Set Up a Sequencing Run (Manual Mode)

Setting up a run in Manual mode means specifying run parameters in the control software and performing analysis off-instrument using an application of your choice. The software saves data to the output folder for the analysis. Generating FASTQ files requires an additional step. If using custom primers, refer to Custom Primers for run setup instructions.

1. If the system is configured for run analysis, collaboration, and storage using BaseSpace Sequence Hub, create a sample sheet for the run:
a. Download the iSeq 100 System Sample Sheet Template for Manual Mode. Refer to the iSeq 100 Sequencing System Software Downloads page on the Illumina support center for more information.
b. Modify the template as needed. Make sure that:
Index 2 (i5) adapter sequences are in the correct orientation. For orientations, refer to Illumina Adapter Sequences (1000000002694).
Sample sheet values must match control software values. For example, enter 151 in the Read 1 field of both the sample sheet and the Run Setup screen.
c. Save the template in CSV file format.
2. In the control software, select Sequence.

The software opens the door at an angle, ejects the tray, and initiates the series of run setup screens.

3. [Optional] Select Help to view an onscreen prompt.

Help prompts appear on each screen to provide supplementary guidance.