Attaching a File to a File Placeholder with REST

Compatibility: API version 2 revision 14

Many lab processes create small files to summarize results. This example attaches a file to the LIMS server file storage repository, rather than linking to an existing file on the network. Linking to an existing file is covered in Attaching Files Located Outside the Default File Storage Repository. Both examples are useful in practice, depending on your network, storage architecture, and the size of the file.

The file attachment method used in this example is equivalent to a lab scientist manually importing a file and attaching it to a file placeholder in the LIMS user interface.

Make sure the file exists

Before POSTing to the files resource, you must make sure that the file exists in the location referenced by the content-location element. If the file does not exist in this location, the POST will fail.

Using EPP / automation to attach files

Note that you can also use EPP / automation to handle attaching files. Though this method is not as flexible, it does attach a file automatically. With this method, the attached files are copied to the LIMS server and attached based on a LIMS ID in the created file name. Automation scripts are very powerful, but they are only called when the process / step is created, whereas the code in this example can run at any time.

Prerequisites

You have added samples to the system.
You have defined a process with analytes (derived samples) as inputs and result files as outputs.
You have run the process on some samples.
The file to be attached (example uses results.csv) exists in the same directory from which the script is executed.
You have imported the JSch library on the classpath. JSch is used by some Groovy closures (see fileSFTPHelper) in the attached script for SFTP logic. For more information, see http://www.jcraft.com/jsch/

Code Example

After you run a process with a result file output, a file placeholder icon displays next to the result file. The following image shows this in the LIMS Operations Interface.

Attaching a file to a process output requires the use of glsstorage and files resources.

glsstorage assigns a file location, much like a file placeholder in the user interface. The files resource is used to associate the physical disk location to the ResultFile artifact. In combination, these resources allow very flexible management of files, and the integration of external file manipulation and transfer tools.

Attaching a file requires three main steps:

1. Create a storage location for the file, using a POST to glsstorage. This returns the XML needed to POST to the files resource. The XML includes a content-location.
2. Copy the file to the new storage location.
3. POST the file XML to attach the file to the result file placeholder.

Key concept

The glsstorage POST is a request to create a unique file name (name anddirectory) for a future disk file.

The file name is returned in XML inside the content-location element of a new file resource. Much like a file placeholder in the LIMS client, the new file resource represents a location and not the actual file.

The XML returned from glsstorage is also used to POST to the files resource, to link the file created to the result file placeholder

Step 1. Create a storage location for the file

The first step in attaching a file to the placeholder is to create a location for the file on the LIMS server, using a POST to the glsstorage resource.

A POST to glsstorage requires the attached-to and original-location child elements to be defined within the file XML content. (These are the only two elements required to post to glsstorage.)

The file XML content is created using Groovy's StreamingMarkupBuilder .

Our example code begins by defining the current location of the file in the variable fileOriginalLocation .

In addition, the artifactURI variable is defined as the URI with which to associate the file. In this case, that resource is a ResultFile artifact.

// Create a new file Node using StreamingMarkupBuilder def fileDoc = new StreamingMarkupBuilder().bind { mkp.xmlDeclaration() mkp.declareNamespace(file:'http://genologics.com/ri/file') 'file:file'{ 'attached-to'("${artifactURI}") 'original-location'("${fileOriginalLocation}") } }

Before the POST to glsstorage the XML created above would appear as follows:

<file:file xmlns:file="http://genologics.com/ri/file"> <attached-to>http://yourIPaddress/api/v2/artifacts/AFF853A40AP2</attached-to> <original-location>/home/glsftp/Testing/results.csv</original-location></file:file>

The following code shows the POST to glsstorage. The XML returned by the POST to glsstorage is stored in the variable resolvedFileNode.

// Post the file Node to the API unresolvedFileNode = GLSRestApiUtils.xmlStringToNode(fileDoc.toString()) resolvedFileNode = GLSRestApiUtils.httpPOST(unresolvedFileNode, "${apiIndexURI}/glsstorage", username, password) println GLSRestApiUtils.nodeToXmlString(resolvedFileNode)

The XML returned by the POST method includes a new child element – content-location. The content-location is the new directory and file name on the LIMS server to which the file should be copied.

<file:file xmlns:file="http://genologics.com/ri/file"> <content-location>sftp://yourIPaddress/home/glsftp/Process/2010/9/A54-BMJ-100921-24-2178/AFF853A40AP2-40-2911.csv</content-location> <attached-to>http://yourIPaddress/api/v2/artifacts/AFF853A40AP2</attached-to> <original-location>/home/glsftp/Testing/results.csv</original-location></file:file>

If the POST to glsstorage was unsuccessful, an XML document explaining the error is returned. For example, if the artifact specified by artifactURI does not exist, then the POST will fail, and resolvedFileNode will hold the content shown below:

<exc:exception xmlns:exc="http://genologics.com/ri/exception"> <message>Requested artifact not found</message></exc:exception>

Step 2. Copy the file to the new storage location

The next step is to copy the file from current to new location. As mentioned in Prerequisites, SFTP is used through the fileSFTPHelper closure.

// Copy the file from its current location def fileFinalLocation = new URI(resolvedFileNode.'content-location'[0].text()) File sourceFileObj = new File(fileOriginalLocation) def sftpSuccess = fileSFTPHelper(fileFinalLocation, sourceFileObj, glsftpUsername, glsftpPassword) println("SFTP success : $sftpSuccess")

If there has never been a file attached to the ResultFile placeholder, the directory specified by the content-location element will not exist.

The fileSFTPHelper Groovy closure takes care of creating the directory required, using the destRemoteFileURI parameter. It then copies the file to the new file location using SFTP.

To make sure the copy was successful, you can check that the $sftpSuccess variable was 'true'.

Key concept

The glsstorage methods do not create, move, or delete disk files. Use file operations that fit within your scripting or system automation practices.

Step 3. Attach file to the result file placeholder

After you SFTP the file to the content-location, you can POST resolvedFileNode to the files (list) REST resource.

POSTs to this resource require file XML content with the attached-to, original-location, and content-location child elements defined.

After the POST to the files resource, the file information is attached to the resource specified in the attached-to child element:

// Post the file node to the API returnNode = GLSRestApiUtils.httpPOST(resolvedFileNode, "${apiIndexURI}/files", username, password) println GLSRestApiUtils.nodeToXmlString(returnNode)

Key concept

The POST to the files (list) REST resource associates a physical file with the ResultFile artifact. In the user interface, this POST changes the icon from a placeholder to an attached file.

If the POST was successful the updated XML will be in the returnNode variable, and will contain a LIMS ID and a URI, as shown below:

<file:file uri="http://yourIPaddress/api/v2/files/AFF853A40AP2-40-2908" limsid="AFF853A40AP2-40-2908"> <content-location>sftp://yourIPaddress/home/glsftp/Process/2010/9/A54-BMJ-100921-24-2178/AFF853A40AP2-40-2907.csv</content-location> <attached-to>http://yourIPaddress/api/v2/artifacts/AFF853A40AP2</attached-to> <original-location>/home/glsftp/Testing/results.csv</original-location> <is-published>false</is-published></file:file>

The artifact resource for the result file now contains a file element with the file LIMS ID and URI:

<art:artifact uri="http://yourIPaddress/api/v2/artifacts/AFF853A40AP2?state=20968" limsid="AFF853A40AP2"> <name>Calcaneus-1</name> <type>ResultFile</type> <output-type>ResultFile</output-type> <parent-process uri="http://yourIPaddress/api/v2/processes/A54-BMJ-100921-24-2178" limsid="A54-BMJ-100921-24-2178"/> <qc-flag>UNKNOWN</qc-flag> <sample uri="http://yourIPaddress/api/v2/samples/AFF853A40" limsid="AFF853A40"/> <file:file limsid="AFF853A40AP2-40-2908" uri="http://yourIPaddress/api/v2/files/AFF853A40AP2-40-2908"/></art:artifact>

Expected Output and Results

When the script completes, an attached file icon will display in the LIMS client (Operations Interface shown here). The lab scientist can download the file from this view.

Attachments

cookbookExamples.properties.txt

PostFileToProcess.groovy.txt