First Time Setup
The MiSeq i100 Series Control Software guides you though the first time setup. The following sections summarize the configuration settings to be configured during initial setup.
Do not disturb the instrument if the spinning wait cursor is displayed. Disruption of the process can result in an unrecoverable critical system error.
To create accurate run result data, you must set the time zone of the instrument after the installation is complete. Refer to Time Settings.
Administrator Account
You can only create one administrator account during the first time setup. After it is set up, you can create additional administrator accounts. For more information, refer to Add Users.
Instrument Nickname
|
•
|
[Optional] Instrument nickname |
If you enter an instrument nickname, it displays at the bottom of the screen in the MiSeq i100 Series Control Software.
Network Connection
Configuring the network connection during the first time setup procedure is optional, but recommended. If you do not configure the network, you must configure a USB or external storage. You cannot use Illumina Proactive, BaseSpace Sequence Hub, or any other cloud services until the network has been set up.
IP Address
To use a static IP address, manually enter the IP address, or use Dynamic Host Configuration Protocol (DHCP) to automate the assignment of the IP address.
|
•
|
Automatically assign IP address (DHCP) |
|
•
|
Manually enter IP address |
DNS Server
If you are inputting the DNS servers manually, you can include multiple servers by separating them with commas. If the MiSeq i100 is not on the domain, you can search for the domain to get name resolution.
|
•
|
Automatically assign DNS server IP address |
|
•
|
[Optional] Manually enter DNS server IP address |
|
–
|
DNS server(s) IP address |
|
•
|
[Optional] Search domain |
Proxy Server
If a proxy server is enabled, an option to enter a user name and password displays for an authenticated proxy.
|
•
|
[Optional] Enable proxy |
|
–
|
Requires user name and password |
Firewall
If you must access the MiSeq i100 remotely, you must enable ports 80 and 443.
|
•
|
Enable network ports 80 and 443 for remote access |
Illumina Proactive
Illumina Proactive is selected by default.
|
•
|
Send instrument performance data to Illumina. No sequencing data are sent. |
System Checks
After the required configurations are in place, system checks are initiated to ensure the proper functioning of all the MiSeq i100 components. The system checks include testing the flow cell door, internal cooling fan, and reagent loading mechanisms. Do not disturb the instrument as it undergoes the system checks. The system checks use the reusable wet and dry test cartridges, which are included with the MiSeq i100.
Load the reusable test cartridges as follows.
|
1.
|
Select Next to extend the dry tray. |
|
2.
|
Load the dry test cartridge after the dry tray extends. |
|
3.
|
Select Next to retract the dry tray and extend the wet tray. |
|
4.
|
Load the wet test cartridge after the wet tray extends. |
|
5.
|
Select Next to retract the wet tray and start the system checks. |
Do not adjust the trays manually. Doing so can lead to an unrecoverable critical system error.
If the system checks identify any failures, the system checks continue until all the components have been checked. A comprehensive list of the failed components is recorded in the log files. Contact Illumina Technical Support to share the log files and address any issues through troubleshooting.
After the system checks have completed, unload the reusable wet test cartridge and reusable dry test cartridge by selecting Eject Consumables from the Start screen. Store the cartridges at the ambient temperature for future use.
External Storage
Local network storage
Network storage - SMB
|
1.
|
Enter the following information: |
Encryption
|
•
|
Require encryption during file transfer. |
|
•
|
Do not require encryption during file transfer. |
|
2.
|
Select Test configuration to test the network storage connection. |
|
3.
|
After the test completes, select Save. |
Network storage - NFS Storage
|
1.
|
Enter the following information: |
|
2.
|
Select Test configuration to test the network storage connection. |
|
3.
|
After the test completes, select Save. |
USB Storage
Adding a USB drive for external storage is only recommended when MiSeq i100 is not connected to a network. A USB drive can also be used to import Sample Sheets and resource files.
The USB drive must be configured as follows.
|
•
|
Formatted to exFAT or NTFS. |
|
•
|
Contain a folder to be used as the output folder. The folder name cannot contain a space. |
The folder cannot be created in the MiSeq i100 Series Control Software, it must be created before adding the USB to the instrument.
If the USB is encrypted, then enter the password. Do not enter a password if the USB is not encrypted.
Specify default folder
After an external storage location has been added, the MiSeq i100 Series Control Software will take you to the Start screen. A default folder must be set up before you can start to a sequencing run. Use the following steps to set the default folder.
|
1.
|
Select the menu icon in the upper-left corner. |
|
2.
|
Select Settings, and then select External storage. |
|
4.
|
Select a server location from the drop-down list, and then select the volume. |
|
5.
|
Select the desired default output folder from Available folders. |
|
6.
|
[Optional] Enter a folder nickname. |
Cloud storage
If you are subscribed to Professional BaseSpace Sequence Hub (BSSH), the private domain name is required.
|
•
|
[Optional] Private domain name |