Sample Sheet
A sample sheet (SampleSheet.csv) records information about samples, the corresponding indexes, and other information that dictates the behavior of DRAGEN. The default location of the sample sheet is the input folder. To specify any CSV file in any location, use the command --sample-sheet. When a sample sheet does not exist in the default location and no sample sheet is specified in the command line, DRAGEN produces an error unless the --no-sample-sheet true option is specified (provided for legacy applications with no demultiplexing, adapter trimming, or other sample-sheet-specified settings supported).
In addition to the command line options that control the behavior of BCL conversion, use the [Settings] section in the sample sheet configuration file to specify how the samples are processed. The following are the sample sheet settings for BCL conversion.
DRAGEN supports both sample sheets v1 and v2. The following table displays the different supported options for v1 and v2.
|
Sample Sheet v1 |
Sample Sheet v2 |
|---|---|
|
Supports both [Settings] and [settings]. Neither are required. |
Supports only [BCLConvert_Settings]. Required. |
|
Unrecognized settings trigger a warning. |
Unrecognized settings produce an error and analysis aborts. |
In addition to the command line options that control the behavior of BCL conversion, you can use the [Settings] section in the sample sheet configuration file to specify how the samples are processed. The following are the sample sheet settings for BCL conversion.
DRAGEN does not support the following sample sheet settings from bcl2fastq:
| • | ReverseComplement |
|
Setting |
Default |
Value |
Description |
|---|---|---|---|
|
AdapterBehavior |
trim |
trim, mask |
Whether the adapter should be trimmed or masked. |
|
AdapterRead1 |
Not applicable |
Read 1 adapter sequence containing A, C, G, or T. |
The sequence to trim or mask from the end of Read 1. This option can only be specified if the first genomic read is included according to RunInfo.xml or OverrideCycles. |
|
AdapterRead2 |
Not applicable |
Read 2 adapter sequence containing A, C, G, or T. |
The sequence to trim or mask from the end of Read 2. This option can only be specified if the second genomic read is included according to RunInfo.xml or OverrideCycles. |
|
AdapterStringency |
0.9 |
Float between 0.5 and 1.0 |
The stringency for matching the read to the adapter using the sliding window algorithm. This option can only be specified if AdapterRead1 or AdapterRead2 is specified. |
|
BarcodeMismatchesIndex1 |
1 |
0, 1, or 2 |
The number of allowed mismatches between the first Index Read and index sequence. Can only be specified when index1 is present and used for demultiplexing for all samples according to the index column and the OverrideCycles setting. |
|
BarcodeMismatchesIndex2 |
1 |
0, 1, or 2 |
The number of allowed mismatches between the second Index Read and index sequence. Can only be specified when index2 is present and used for demultiplexing for all samples according to the index2 column and the OverrideCycles setting. |
|
MinimumTrimmedReadLength |
The minimum of 35 and the shortest non-indexed read length. |
0 to the shortest non-indexed read length |
Reads trimmed below this point become masked at that point. Can only be specified when index2 is present and used for demultiplexing for all samples according to the index column and the OverrideCycles setting. |
|
MinimumAdapterOverlap |
1 |
1, 2, or 3 |
Do not trim detected adapter sequences shorter than this value. |
|
MaskShortReads |
The minimum of 22 and MinimumTrimmedReadLength. |
0 to MinimumTrimmedReadLength |
Reads trimmed below this point become masked out. |
|
OverrideCycles |
None |
Y: Specifies a sequencing read I: Specifies an indexing read U: Specifies a UMI length to be trimmed from read |
String used to specify UMI cycles and mask out cycles of a read. |
|
TrimUMI |
true |
true or false (1 or 0) |
If set to false, UMI sequences are not trimmed from output FASTQ reads. The UMI is still placed in sequence header. Can only be enabled if a UMI is present for at least 1 sample according to the OverrideCycles setting. |
|
CreateFastqForIndexReads |
0 |
true or false (1 or 0) |
If set to true, output FASTQ files for index reads as well as genomic reads. Can only be enabled when an index is present and used for demultiplexing according to the index/index2 columns and the OverrideCycles setting. |
|
NoLaneSplitting |
false |
true or false |
If set to true, output all lanes of a flow cell to the same FASTQ files consecutively. |
|
FastqCompressionFormat |
gzip |
gzip, dragen, dragen-interleaved |
Define the compression format: If the value is gzip, output FASTQ.GZ. If the value is dragen, output FASTQ.ORA not interleaving paired reads. If the value is dragen-interleaved, output FASTQ.ORA interleaving paired reads in a single FASTQ.ORA file for a higher compression rate. Specify the directory of the DRAGEN ORA reference files using the --ora-reference command. This can also be controlled using the command line options. |
|
FindAdaptersWithIndels |
false |
true or false |
Use single-indel-detection adapter trimming (for matching default bcl2fastq2 behavior) |
|
IndependentIndexCollisionCheck |
empty |
Integer between 1 and the number of lanes that exist according to the RunInfo.xml. |
Semi-colon-separated list of lanes which will use stricter validation. When enabled for any given lane, a barcode collision among samples in the corresponding lane(s) will be identified if at least one index (index or index2) have a collision. When disabled (default), a barcode collision among samples in the corresponding lane(s) will be identified if both indexes (index and index2) have a collision. |
The OverrideCycles mask elements are semicolon separated. For example:
OverrideCycles,U7N1Y143;I8;I8;U7N1Y143
DRAGEN supports flexible UMI processing during BCL conversion to support more third-party assays, including UMI sequences in index reads and multiple UMI regions per read. UMI sequences are trimmed from FASTQ read sequences and placed in the sequence identifier for each read, as normal.
The following are examples of OverrideCycles settings using 2x151 reads:
|
Setting |
Description |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
OverrideCycles,U7N1Y143;I8;I8;U7N1Y143 |
UMI is composed of the first 7 bp of each genomic read, linked by 1 bp of ignored sequence, and is the format for Illumina nonrandom UMIs, used in the following products:
|
|||||||||
|
OverrideCycles,Y151;I8;U10;Y151 |
Index Read 2 is a 10 bp UMI, and is the format for Agilent XT HS. |
|||||||||
|
OverrideCycles,Y151;I8U9;I8;Y151 |
Index Read 1 contains both an index and a 9 bp UMI, and is the format for IDT Dual Index Adapters with UMIs. |
|||||||||
|
OverrideCycles,U3N2Y146;I8;I8;U3N2Y146 |
UMI is composed of the first 3 bp of each genomic read, linked by 2 bp of ignored sequence, and is the format for UMIs in SureSelect XT HS 2 and IDT xGen Duplex Seq Adapter. |
|||||||||
|
OverrideCycles,Y151;I8;I8;U10N12Y127 |
UMI is at the beginning of Read 2, attached with a linker sequence of length 12. |
When using --no-lane-splitting true or the corresponding sample sheet setting NoLaneSplitting-true, DRAGEN FASTQ file name convention and FASTQ contents match bcl2fastq2 for the same feature.
DRAGEN only supports this mode when the Lanecolumn is specified in the sample sheet to make sure that all samples are present in all lanes in the same order listed. This order is expected for flow cells with no fluidic boundaries between lanes.
When this mode is enabled for a lane, each index (i7 & i5) must resolve to a single barcode within mismatch tolerances, because ambiguities can be resolved by the other index by default. Combinatorial (exact) matches are still allowed. Refer to Demultiplexing for more information.
The data section is required. Headers for the data section should be [Data] or [data] for sample sheet v1 and [BCLConvert_Data] for sample sheet v2. DRAGEN uses columns in the Data section to sort samples and index adapters.
|
Column |
Description |
|---|---|
|
Lane |
When specified, DRAGEN generates FASTQ files only for the samples with the specified lane number. Only one valid integer is allowed, as defined by the RunInfo.xml. |
|
Sample_ID |
The sample ID. |
|
index |
The Index1 (i7) index adapter sequence. The length of string must match the number of first index cycles in RunInfo.xml or number specified in OverrideCycles. Reverse-complement of listed sequence is used if RunInfo has an IsReverseComplement tag with value Y. |
|
index2 |
The Index2 (i5) Index adapter sequence. The length of string must match number of second index cycles in RunInfo.xml or number specified in OverrideCycles. Reverse-complement of listed sequence is used if RunInfo has an IsReverseComplement tag with value Y. |
|
Sample_Project |
Can only contain alphanumeric characters, dashes, and underscores. Duplicate data strings with different cases (eg, sampleProject and SampleProject) are not allowed. If these data strings are used, analysis fails. This column is not used unless you are using the command line option --bcl-sampleproject-subdirectories. See Command Line Options for more information on command line options. |
|
Sample_Name |
If present, and both --sample-name-column-enabled true and --bcl-sampleproject-subdirectories true command lines are used, then output FASTQ files to subdirectories based upon Sample_Project and Sample_ID, and name FASTQ files by Sample_Name |
DRAGEN/bcl-convert 4.1 and later supports the following settings as columns in the BCLConvert_Data section, allowing them to be specified differently for each sample:
| • | OverrideCycles |
| • | BarcodeMismatchesIndex1 |
| • | BarcodeMismatchesIndex2 |
| • | AdapterRead1 |
| • | AdapterRead2 |
| • | AdapterBehavior |
| • | AdapterStringency |
The per-sample settings can be specified by omitting the setting from the BCLConvert_Settings section, and instead adding a column to the BCLConvert_Data section with the setting name. Settings that do not apply to a sample, e.g index2 if i5 is masked out for that sample, must be blank or na in the entry for that sample.
This feature is only supported on version two v2 sample sheets, and setting cannot be specified both globally and per-sample. Specifying OverrideCycles differently per-sample allows mixing of different pools into the same lane, but must follow barcode mismatch constraints for all cycles that are used for demultiplexing by any sample in that lane. DRAGEN software will detect all conflicts between samples at the beginning of the conversion run, even between different pools.
Different strategies, such as UMI indexes and dual-index inputs, can be combined if IndependentIndexCollisionCheck is not enabled.
The following is an example sample sheet using per-sample-settings:
[Header] FileFormatVersion,2
[BCLConvert_Settings] AdapterRead1,AGATCGGAAGAGCACACGTCTGAACTCCAGTCA AdapterRead2,AGATCGGAAGAGCGTCGTGTAGGGAAAGAGTGT
[BCLConvert_Data] Sample_ID,index,index2,OverrideCycles 21599,ATAGAGGC,TATAGCCT,Y151;I8;I8;Y151 21600,na,ATAGAGGC,Y151;U8;I8;Y151 21601,GGCTCTG,CCTATCC,Y151;I7N1;I7N1;Y151 21602,ATTACTCG,GGCTCTGA,Y151;I8;I8;U10Y141
DRAGEN does not support the following settings, and new formats must replace their corresponding old formats, when applicable. Manual changes to the sample sheet can be made to the [Settings] section, but the [Data] section must remain unchanged. If any of the obsolete settings are used in the command line or the sample sheet, DRAGEN aborts and returns an error. Also note that some obsolete settings that were previously specified on the command line are now correctly specified in the sample sheet.
|
Behavior |
Obsolete Settings |
New Settings |
|---|---|---|
|
Designate the adapter sequences for Read 1 and Read 2 and specify the behavior as trim. |
(sample sheet) Adapter, AGATCGGAAGAGCACACGTCTGAACTCCAGTCA OR TrimAdapter, AGATCGGAAGAGCACACGTCTGAACTCCAGTCA |
(sample sheet) AdapterRead1, AGATCGGAAGAGCACACGTCTGAACTCCAGTCA AND AdapterRead2, AGATCGGAAGAGCACACGTCTGAACTCCAGTCA |
|
Designate the same adapter sequence for Read 1 and Read 2 and specify the behavior as mask. |
(sample sheet) MaskAdapter, AGATCGGAAGAGCACACGTCTGAACTCCAGTCA |
(sample sheet) AdapterRead1, AGATCGGAAGAGCACACGTCTGAACTCCAGTCA AND AdapterRead2, AGATCGGAAGAGCACACGTCTGAACTCCAGTCA AND AdapterBehavior, mask |
|
Designate the adapter sequences for Read 1 and Read 2 and specify the behavior as mask. |
(sample sheet) MaskAdapter, AGATCGGAAGAGCACACGTCTGAACTCCAGTCA OR MaskAdapterRead2, AGATCGGAAGAGCGTCGTGTAGGGAAAGAGTGT |
(sample sheet) AdapterRead1, AGATCGGAAGAGCACACGTCTGAACTCCAGTCA AND AdapterRead2, AGATCGGAAGAGCGTCGTGTAGGGAAAGAGTGT AND AdapterBehavior, mask |
|
Designate the adapter sequences for Read 1 and Read 2 and specify the behavior as trim. Also specify 0.5 as the adapter stringency. |
(sample sheet) Adapter, AGATCGGAAGAGCACACGTCTGAACTCCAGTCA OR TrimAdapter, AGATCGGAAGAGCACACGTCTGAACTCCAGTCA (command line) --adapter-stringency 0.5 |
(sample sheet) AdapterRead1, AGATCGGAAGAGCACACGTCTGAACTCCAGTCA AND AdapterRead2, AGATCGGAAGAGCACACGTCTGAACTCCAGTCA(sample sheet) AND AdapterStringency, 0.5 |
|
Behavior |
Obsolete Settings |
New Settings |
|---|---|---|
|
Trim the first 7 bases and last 6 bases of Read 1 for a 151 x 8 x 8 x 151 run. |
(sample sheet) Read1StartFromCycle, 8 Read1EndWithCycle, 145 |
(sample sheet) N7Y137N6;I8;I8;Y151 |
|
|
Obsolete Settings |
New Settings |
|---|---|---|
|
Designate the first 8 cycles of Read 1 and Read 2 as UMIs and trim the trailing base for a 151 x 8 x 8 x 151 run. |
(sample sheet) Read1UMIStartFromCycle, 1 Read1UMILength, 8 Read1StartFromCycle, 10 Read2UMIStartFromCycle, 1 Read2UMILength, 8 Read2StartFromCycle, 10 |
(sample sheet) U8N1Y142;I8;I8;U8N1Y142 |
|
Behavior |
Obsolete Command Line Settings |
New Sample Sheet Settings |
|---|---|---|
|
Allow 1 mismatch in the i7 index sequence and 1 mismatch i5 index sequence. |
--barcode-mismatches 1 OR --barcode-mismatches 1,1 |
BarcodeMismatchesIndex1, 1 AND BarcodeMismatchesIndex2, 1 |
|
Allow 2 mismatches in the i7 index sequence and 2 mismatches in the i5 index sequence. |
--barcode-mismatches 2 OR --barcode-mismatches 2,2 |
BarcodeMismatchesIndex1, 2 AND BarcodeMismatchesIndex2, 2 |
|
Behavior |
Obsolete Command Line Settings |
New Sample Sheet Settings |
|---|---|---|
|
Make sure that all trimmed reads are at least 10 base pairs long after adapter trimming by appending Ns to any read shorter than 10 base pairs. |
--minimium-trimmed-read-length 10 |
MinimumTrimmedReadLength, 10 |
|
Make sure that all trimmed reads below 5 base pairs long are masked with Ns. |
--mask-short-adapter-reads, 5 |
MaskShortReads, 5 |
