Load XLEAP-SBS Consumables Into the Cartridge
This step prepares the XLEAP-SBS cartridge for sequencing by mixing the prefilled reagents and loading diluted libraries and the flow cell.
| 1. | Open the cartridge bag by tearing from the top notch on either side and remove the cartridge from the bag. |
Discard the bag and desiccant (moisture-absorbing pouch).
| 2. | Slowly invert the cartridge 10 times |
Internal components can rattle during inversion, which is normal.
| 3. | [XLEAP-SBS Chemistry] Place the cartridge on a flat surface. Lift the cartridge by the arms approximately 1.5–2.5 in (4–7 cm) above the flat surface as shown, and release onto the bench five times. |
This process reduces air bubbles and stabilizes cartridge components. If cartridge is accidentally dropped, repeat this process.
| 1. | Open the |
If unable to use the flow cell immediately, refer to Return Consumables to Storage.
| 2. | Pull the flow cell out of the package. |
Set aside the foil package and desiccant in case that you must return the flow cell to storage. The desiccant is contained in a pouch at the bottom of the foil package. Discard them when sequencing begins.
| 3. | Hold the flow cell by the gray tab with the label on the tab facing up. |
| 4. | Push to insert the flow cell into |
An audible click indicates that the flow cell is in place. When properly loaded, gray tab protrudes from the cartridge.
| 5. | Pull back and remove the gray tab to expose the flow cell. |
| 1. |
|
| 2. | Discard the pipette tip to prevent contamination. |
| 3. | Add the diluted library to the bottom of the reservoir by slowly lowering the pipette tip to the bottom of reservoir before dispensing. Avoid touching the foil. |
| • | If using the onboard denature and dilute process, add 20 µl diluted library. |
| • | If using the manual denature and dilute process, add 200 µl diluted library. |
