MiSeq Integration v8.1 Configuration and User Interaction
The Illumina MiSeq Integration Package v8.1 supports the integration of BaseSpace Clarity LIMS to Illumina MiSeq Sequencing Systems.
The integration allows for automated tracking of an Illumina sequencing run in Clarity LIMS. This capability includes tracking sequencing run status, generating run report, and capturing and parsing run statistics. In addition, this integration provides automated generation of a sample sheet file for use with the MiSeq Control Software (MCS) and Local Run Manager.
The Illumina MiSeq Integration Package v8.1 includes the following features:
| • | Preconfigured MiSeq Sequencing v3.1 workflow that maps to lab protocol steps and sequencing runs. |
| • | Automated generation of a sample sheet for use with the MiSeq Control Software (MCS) and Local Run Manager. |
| • | Supported Local Run Manager analysis modules are GenerateFastQ, Library QC, Resequencing, DNA Enrichment, and DNA Amplicon. |
| • | Automated tracking of the MiSeq sequencing run and parsing of run statistics into Clarity LIMS, including: |
| – | Run status and metrics of sequencing run |
| – | Sequencing run parameters |
| – | Real-Time Analysis (RTA) v1 run directory location and other run-specific information |
This integration also includes a preconfigured Library Prep Validation workflow that is used for validation purposes only. The workflow contains a single-step protocol that models the library prep required to produce normalized libraries that are ready for the MiSeq Sequencing v3.1 workflow. For details, see MiSeq Integration v8.1 Validation and Troubleshooting
The MiSeq Integration Package v8.1 only supports Clarity LIMS v5.1.x to v6.x.
As support for Clarity LIMS 4.x is removed, there is no preconfigured MiSeq Sequencing workflow provided through PreConfigured Workflow Package (PCWP).
Using Illumina Preset Protocols (IPP) v.2.3, install the following workflows:
| • | MiSeq Sequencing v3.1 workflow |
| • | Library Prep Validation v2.0.4 workflow |
The following sections describe the MiSeq Sequencing v3.1 workflow.
For instructions on validating and troubleshooting the Illumina MiSeq Integration Package, see MiSeq Integration v8.1 Validation and Troubleshooting.
| • | The MiSeq Sequencing v3.1 workflow is only compatible with MiSeq Integration Package v8.1. |
| • | MiSeq Integration v8.1 is not supported on Clarity LIMS v4.x and v5.0. |
| • | MiSeq Integration v8.1 is only required for MiSeq Control Software (MCS) v4.0. For MCS v3.1 or earlier, do not upgrade to MiSeq Integration v8.1. This upgrade breaks the integration. |
Before samples are assigned to the MiSeq Sequencing v3.1 workflow, make sure that the following prerequisites are completed:
| • | Samples have been accessioned into Clarity LIMS. |
| • | Samples have been run through QC and library prep. |
| • | Samples have been normalized, and the value is captured in a field called Normalized Molarity (nM). |
For more information on sample accessioning, see Sample Accessioning and Upload and Modify Samples in the Clarity LIMS Core Functionality documentation.
Samples can be assigned to the MiSeq Sequencing v3.1 workflow automatically using a routing script or manually from the Projects & Samples dashboard. See Assign and Process Samples in the Clarity LIMS Core Functionality documentation.
The Illumina MiSeq Integration includes the MiSeq Sequencing v3.1 workflow, which contains a single protocol of the same name.
The protocol includes the following steps and automations.
|
Step |
Automation |
|||||||||
|---|---|---|---|---|---|---|---|---|---|---|
|
|
|||||||||
|
|
|||||||||
|
|
This integration supports BaseSpace Clarity LIMS v5.1 and later.
As of Clarity LIMS v5, the term user-defined field (UDF) has been replaced with custom field in the user interface. However, the API resource is still called UDF.
There are two types of custom fields:
| • | Master step fields—Configured on master steps. Master step fields only apply to the following: |
| • | The master step on which the fields are configured. |
| • | The steps derived from those master steps. |
| • | Global fields—Configured on entities (eg, submitted sample, derived sample, measurement, etc.). Global fields apply to the entire Clarity LIMS system. |