the changes to the external storage output parameters. |