Scan BeadChips
Before starting a scan, the lasers must stabilize. Make sure that the iScan Reader has been on for at least 5 minutes before beginning a scan.
After selecting the BeadChips to scan and confirming the settings, start the scan. For instructions on omitting one or more BeadChips from a scan, refer to Omit BeadChips and Stripes from a Scan.
To start a scan:
1. | On the iScan Control Software Setup screen, select Scan. |
The iCS performs
As each section is scanned, image and intensity data are saved on the instrument control computer or in a network location in the output path specified on the iCS Setup screen.

Steps |
Description |
||||||
---|---|---|---|---|---|---|---|
Checks available disk space |
If there is less than 75 GB of free disk space for the image and intensity files, the iCS displays a warning. If there is less than 12 GB of free disk space on the drive, the scan is not performed. |
||||||
Loads the DMAP files from the user-defined input folder for each BeadChip in the carrier |
If any DMAP files are missing, the iCS displays a warning. |
||||||
Calibrates the iScan Reader |
Calibration includes the following processes:
Calibration can take several minutes to complete. If there are defective or dirty sections at any of the three alignment corners, the software attempts to use alternate sections until satisfactory calibration is achieved. If no alternate sections are available, calibration fails and an error message is displayed. For troubleshooting information, refer to iScan Reader Issues. |
||||||
Tilts and aligns the BeadChips in the carrier |
The iScan Reader Autofocus feature records the Z-position (height) of three corners of the BeadChip to determine its current tilt, and adjusts the BeadChip until it is flat. The iScan Reader then identifies the X-Y position (lateral location) of the fiducials (focus points) on the BeadChip edges, and adjusts the stage to align the BeadChips under the optics. |