Import Sample Sheet
|
2.
|
Navigate to the Start screen, and then select Start. |
|
4.
|
On the NovaSeq X Plus, select an instrument side for a single-sided run or select both sides for a dual-sided run. |
If a single-sided run is in progress, the idle side of the instrument displays the next available time for a staggered start. For more information, refer to Staggered Start of Runs.
|
5.
|
At the top of the runs list, select Import sample sheet. |
To change the default run setting, refer to Customize System Settings.
|
6.
|
Select Select file, and then open the sample sheet v2 file. |
|
7.
|
Select Review, and then review the run information. |
|
8.
|
To use custom primers, select the appropriate Custom primer checkboxes. |
For more information on custom primers, refer to Custom Primers.
|
9.
|
If the instrument is connected to the cloud, confirm the cloud run setting. |
|
10.
|
Confirm the output folder. |
For runs with cloud storage enabled, the following options are available:
|
•
|
To transfer run data to a local storage folder in addition to the cloud, provide an output folder. |
|
•
|
To only transfer run data to the cloud, select Don't transfer run data to external output folder. |
To change the default output folder, refer to Specify the Default Output Folder Location.
|
11.
|
If performing local secondary analysis, confirm BCL data folder transfer setting. |
To update the default run data management selection, refer to Customize System Settings.
|
12.
|
[Optional] Select a custom recipe file. |
If using the Illumina Stranded Total RNA Prep with Ribo-Zero Plus kit or the Illumina Stranded mRNA Prep kit, selecting the custom recipe is required. Refer to Dark Cycle Sequencing for more information.
|
13.
|
When finished, select Load consumables. |